Araştırma Makalesi
BibTex RIS Kaynak Göster

A validated HPLC method for simultaneous estimation of Melatonin and Octyl Methoxycinnamate in combined pharmaceutical applications

Yıl 2017, Cilt: 21 Sayı: 4, 921 - 930, 01.12.2017

Öz

A simple, fast and precise reverse phase high performance
liquid chromatographic method has been developed for
the simultaneous determination of Melatonin and Octyl
Methoxycinnamate. Melatonin has become an attractive
substrate in sunscreen formulations because of its high
antioxidant and photo-protection properties. Octyl
methoxycinnamate is one of the chemical UV filter that can be
found most of the sunscreen formulations up to 7.5 % according
Food and Drug Administration. The aim of the present study
was to develop and validate a High-Performance Liquid
Chromatography method for the determination of Melatonin
and Octyl Methoxycinnamate in combined pharmaceutical or
cosmetic applications. As a model of combined pharmaceutical
applications, a microemulsion consisting of Melatonin and
Octyl Methoxycinnamate was also prepared and characterized
in terms of droplet size, pH and viscosity. The separation was
performed with a Waters XTerra RP C18 (5 μm, 4.6 x 150 mm).
All HPLC assays were performed with 10 μl injection volume,
mobile phase consisting of acetonitrile and water, using gradient
elution starting at 20% and ending at 90% of acetonitrile with
a flow rate of 1.5 ml min-1.The eluent was monitored with
UV detection at 222 nm for Melatonin and 306 nm for Octyl
Methoxycinnamate. The method was validated according
to ICH guidelines. Validation parameters were specificity,
accuracy, precision (repeatability and reproducibility), linearity,
limit of detection (LOD) and limit of quantification (LOQ).
Analytical method development results indicated that the LOD
values were 0.132 and 0.049 μg/ml; LOQ values were 0.4 and
0.15 μg/ml and assay exhibited a linear range of 0.5- 60 μg/ml
for Melatonin and Octyl Methoxycinnamate, respectively.

Kaynakça

  • 1. Slominski A, Fischer TW, Zmijewski MA, Wortsman J, Semak I, Zbytek B, Slominski RM, Tobin DJ. On the role of melatonin in skin physiology and pathology. Endocr 2005; 27: 137-48. 2. Slominski AT, Kleszczyński K, Semak I, Semak I, Janjetovic Z, Zmijewski MA, Kim TK, Slominski RM, Reiter RJ, Fischer TW. local melatoninergic system as the protector 3. Sierra AF, Ramírez MLG, Campmany ACC, Martínez AR, Naveros BC. In vivo and in vitro evaluation of the use of a newly developed melatonin loaded emulsion combined with UV filters as a protective agent against skin irradiation. J Dermatol Sci 2013; 69: 202-14. 4. Marto J, Ascenso A, Gonçalves LM, Gouveia LF, Manteigas P, Pinto P, Oliveira E, Almeida AJ, Ribeiro HM. Melatoninbased pickering emulsion for skin’s photoprotection. Drug Deliv 2016; 23: 1594-607. 5. Wang SQ, Lim HW. Current status of the sunscreen regulation in the United States: 2011 Food and Drug Administration’s final rule on labeling and effectiveness testing. J Am Acad Dermatol 2011; 65: 863-9. 6. Heuschkel S, Goebel A, Neubert R.H.H. Microemulsions— modern colloidal carrier for dermal and transdermal drug delivery. J Pharm Sci 2008; 97: 603–31. 7. Lopes LB. Overcoming the cutaneous barrier with microemulsions. Pharmaceutics 2014; 6: 52-77. 8. Carlotti ME, Gallarate M, Rossatto V. O/W microemulsion as a vehicle for sunscreens. J Cosmet Sci 2003; 54: 451–62. 9. Lin J, Zhang C, Gao Y, Zhao X, Li X. A validated HPLC method for determining melatonin in capsule dosage form. Spatula DD 2012; 2: 147-51. 10. Kedor-Hackmann ERM., De Lourdes Pérez González ML, Singh AK, Santoro MIRM. Validation of a HPLC method for simultaneous determination of five sunscreens in lotion preparation. Int J Cosmet Sci 2006; 28: 219–24. 11. Andrisano V, Bertucci C, Battaglia A, Cavrini V. Photostability of drugs: Photodegradation of melatonin and its determination in commercial formulations. J Pharm Biomed Anal 2000; 23: 15-23. 12. ICH (Q2 (R1); Validation of Analytical Procedures: Text and methodology. International Conference on Harmonization, Geneva. 2005. 13. Songkro S, Lo NL, Tanmanee N, Maneenuan D, Boonme P. In vitro release, skin permeation and retention of benzophenone-3 from microemulsions (o/w and w/o). J Drug Deliv Sci Technol 2014; 24: 703-11. 14. FDA, Reviewer Guidance—Validation of chromatographic methods. Rockville, United States of America. 1994.
Yıl 2017, Cilt: 21 Sayı: 4, 921 - 930, 01.12.2017

