TY - JOUR TT - Radiation Management for C-Arm Fluoroscopy Devices AU - Şahiner, Tuğba AU - Gül, Serdar Savaş AU - İlçe, Huri Tilla AU - Deveci, Emel Koçyiğit AU - Erdem, Ali Fuat AU - Kurt, Mustafa AU - Eker, Sıtkı PY - 2014 DA - June JF - Gaziosmanpaşa Üniversitesi Tıp Fakültesi Dergisi JO - Gaziosmanpaşa Tıp Dergisi PB - Tokat Gaziosmanpaşa Üniversitesi WT - DergiPark SN - 1309-3320 SP - 101 EP - 109 VL - 6 IS - 2 KW - C-kollu sopi KW - radyasyon yönetimi N2 - Purpose: In our study, C-arm fluoroscopy devices that used in the operating room have been identified radioactivity rate for during the operation and possible radiation exposure of personnel. It has been tested the strength of the radiation protective equipment. Operating room personnels have been tried to put forward awareness of the radiation. Materials and Methods: Used in the operating room 3 units C-arm fluoroscopy devices tested radiation leakage during work and employees were been determined the radiation dose quantities. For radiation leakage measurements each device is made of 100 kVp and exposure in 50mas value. Surface measuring device has been detected by the enviroment radiation dose. Robustness of the lead apron, gonads, thyroid and eye protective equipments that used radiation shield were determined under fluoroscopic displayed. Results and Conclusion: Dose measurements made outside the operating room as a result, the amount of radioactivity was not found to be significantly higher. So fluoroscopic devices to be coated with lead plates of the room where it was understood that there is no need. As a result; fluoroscopic device closest to the doctors and nurses without the protective lead apron will be exposed to the highest radiation dose UR - https://dergipark.org.tr/tr/pub/gutfd/article/378934 L1 - https://dergipark.org.tr/tr/download/article-file/404536 ER -