Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS

Volume: 4 Number: 1 March 1, 2020
EN

Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS

Abstract

Purpose: Polymer gel materials are used as dosimeters to provide three dimensional radiation dose distribution. An ideal gel dosimeter should offer radiological properties equivalent to water. This study aims to determine and compare the water equivalencies of various new polymer gel mixtures using Geant4-based Architecture for Medicine-Oriented Simulations GAMOS software program. Methods: To do this, percentage depth dose curves of the simulated polymer gel mixtures were compared with water curves. Different combinations of materials from 2 main classes were used in the mixtures: 1 Sodium Carboxymethyl Cellulose SODIUM CMC , Polyvinyl Alcohol PVA and Polyethylene Glycol PEG as gel matrix, 2 2-Hydroxyethyl Methacrylate HEMA , 1-Vinyl-2-Pyrrolidinone VP and Diethylene Glycol Dimethacrylate DEGDMA as monomer. Each mixture also contained ultra-pure water and an antioxidant material called Tetrakis Hydroxymethyl Phosphonium Chloride THPC . Results: Percentage depth doses of gel mixtures and water was calculated for 6, 18 MV and 50, 70, 100, 150 kVp X rays. The mixture sample showing radiological behavior nearest to water over all energies was observed to be 31a Water, PEG, HEMA, THPC , while the largest difference is observed in 11a Water, SODIUM CMC, HEMA, THPC .Conclusion: As a result, simulated gel formulations were found to be radiologically water equivalent.

Keywords

References

  1. Baldock C, De Deene Y, Doran S, et al. Polymer gel dosimetry. Phys Med Biol 2010;55:R1–R63. [CrossRef]
  2. McJury M, Oldham M, Cosgrove VP, et al. Radiation dosimetry using polymer gels: methods and applications. Br J Radiol 2000;73:919– 929. [CrossRef]
  3. De Deene Y, De Wagter C, Van Duyse B, et al. Validation of MR-based polymer gel dosimetry as a preclinical three dimensional verification tool in conformal radiotherapy. Magn Reson Med 2000;43:116–125. [CrossRef]
  4. Maryanski MJ, Schulz RJ, Ibbott GS, et al. Magnetic resonance imaging of radiation dose distributions using a polymer-gel dosimeter. Phys Med Biol 1994;39:1437–1455. [CrossRef]
  5. Brindha S, Venning AJ, Hill B, Baldock C. Experimental study of attenuation properties of normoxic polymer gel dosimeters. Phys Med Biol 2004;49:N353–N361. [CrossRef]
  6. Hill B, Venning A, Baldock C. The dose response of normoxic polymer gel dosimeters measured using X-ray CT. Br J Radiol 2005;78:623–630. [CrossRef]
  7. Zahran RR, Kandeil AY, Higazy AA, Kassem ME. Ultrasonic and thermal properties of gamma irradiated low-density polyethylene. J Appl Poly Sci 1993;49:1291–1297. [CrossRef]
  8. Zahran RR. Effects of gamma irradiation on the ultrasonic and structural properties of polyoxymethylene. Mater Lett 1998;37:83-89. [CrossRef]

Details

Primary Language

English

Subjects

-

Journal Section

-

Authors

Ayşegül Yurt This is me

İsmail Özsoykal This is me

Publication Date

March 1, 2020

Submission Date

-

Acceptance Date

-

Published in Issue

Year 2020 Volume: 4 Number: 1

APA
Özbay, T., Yurt, A., & Özsoykal, İ. (2020). Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS. Journal of Basic and Clinical Health Sciences, 4(1), 51-58. https://doi.org/10.30621/jbachs.2020.899
AMA
1.Özbay T, Yurt A, Özsoykal İ. Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS. JBACHS. 2020;4(1):51-58. doi:10.30621/jbachs.2020.899
Chicago
Özbay, Türkan, Ayşegül Yurt, and İsmail Özsoykal. 2020. “Simulation of Water Equivalency of Polymer Gel Dosimeters With GAMOS”. Journal of Basic and Clinical Health Sciences 4 (1): 51-58. https://doi.org/10.30621/jbachs.2020.899.
EndNote
Özbay T, Yurt A, Özsoykal İ (March 1, 2020) Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS. Journal of Basic and Clinical Health Sciences 4 1 51–58.
IEEE
[1]T. Özbay, A. Yurt, and İ. Özsoykal, “Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS”, JBACHS, vol. 4, no. 1, pp. 51–58, Mar. 2020, doi: 10.30621/jbachs.2020.899.
ISNAD
Özbay, Türkan - Yurt, Ayşegül - Özsoykal, İsmail. “Simulation of Water Equivalency of Polymer Gel Dosimeters With GAMOS”. Journal of Basic and Clinical Health Sciences 4/1 (March 1, 2020): 51-58. https://doi.org/10.30621/jbachs.2020.899.
JAMA
1.Özbay T, Yurt A, Özsoykal İ. Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS. JBACHS. 2020;4:51–58.
MLA
Özbay, Türkan, et al. “Simulation of Water Equivalency of Polymer Gel Dosimeters With GAMOS”. Journal of Basic and Clinical Health Sciences, vol. 4, no. 1, Mar. 2020, pp. 51-58, doi:10.30621/jbachs.2020.899.
Vancouver
1.Türkan Özbay, Ayşegül Yurt, İsmail Özsoykal. Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS. JBACHS. 2020 Mar. 1;4(1):51-8. doi:10.30621/jbachs.2020.899

Cited By