Araştırma Makalesi
BibTex RIS Kaynak Göster

A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES

Yıl 2024, Cilt: 29 Sayı: 1, 139 - 154, 22.04.2024
https://doi.org/10.17482/uumfd.1269483

Öz

Wireless power transfer (WPT) presents numerous possibilities for recharging electronic devices in challenging environments. Charging of biomedical devices within the body is among the available opportunities. Inductively coupled WPT is a dependable and effective solution for powering these devices. Energy is transferred from the transmitter to the receiver in the inductively coupled WPT system through the use of coils and magnetic coupling. A WPT system was designed for this study, with dimensions of 4 cm by 4 cm, power output of 1 mW, and a frequency of 13.56 MHz. Series Series (SS) topology was selected for its ease of handling and simple architecture. A square coil was selected as the receiver and transmitter coil structure due to its higher coupling factor than circular coils. ANSYS® Maxwell 3D was used to design the coils and perform magnetic analysis. In the ANSYS® HFSS program, the WPT system was placed inside the male human model and the electromagnetic exposure of the WPT on humans was examined. The magnetic scattering of the WPT system was within the safe values specified by IEEE and ICNIRP standards.

Kaynakça

  • 1. Agarwal, K., Jegadeesan, R., Guo, Y. X., and Thakor, N. V. (2017) Wireless power transfer strategies for implantable bioelectronics, IEEE reviews in biomedical engineering, 10, 136-161. doi: 10.1109/RBME.2017.2683520
  • 2. Ağçal, A., Doğan, T. H., and Aksu, G. (2022) The Effects of Operating Frequency on Wireless Power Transfer System Design and Human Health in Electric Vehicles, Electrica, 22(2), 188-197. doi: 10.54614/electrica.2022.22020
  • 3. Brown, W.C. (1969) Experiments involving a microwave beam to power and position a helicopter, IEEE Transactions on Aerospace and Electronic Systems, 5(1), 692-702. doi: 10.1109/TAES.1969.309867
  • 4. Campi, T., Cruciani, S., Palandrani, E., De Santis, V., Hirata, A., and Feliziani, M. (2016) Wireless Power Transfer Charging System for AIMDs and Pacemakers, IEEE Transactions on Microwave Theory and Techniques, 64(2), 633-642. doi: 10.1109/TMTT.2015.2511011
  • 5. Dai, J., and Ludois, D. C. (2015) A survey of wireless power transfer and a critical comparison of inductive and capacitive coupling for small gap applications, IEEE Transactions on Power Electronics, 30(11), 6017-6029. doi: 10.1109/TPEL.2015.2415253
  • 6. Demirci, Y.E. (2020). Kalp Pili Uygulamaları için kablosuz enerji transferi devre tasarımı, Master thesis, Pamukkale Üniversitesi, Türkiye.
  • 7. Hong, S., Jeong, S., Lee, S., Sim, B., Kim, H., and Kim, J., (2020) Low EMF design of cochlear implant wireless power transfer system using a shielding coil”, In 2020 IEEE International Symposium on Electromagnetic Compatibility & Signal/Power Integrity (EMCSI), Reno, 623-625. doi: 10.1109/EMCSI38923.2020.9191460
  • 8. IEEE, (2019) Standard for safety levels with respect to human exposure to electric, magnetic, and electromagnetic fields (0 Hz to 100 kHz), Std, Vol. C95, no. 1™-2019,.
  • 9. Imura, T., and Hori, Y. (2011) Maximizing air gap and efficiency of magnetic resonant coupling for wireless power transfer using equivalent circuit and Neumann formula, IEEE Transactions on industrial electronics, 58(10), 4746-4752. doi: 10.1109/TIE.2011.2112317
  • 10. International Commission on Non-Ionizing Radiation Protection (ICNIRP), (2010) Guidelines for limiting exposure to time-varying electric and magnetic fields for low frequencies (1 Hz–100 kHz),” Health Physics, 99(6), 818–836. doi: 10.1097/HP.0b013e3181f06c86
  • 11. Jegadeesan, R., Nag, S., Agarwal, K., Thakor, N., and Guo, Y.-X. (2015) Enabling wireless powering and telemetry for peripheral nerve implants, IEEE J. Biomed. Health Informat., 19(3), 958–970. doi: 10.1109/JBHI.2015.2424985
  • 12. Karakaya, U., (2007) Motor control via wireless energy and information transfer”, Master thesis, İstanbul Teknik Üniversitesi, Türkiye.
  • 13. Laird, (2023). Flexible Magnetic Sheet MULL6060-300. Erişim Adresi: http:// https://www.laird.com/products/inductive-components-emc-components-and-ferrite-cores/ferrite-sheets/mull- series/mull6060-300 (Erişim Tarihi: 09.11.2023)
  • 14. Mohan, S. S., del Mar Hershenson, M., Boyd, S. P., and Lee, T. H. (1999) Simple accurate expressions for planar spiral inductances, IEEE Journal of solid-state circuits, 34(10), 1419-1424. doi: 10.1109/4.792620
  • 15. RamRakhyani, A. K., Mirabbasi, S., and Chiao, M. (2010) Design and optimization of resonance-based efficient wireless power delivery systems for biomedical implants, IEEE transactions on biomedical circuits and systems, 5(1), 48-63. doi: 10.1109/TBCAS.2010.2072782
  • 16. Sahai, A. and Graham, D. (2001) Optical wireless power transmission at long wavelengths, IEEE International Conference on Space Optical Systems and Applications (ICSOS), Santa Monica, 164- 170. doi: 10.1109/ICSOS.2011.5783662
  • 17. Sample, A. P., Meyer, D. T., and Smith, J. R. (2010) Analysis, experimental results, and range adaptation of magnetically coupled resonators for wireless power transfer, IEEE Transactions on industrial electronics, 58(2), 544-554. doi: 10.1109/TIE.2010.2046002
  • 18. Shuvo, M. M. H., Titirsha, T., Amin, N., and Islam, S. K. (2022) Energy harvesting in implantable and wearable medical devices for enduring precision healthcare. Energies, 15(20), 7495. doi: 10.3390/en15207495
  • 19. Tesla, N. (1905) Art of Transmitting Electrical Energy through Natural Mediums, U.S. Patent No. 787,412.

