A comparative study to evaluate the heat generated during osteotomy with conventional drill, trephine and alveolar expander

Abstract

Purpose : Excessively produced heat could lead to clinical failure of osseointegration. This study was done to compare the heat generated during osteotomy with the conventional drill, trephine, and alveolar expander Material and methods: This in vitro study was performed on ten bovine femoral bones. In each femoral bone, three osteotomy sites were prepared at a distance of 1.5cm using the conventional drill, trephine, and alveolar expander. During osteotomy, the site was irrigated with a copius amount of normal saline. Osteotomy sites of 3.6 mm in diameter and 11.5 mm in length were prepared using the conventional drill and bone trephines. The alveolar expander used for preparing the osteotomy site was 3.5mm, the nearest dimensions available. The temperature rise was measured using a thermocouple thermometer. Repeated measures ANOVA and Fisher’s least significant difference pairwise comparison test was done for statistical analysis. Results: Repeated measures ANOVA revealed a significant difference in the heat generation with the conventional drill, trephine, and alveolar expander (P< 0.001). The mean heat generated was maximum with the trephine (28.26±0.246) followed by the conventional drill (27.27±0.297) and least with alveolar expander (25.64±0.142). Pairwise comparison showed a significant difference in heat generated during osteotomy with conventional drill compared to trephine (P=0.023), conventional drill compared to alveolar expander (P=0.014), and trephine compared to alveolar expander (P< 0.001). Conclusion: The heat generated with trephine was maximum compared to the alveolar expander and conventional drills. If in case trephine is to be used, both internal and external irrigation must be used. Less heat generation during osteotomy by alveolar expander seems very promising and advantageous for better osseointegration.

Keywords

Supporting Institution

None

Project Number

Nil

References

  1. Reference 1. Ogle OE. Implant surface material, design, and osseointegration. Dent Clin North Am 2015;59:505-20.
  2. Reference 2. Guglielmotti MB, Olmedo DG, Cabrini RL. Research on implants and osseointegration. Periodontol 2000.2019;79:178-89.
  3. Reference 3. Möhlhenrich SC, Modabber A, Steiner T, Mitchell DA, Hölzle F. Heat generation and drill wear during dental implant site preparation: systematic review. Br J Oral Maxillofac Surg 2015;53:679-89.
  4. Reference 4.Tehemar SH. Factors affecting heat generation during implant site preparation: a review of biologic observations and future considerations. Int J Oral Maxillofac Implants 1999;14:127-36.
  5. Reference 5. Kuster M, Mukaddam K, Zitzmann NU, Filippi A, Kühl S. Influence of a Novel Drill Design on Heat Generation During Conventional and Guided Implant Osteotomy. Int J Oral Maxillofac Implants 2021;36:e31-41.
  6. Reference 6. Eriksson R, Adell R. Temperatures during drilling for the placement of implants using the osteointegration technique. J Oral Maxillofac Surg 1986;36:968-73.
  7. Reference 7.Flanagan D. Heat generated during seating of dental implant fixtures. J Oral Implantol 2014;40:174-81.
  8. Reference 8. Isler SC, Cansiz E, Tanyel C, Soluk M, Selvi F, Cebi Z. The effect of irrigation temperature on bone healing. Int J Med Sci 2011;8:704-8.

Details

Primary Language

English

Subjects

Dentistry

Journal Section

Research Article

Publication Date

January 31, 2023

Submission Date

February 11, 2022

Acceptance Date

May 25, 2022

Published in Issue

Year 2023 Volume: 57 Number: 1

APA
Omar, S., Jaiswal, H., Mishra, S., Bhargava, D., & Kumar, P. (2023). A comparative study to evaluate the heat generated during osteotomy with conventional drill, trephine and alveolar expander. European Oral Research, 57(1), 22-27. https://doi.org/10.26650/eor.20221071705
AMA
1.Omar S, Jaiswal H, Mishra S, Bhargava D, Kumar P. A comparative study to evaluate the heat generated during osteotomy with conventional drill, trephine and alveolar expander. EOR. 2023;57(1):22-27. doi:10.26650/eor.20221071705
Chicago
Omar, Shubhanshi, Honey Jaiswal, Sunil Mishra, Darpan Bhargava, and Prince Kumar. 2023. “A Comparative Study to Evaluate the Heat Generated During Osteotomy With Conventional Drill, Trephine and Alveolar Expander”. European Oral Research 57 (1): 22-27. https://doi.org/10.26650/eor.20221071705.
EndNote
Omar S, Jaiswal H, Mishra S, Bhargava D, Kumar P (January 1, 2023) A comparative study to evaluate the heat generated during osteotomy with conventional drill, trephine and alveolar expander. European Oral Research 57 1 22–27.
IEEE
[1]S. Omar, H. Jaiswal, S. Mishra, D. Bhargava, and P. Kumar, “A comparative study to evaluate the heat generated during osteotomy with conventional drill, trephine and alveolar expander”, EOR, vol. 57, no. 1, pp. 22–27, Jan. 2023, doi: 10.26650/eor.20221071705.
ISNAD
Omar, Shubhanshi - Jaiswal, Honey - Mishra, Sunil - Bhargava, Darpan - Kumar, Prince. “A Comparative Study to Evaluate the Heat Generated During Osteotomy With Conventional Drill, Trephine and Alveolar Expander”. European Oral Research 57/1 (January 1, 2023): 22-27. https://doi.org/10.26650/eor.20221071705.
JAMA
1.Omar S, Jaiswal H, Mishra S, Bhargava D, Kumar P. A comparative study to evaluate the heat generated during osteotomy with conventional drill, trephine and alveolar expander. EOR. 2023;57:22–27.
MLA
Omar, Shubhanshi, et al. “A Comparative Study to Evaluate the Heat Generated During Osteotomy With Conventional Drill, Trephine and Alveolar Expander”. European Oral Research, vol. 57, no. 1, Jan. 2023, pp. 22-27, doi:10.26650/eor.20221071705.
Vancouver
1.Shubhanshi Omar, Honey Jaiswal, Sunil Mishra, Darpan Bhargava, Prince Kumar. A comparative study to evaluate the heat generated during osteotomy with conventional drill, trephine and alveolar expander. EOR. 2023 Jan. 1;57(1):22-7. doi:10.26650/eor.20221071705