@article{article_1598072, title={Evaluating Polyetheretherketone (PEEK) as a Superior Alternative to Titanium in Middle Ear Implants: A Finite Element Modeling Study}, journal={The Eurasia Proceedings of Science Technology Engineering and Mathematics}, volume={32}, pages={154–159}, year={2024}, DOI={10.55549/epstem.1598072}, author={Khatir, Omar and Benkhettou, Abdelkader and Fekih, Sidi Mohamed and Sahli, Abderahmene and Boudjemaa, Ismail}, keywords={Middel ear, Prosthesis, Ossicles, Vibration analysis, Finite-element method.}, abstract={This study presents a comprehensive finite element analysis (FEA) of hydroxyapatite (HA)-coated middle ear prostheses, evaluating the biocompatibility, mechanical integrity, and material suitability of polyetheretherketone (PEEK), titanium, and stainless steel in middle ear environments. Simulations assessed stress distribution, displacement behavior, and coating performance under auditory frequencies of 250–8000 Hz and an applied acoustic pressure of 90 dB. Results indicated that PEEK closely parallels titanium in stress tolerance and mechanical displacement, demonstrating reduced fabrication complexity and enhanced adaptability for HA coatings. This analysis supports PEEK as a practical and potentially superior alternative, offering insights into innovative material applications for middle ear prosthetic devices.}, publisher={ISRES Publishing}