Vücut İçi İmplant Cihazların Şarjı için Kablosuz Enerji Transfer Sistemi Tasarımı

Yıl 2024, Cilt: 29 Sayı: 1, 139 - 154, 22.04.2024
https://doi.org/10.17482/uumfd.1269483

Öz

Kablosuz enerji transferi (KET), zorlu ortamlarda elektronik cihazları şarj etmek için çok sayıda olanak sunar. Biyomedikal cihazların vücuttan şarj edilmesi de mevcut imkanlar arasındadır. Endüktif bağlantı ile birleştirilmiş kablosuz güç aktarımının kullanılması, bu cihazlara güç sağlamak için güvenilir ve etkili bir çözümdür. Endüktif olarak bağlanmış KET sisteminde vericiden alıcıya enerji, bobinler ve manyetik bağlantı kullanılarak aktarılır. Bu çalışma için 4 cm'ye 4 cm boyutlarında, 1 mW güç çıkışı ve 13,56 MHz frekansta bir WPT sistemi tasarlanmıştır. Seri-Seri (SS) topoloji, kullanım kolaylığı ve basit mimarisi nedeniyle seçilmiştir. Dairesel bobinlere kıyasla daha yüksek kuplaj faktörü nedeniyle alıcı ve verici bobin yapısı olarak kare bir bobin seçilmiştir. Bobinleri tasarlamak ve manyetik analizleri yapmak için ANSYS® Maxwell 3D kullanıldı. ANSYS® HFSS programında KET sistemi erkek insan modelinin içine yerleştirilmiş ve KET'in insanlar üzerindeki elektromanyetik maruziyeti incelenmiştir. KET sisteminin manyetik saçılımının IEEE ve ICNIRP standartlarında belirtilen güvenli değerler içerisinde olduğu görülmüştür.