Öz

Kaynakça

  • 1. Slominski A, Fischer TW, Zmijewski MA, Wortsman J, Semak I, Zbytek B, Slominski RM, Tobin DJ. On the role of melatonin in skin physiology and pathology. Endocr 2005; 27: 137-48. 2. Slominski AT, Kleszczyński K, Semak I, Semak I, Janjetovic Z, Zmijewski MA, Kim TK, Slominski RM, Reiter RJ, Fischer TW. local melatoninergic system as the protector 3. Sierra AF, Ramírez MLG, Campmany ACC, Martínez AR, Naveros BC. In vivo and in vitro evaluation of the use of a newly developed melatonin loaded emulsion combined with UV filters as a protective agent against skin irradiation. J Dermatol Sci 2013; 69: 202-14. 4. Marto J, Ascenso A, Gonçalves LM, Gouveia LF, Manteigas P, Pinto P, Oliveira E, Almeida AJ, Ribeiro HM. Melatoninbased pickering emulsion for skin’s photoprotection. Drug Deliv 2016; 23: 1594-607. 5. Wang SQ, Lim HW. Current status of the sunscreen regulation in the United States: 2011 Food and Drug Administration’s final rule on labeling and effectiveness testing. J Am Acad Dermatol 2011; 65: 863-9. 6. Heuschkel S, Goebel A, Neubert R.H.H. Microemulsions— modern colloidal carrier for dermal and transdermal drug delivery. J Pharm Sci 2008; 97: 603–31. 7. Lopes LB. Overcoming the cutaneous barrier with microemulsions. Pharmaceutics 2014; 6: 52-77. 8. Carlotti ME, Gallarate M, Rossatto V. O/W microemulsion as a vehicle for sunscreens. J Cosmet Sci 2003; 54: 451–62. 9. Lin J, Zhang C, Gao Y, Zhao X, Li X. A validated HPLC method for determining melatonin in capsule dosage form. Spatula DD 2012; 2: 147-51. 10. Kedor-Hackmann ERM., De Lourdes Pérez González ML, Singh AK, Santoro MIRM. Validation of a HPLC method for simultaneous determination of five sunscreens in lotion preparation. Int J Cosmet Sci 2006; 28: 219–24. 11. Andrisano V, Bertucci C, Battaglia A, Cavrini V. Photostability of drugs: Photodegradation of melatonin and its determination in commercial formulations. J Pharm Biomed Anal 2000; 23: 15-23. 12. ICH (Q2 (R1); Validation of Analytical Procedures: Text and methodology. International Conference on Harmonization, Geneva. 2005. 13. Songkro S, Lo NL, Tanmanee N, Maneenuan D, Boonme P. In vitro release, skin permeation and retention of benzophenone-3 from microemulsions (o/w and w/o). J Drug Deliv Sci Technol 2014; 24: 703-11. 14. FDA, Reviewer Guidance—Validation of chromatographic methods. Rockville, United States of America. 1994.
Toplam 1 adet kaynakça vardır.

Ayrıntılar

Konular Sağlık Kurumları Yönetimi
Bölüm Makaleler
Yazarlar

Gülçin Arslan Azizoğlu Bu kişi benim

Erkan Azizoğlu Bu kişi benim

Sakine Tuncay Tanrıverdi

Özgen Özer Bu kişi benim

Yayımlanma Tarihi 1 Aralık 2017
Yayımlandığı Sayı Yıl 2017 Cilt: 21 Sayı: 4

Kaynak Göster

APA Arslan Azizoğlu, G., Azizoğlu, E., Tuncay Tanrıverdi, S., Özer, Ö. (2017). A validated HPLC method for simultaneous estimation of Melatonin and Octyl Methoxycinnamate in combined pharmaceutical applications. Marmara Pharmaceutical Journal, 21(4), 921-930.
AMA Arslan Azizoğlu G, Azizoğlu E, Tuncay Tanrıverdi S, Özer Ö. A validated HPLC method for simultaneous estimation of Melatonin and Octyl Methoxycinnamate in combined pharmaceutical applications. mpj. Aralık 2017;21(4):921-930.
Chicago Arslan Azizoğlu, Gülçin, Erkan Azizoğlu, Sakine Tuncay Tanrıverdi, ve Özgen Özer. “A Validated HPLC Method for Simultaneous Estimation of Melatonin and Octyl Methoxycinnamate in Combined Pharmaceutical Applications”. Marmara Pharmaceutical Journal 21, sy. 4 (Aralık 2017): 921-30.
EndNote Arslan Azizoğlu G, Azizoğlu E, Tuncay Tanrıverdi S, Özer Ö (01 Aralık 2017) A validated HPLC method for simultaneous estimation of Melatonin and Octyl Methoxycinnamate in combined pharmaceutical applications. Marmara Pharmaceutical Journal 21 4 921–930.
IEEE G. Arslan Azizoğlu, E. Azizoğlu, S. Tuncay Tanrıverdi, ve Ö. Özer, “A validated HPLC method for simultaneous estimation of Melatonin and Octyl Methoxycinnamate in combined pharmaceutical applications”, mpj, c. 21, sy. 4, ss. 921–930, 2017.
ISNAD Arslan Azizoğlu, Gülçin vd. “A Validated HPLC Method for Simultaneous Estimation of Melatonin and Octyl Methoxycinnamate in Combined Pharmaceutical Applications”. Marmara Pharmaceutical Journal 21/4 (Aralık 2017), 921-930.
JAMA Arslan Azizoğlu G, Azizoğlu E, Tuncay Tanrıverdi S, Özer Ö. A validated HPLC method for simultaneous estimation of Melatonin and Octyl Methoxycinnamate in combined pharmaceutical applications. mpj. 2017;21:921–930.
MLA Arslan Azizoğlu, Gülçin vd. “A Validated HPLC Method for Simultaneous Estimation of Melatonin and Octyl Methoxycinnamate in Combined Pharmaceutical Applications”. Marmara Pharmaceutical Journal, c. 21, sy. 4, 2017, ss. 921-30.
Vancouver Arslan Azizoğlu G, Azizoğlu E, Tuncay Tanrıverdi S, Özer Ö. A validated HPLC method for simultaneous estimation of Melatonin and Octyl Methoxycinnamate in combined pharmaceutical applications. mpj. 2017;21(4):921-30.