THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY

Volume: 49 Number: 1 March 25, 2015
TR EN

THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY

Abstract

Purpose: The objective of this in vitro study was to investigate and compare the precisions of several radiodiagnostic methods used in dentistry for the measurement of peri-implantary sites.
Materials and Methods: Six dental implants were placed in a human cadaver mandible. Periapical radiographs obtained with the parallel as well as the bisecting angle technique, digital and conventional panoramic radiographs were used for implant and periimplant bone measurements. The measurement results at each implant were statistically analyzed.
Results: The ICC values for the inter-observer reliability were 0.79 for implant diameters and 0.96 for implant lengths. Statistical significance was not detected between the differences of the measurements of the 2 examiners from the original implant dimensions related to anatomic locations. For both of the examiner measurements, significantly less difference from the original implant dimensions was detected in the parallel technique compared to the other techniques (p<0.05).
Conclusion: The present study showed that the most precise peri-implant bone measurements can be obtained from periapical radiographies by using the parallel technique.

Keywords

References

  1. Chaytor DV, Zarb GA, Schmitt A, Lewis DW. The longitudinal effectiveness of osseointegrated dental implants. The Toronto study: bone level changes. Int J Periodontics Restorative Dent 1991;11(2):112–125.
  2. Monsour PA, Dudhia R. Implant radiography and radiology. Aust Dent J 2008;53 (Suppl 1):11-25.
  3. Dula K, Mini R, van der Stelt PF, Buser D. The radiographic assessment of implant patients: decision-making criteria. Int J Oral Maxillofac Implants 2001;16(1):80–89.
  4. Mupparapu M, Singer SR. Implant imaging for the dentist. J Can Dent Assoc 2004; 70(1):32.
  5. Donatsky O. Osseointegrated dental implants with ball attachments supporting over-dentures in patients with mandibular alveolar ridge atrophy. Int J Oral Maxillofac Implants 1993;8(2):162–166.
  6. Boerrigter EM, van Oort RP, Raghoebar GM, Stegenga B, Schoen PJ, Boering G. A controlled clinical trial of implant-retained mandibular overdentures: clinical aspects. J Oral Rehabil 1997;24(3):182–190.
  7. Versteegh PA, Van Beek GJ, Slagter AP, Ottervanger JP. Clinical evaluation of man-dibular overdentures supported by multiple-bar fabrication: a follow-up study of two implant systems. Int J Oral Maxillofac Implants 1995;10(5):595–603.
  8. Mericske-Stern R, Steinlin Schaffner T, Marti P, Geering AH. Peri-implant mucosal aspects of ITI implants supporting overdentures. A five-year longitudinal study. Clin Oral Implants Res 1994;5(1):9–18.

Details

Primary Language

English

Subjects

-

Journal Section

-

Authors

Selda Arat Bilhan This is me

Berkman Albayrak This is me

Pelin Bozbulut This is me

Fatma Ünalan This is me

Publication Date

March 25, 2015

Submission Date

May 2, 2014

Acceptance Date

-

Published in Issue

Year 2015 Volume: 49 Number: 1

APA
Bilhan, H., Geçkili, O., Arat Bilhan, S., Ayçiçek, F., Albayrak, B., Bozbulut, P., & Ünalan, F. (2015). THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY. Journal of Istanbul University Faculty of Dentistry, 49(1), 1-9. https://doi.org/10.17096/jiufd.55134
AMA
1.Bilhan H, Geçkili O, Arat Bilhan S, et al. THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY. J Istanbul Univ Fac Dent. 2015;49(1):1-9. doi:10.17096/jiufd.55134
Chicago
Bilhan, Hakan, Onur Geçkili, Selda Arat Bilhan, et al. 2015. “THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY”. Journal of Istanbul University Faculty of Dentistry 49 (1): 1-9. https://doi.org/10.17096/jiufd.55134.
EndNote
Bilhan H, Geçkili O, Arat Bilhan S, Ayçiçek F, Albayrak B, Bozbulut P, Ünalan F (March 1, 2015) THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY. Journal of Istanbul University Faculty of Dentistry 49 1 1–9.
IEEE
[1]H. Bilhan et al., “THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY”, J Istanbul Univ Fac Dent, vol. 49, no. 1, pp. 1–9, Mar. 2015, doi: 10.17096/jiufd.55134.
ISNAD
Bilhan, Hakan - Geçkili, Onur - Arat Bilhan, Selda - Ayçiçek, Fatih - Albayrak, Berkman - Bozbulut, Pelin - Ünalan, Fatma. “THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY”. Journal of Istanbul University Faculty of Dentistry 49/1 (March 1, 2015): 1-9. https://doi.org/10.17096/jiufd.55134.
JAMA
1.Bilhan H, Geçkili O, Arat Bilhan S, Ayçiçek F, Albayrak B, Bozbulut P, Ünalan F. THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY. J Istanbul Univ Fac Dent. 2015;49:1–9.
MLA
Bilhan, Hakan, et al. “THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY”. Journal of Istanbul University Faculty of Dentistry, vol. 49, no. 1, Mar. 2015, pp. 1-9, doi:10.17096/jiufd.55134.
Vancouver
1.Hakan Bilhan, Onur Geçkili, Selda Arat Bilhan, Fatih Ayçiçek, Berkman Albayrak, Pelin Bozbulut, Fatma Ünalan. THE COMPARISON OF THE PRECISION OF DIFFERENT DENTAL RADIOGRAPHIC METHODS IN MANDIBULAR PERI-IMPLANTARY MEASUREMENTS: AN IN VITRO STUDY. J Istanbul Univ Fac Dent. 2015 Mar. 1;49(1):1-9. doi:10.17096/jiufd.55134