Kaynakça

  • 1. Agarwal, K., Jegadeesan, R., Guo, Y. X., and Thakor, N. V. (2017) Wireless power transfer strategies for implantable bioelectronics, IEEE reviews in biomedical engineering, 10, 136-161. doi: 10.1109/RBME.2017.2683520
  • 2. Ağçal, A., Doğan, T. H., and Aksu, G. (2022) The Effects of Operating Frequency on Wireless Power Transfer System Design and Human Health in Electric Vehicles, Electrica, 22(2), 188-197. doi: 10.54614/electrica.2022.22020
  • 3. Brown, W.C. (1969) Experiments involving a microwave beam to power and position a helicopter, IEEE Transactions on Aerospace and Electronic Systems, 5(1), 692-702. doi: 10.1109/TAES.1969.309867
  • 4. Campi, T., Cruciani, S., Palandrani, E., De Santis, V., Hirata, A., and Feliziani, M. (2016) Wireless Power Transfer Charging System for AIMDs and Pacemakers, IEEE Transactions on Microwave Theory and Techniques, 64(2), 633-642. doi: 10.1109/TMTT.2015.2511011
  • 5. Dai, J., and Ludois, D. C. (2015) A survey of wireless power transfer and a critical comparison of inductive and capacitive coupling for small gap applications, IEEE Transactions on Power Electronics, 30(11), 6017-6029. doi: 10.1109/TPEL.2015.2415253
  • 6. Demirci, Y.E. (2020). Kalp Pili Uygulamaları için kablosuz enerji transferi devre tasarımı, Master thesis, Pamukkale Üniversitesi, Türkiye.
  • 7. Hong, S., Jeong, S., Lee, S., Sim, B., Kim, H., and Kim, J., (2020) Low EMF design of cochlear implant wireless power transfer system using a shielding coil”, In 2020 IEEE International Symposium on Electromagnetic Compatibility & Signal/Power Integrity (EMCSI), Reno, 623-625. doi: 10.1109/EMCSI38923.2020.9191460
  • 8. IEEE, (2019) Standard for safety levels with respect to human exposure to electric, magnetic, and electromagnetic fields (0 Hz to 100 kHz), Std, Vol. C95, no. 1™-2019,.
  • 9. Imura, T., and Hori, Y. (2011) Maximizing air gap and efficiency of magnetic resonant coupling for wireless power transfer using equivalent circuit and Neumann formula, IEEE Transactions on industrial electronics, 58(10), 4746-4752. doi: 10.1109/TIE.2011.2112317
  • 10. International Commission on Non-Ionizing Radiation Protection (ICNIRP), (2010) Guidelines for limiting exposure to time-varying electric and magnetic fields for low frequencies (1 Hz–100 kHz),” Health Physics, 99(6), 818–836. doi: 10.1097/HP.0b013e3181f06c86
  • 11. Jegadeesan, R., Nag, S., Agarwal, K., Thakor, N., and Guo, Y.-X. (2015) Enabling wireless powering and telemetry for peripheral nerve implants, IEEE J. Biomed. Health Informat., 19(3), 958–970. doi: 10.1109/JBHI.2015.2424985
  • 12. Karakaya, U., (2007) Motor control via wireless energy and information transfer”, Master thesis, İstanbul Teknik Üniversitesi, Türkiye.
  • 13. Laird, (2023). Flexible Magnetic Sheet MULL6060-300. Erişim Adresi: http:// https://www.laird.com/products/inductive-components-emc-components-and-ferrite-cores/ferrite-sheets/mull- series/mull6060-300 (Erişim Tarihi: 09.11.2023)
  • 14. Mohan, S. S., del Mar Hershenson, M., Boyd, S. P., and Lee, T. H. (1999) Simple accurate expressions for planar spiral inductances, IEEE Journal of solid-state circuits, 34(10), 1419-1424. doi: 10.1109/4.792620
  • 15. RamRakhyani, A. K., Mirabbasi, S., and Chiao, M. (2010) Design and optimization of resonance-based efficient wireless power delivery systems for biomedical implants, IEEE transactions on biomedical circuits and systems, 5(1), 48-63. doi: 10.1109/TBCAS.2010.2072782
  • 16. Sahai, A. and Graham, D. (2001) Optical wireless power transmission at long wavelengths, IEEE International Conference on Space Optical Systems and Applications (ICSOS), Santa Monica, 164- 170. doi: 10.1109/ICSOS.2011.5783662
  • 17. Sample, A. P., Meyer, D. T., and Smith, J. R. (2010) Analysis, experimental results, and range adaptation of magnetically coupled resonators for wireless power transfer, IEEE Transactions on industrial electronics, 58(2), 544-554. doi: 10.1109/TIE.2010.2046002
  • 18. Shuvo, M. M. H., Titirsha, T., Amin, N., and Islam, S. K. (2022) Energy harvesting in implantable and wearable medical devices for enduring precision healthcare. Energies, 15(20), 7495. doi: 10.3390/en15207495
  • 19. Tesla, N. (1905) Art of Transmitting Electrical Energy through Natural Mediums, U.S. Patent No. 787,412.
Toplam 19 adet kaynakça vardır.

