Background and Aims: The aim of this study was to determine the maximum dynamic torque and apically directed force generated during the shaping of simulated curved root canals using the Protaper Gold (S1, S2, F1, F2, and F3) and Protaper Ultimate (Slider, Shaper, F1, F2, and F3) systems. Materials and Methods: A specially designed, computercontrolled testing platform integrated with an endodontic motor was used to measure the maximum torque and apical force transmitted during the shaping of curved resin blocks. Resin blocks were mounted on the platform and prepared sequentially using Protaper Gold instruments (S1, S2, F1, F2, and F3 files). The same procedure was performed with Protaper Ultimate instruments. The maximum torque and force values generated during the canal shaping with each file were recorded. Group differences were evaluated using two-way ANOVA, and post-hoc Bonferroni tests. The significance level was set at p≤0.05. Results: The Protaper Ultimate Shaper file and the Protaper Gold S1 file had the highest torque values compared with other files within their respective systems (p≤0.05). Protaper Gold S1, S2, and F1 files showed higher torque values compared with their Protaper Ultimate counterparts, whereas the Protaper Ultimate F2 and F3 files had higher torque values (p≤0.05). The initial files of both systems (Slider and S1) generated the lowest forces, while the highest forces were observed with the F3 files of both systems (p≤0.05). Except for the initial shaping files, the Protaper Gold files produced higher force values across the systems. Conclusion: Protaper Ultimate system showed the lowest values for both maximum torque and force.

Keywords

References

  1. 1. Mustafa M, Attur K, Bagda KK, Singh S, Oak A, Kathiria N. An Appraisal on Newer Endodontic File Systems: A Narrative Review. J Contemp Dent Pract 2022; 23: 944-952.
  2. 2. Martins JNR, Silva E, Marques D, Ajuz N, Rito Pereira M, Pereira da Costa R et al. Characterization of the file-specific heat-treated ProTaper Ultimate rotary system. Int Endod J 2023; 56: 530-542.
  3. 3. Rowan MB, Nicholls JI, Steiner J. Torsional properties of stainless steel and nickel-titanium endodontic files. J Endod 1996; 22: 341- 345.
  4. 4. Ha JH, Cheung GS, Versluis A, Lee CJ, Kwak SW, Kim HC. ‘Screw-in’ tendency of rotary nickel-titanium files due to design geometry. Int Endod J 2015; 48: 666-672.
  5. 5. Ha JH, Park SS. Influence of glide path on the screw-in effect and torque of nickel-titanium rotary files in simulated resin root canals. Restor Dent Endod 2012; 37: 215-219.
  6. 6. Park SY, Cheung GS, Yum J, Hur B, Park JK, Kim HC. Dynamic torsional resistance of nickel-titanium rotary instruments. J Endod 2010; 36: 1200-1204.
  7. 7. Dane A, Capar ID, Arslan H, Akcay M, Uysal B. Effect of Different Torque Settings on Crack Formation in Root Dentin. J Endod 2016; 42: 304-306.
  8. 8. Aslantas EE, Ors SA, Serper A. Effect of Sodium Hypochlorite Irrigation with and Without Surfactant on Maximum Operative Torque and Vertical Force of Nickel-Titanium Endodontic Instruments. Acta Stomatol Croat 2022; 56: 33-41.

Details

Primary Language

English

Subjects

Endodontics

Journal Section

Research Article

Publication Date

August 28, 2025

Submission Date

December 20, 2024

Acceptance Date

May 12, 2025

Published in Issue

Year 2025 Volume: 49 Number: 2

APA
Aslantaş, E. E., & Serper, A. (2025). MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION. Clinical Dentistry and Research, 49(2), 103-109. https://izlik.org/JA44RA39SW
AMA
1.Aslantaş EE, Serper A. MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION. Clin Dent Res. 2025;49(2):103-109. https://izlik.org/JA44RA39SW
Chicago
Aslantaş, Eda Ezgi, and Ahmet Serper. 2025. “MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION”. Clinical Dentistry and Research 49 (2): 103-9. https://izlik.org/JA44RA39SW.
EndNote
Aslantaş EE, Serper A (August 1, 2025) MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION. Clinical Dentistry and Research 49 2 103–109.
IEEE
[1]E. E. Aslantaş and A. Serper, “MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION”, Clin Dent Res, vol. 49, no. 2, pp. 103–109, Aug. 2025, [Online]. Available: https://izlik.org/JA44RA39SW
ISNAD
Aslantaş, Eda Ezgi - Serper, Ahmet. “MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION”. Clinical Dentistry and Research 49/2 (August 1, 2025): 103-109. https://izlik.org/JA44RA39SW.
JAMA
1.Aslantaş EE, Serper A. MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION. Clin Dent Res. 2025;49:103–109.
MLA
Aslantaş, Eda Ezgi, and Ahmet Serper. “MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION”. Clinical Dentistry and Research, vol. 49, no. 2, Aug. 2025, pp. 103-9, https://izlik.org/JA44RA39SW.
Vancouver
1.Eda Ezgi Aslantaş, Ahmet Serper. MAXIMUM DYNAMIC TORQUE AND FORCE GENERATED BY PROTAPER GOLD AND PROTAPER ULTIMATE SYSTEMS DURING ROOT CANAL PREPARATION. Clin Dent Res [Internet]. 2025 Aug. 1;49(2):103-9. Available from: https://izlik.org/JA44RA39SW