Ayrıntılar

Birincil Dil İngilizce
Konular Elektrik Mühendisliği
Bölüm Araştırma Makaleleri
Yazarlar

Edanur Büyüktuna Bu kişi benim 0009-0005-9512-9345

Elif Dilek Bu kişi benim 0009-0002-8826-9786

Fatma Nur Yorgancılar Bu kişi benim 0009-0008-2402-9672

Ramazan Çetin Bu kişi benim 0000-0001-5846-8667

Ali Ağçal 0000-0003-0965-4038

Erken Görünüm Tarihi 28 Mart 2024
Yayımlanma Tarihi 22 Nisan 2024
Gönderilme Tarihi 22 Mart 2023
Kabul Tarihi 9 Ocak 2024
Yayımlandığı Sayı Yıl 2024 Cilt: 29 Sayı: 1

Kaynak Göster

APA Büyüktuna, E., Dilek, E., Yorgancılar, F. N., Çetin, R., vd. (2024). A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES. Uludağ Üniversitesi Mühendislik Fakültesi Dergisi, 29(1), 139-154. https://doi.org/10.17482/uumfd.1269483
AMA Büyüktuna E, Dilek E, Yorgancılar FN, Çetin R, Ağçal A. A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES. UUJFE. Nisan 2024;29(1):139-154. doi:10.17482/uumfd.1269483
Chicago Büyüktuna, Edanur, Elif Dilek, Fatma Nur Yorgancılar, Ramazan Çetin, ve Ali Ağçal. “A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES”. Uludağ Üniversitesi Mühendislik Fakültesi Dergisi 29, sy. 1 (Nisan 2024): 139-54. https://doi.org/10.17482/uumfd.1269483.
EndNote Büyüktuna E, Dilek E, Yorgancılar FN, Çetin R, Ağçal A (01 Nisan 2024) A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES. Uludağ Üniversitesi Mühendislik Fakültesi Dergisi 29 1 139–154.
IEEE E. Büyüktuna, E. Dilek, F. N. Yorgancılar, R. Çetin, ve A. Ağçal, “A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES”, UUJFE, c. 29, sy. 1, ss. 139–154, 2024, doi: 10.17482/uumfd.1269483.
ISNAD Büyüktuna, Edanur vd. “A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES”. Uludağ Üniversitesi Mühendislik Fakültesi Dergisi 29/1 (Nisan 2024), 139-154. https://doi.org/10.17482/uumfd.1269483.
JAMA Büyüktuna E, Dilek E, Yorgancılar FN, Çetin R, Ağçal A. A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES. UUJFE. 2024;29:139–154.
MLA Büyüktuna, Edanur vd. “A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES”. Uludağ Üniversitesi Mühendislik Fakültesi Dergisi, c. 29, sy. 1, 2024, ss. 139-54, doi:10.17482/uumfd.1269483.
Vancouver Büyüktuna E, Dilek E, Yorgancılar FN, Çetin R, Ağçal A. A WIRELESS POWER TRANSFER SYSTEM DESIGN FOR CHARGING OF INTRA-BODY IMPLANT DEVICES. UUJFE. 2024;29(1):139-54.

DUYURU:

30.03.2021- Nisan 2021 (26/1) sayımızdan itibaren TR-Dizin yeni kuralları gereği, dergimizde basılacak makalelerde, ilk gönderim aşamasında Telif Hakkı Formu yanısıra, Çıkar Çatışması Bildirim Formu ve Yazar Katkısı Bildirim Formu da tüm yazarlarca imzalanarak gönderilmelidir. Yayınlanacak makalelerde de makale metni içinde "Çıkar Çatışması" ve "Yazar Katkısı" bölümleri yer alacaktır. İlk gönderim aşamasında doldurulması gereken yeni formlara "Yazım Kuralları" ve "Makale Gönderim Süreci" sayfalarımızdan ulaşılabilir. (Değerlendirme süreci bu tarihten önce tamamlanıp basımı bekleyen makalelerin yanısıra değerlendirme süreci devam eden makaleler için, yazarlar tarafından ilgili formlar doldurularak sisteme yüklenmelidir).  Makale şablonları da, bu değişiklik doğrultusunda güncellenmiştir. Tüm yazarlarımıza önemle duyurulur.

Bursa Uludağ Üniversitesi, Mühendislik Fakültesi Dekanlığı, Görükle Kampüsü, Nilüfer, 16059 Bursa. Tel: (224) 294 1907, Faks: (224) 294 1903, e-posta: mmfd@uludag.edu.tr