Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults

Volume: 40 Number: 2 July 19, 2023
EN

Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults

Abstract

Purpose: To compare the refraction values obtained with Plusoptix A09 photorefractometer and Topcon autorefractometer in adults. Materials and Methods: 230 eyes of 115 patients who had no eye disease other than refractive error were included in the study. Refraction measurements were made with Plusoptix A09 photorefractometer and Topcon KR-8100P autorefractometer devices. Measurements were taken 3 times and the averages were recorded. Median spherical, cylindrical, spherical equivalent and cylindrical axis measurements obtained with both devices were statistically compared. Interdevice compatibility was evaluated with intraclass correlation coefficient (ICC) and Spearman correlation analysis. Results: The median age of 115 patients included in the study was 37 years (range, 20-67), and 50 (43.5%) were female and 65 (56.5%) were male. Median spherical value obtained with Plusoptix A09 photorefractometer 0.50 D (range, -4.00-(3.25)), median cylindrical value -0.50 D (range, -2.00-(0.00)), spherical equivalent median 0.38 D (range, -4.00- (3.00)), the J0 power median was 0 (range, -1.00-(1.00)), and the J45 power median was 0 (range, -0.49-(0.37)). The median of the spherical value obtained by autorefractometer is 0.0 D (range, -3.75-(2.00)), the median of the cylindrical value is -0.50 D (range, -2.00-(0)), the median of the spherical equivalent is -0.25 D (range, -3.75- (1.75)), the J0 power median was 0 (range, -0.87-(0.98)), and the J45 power median was 0 (range, -0.59-(0.50)). There was a significant difference between spherical, cylindrical and spherical equivalent measurements between devices. There was no significant difference between J0 and J45 measurements. Conclusion: Besides being used in childhood, Plusoptix A09 photorefractometer can also be used as a fast and easy refraction measurement method, especially in physically or mentally handicapped adults with adjustment problems.

Keywords

References

  1. 1. Dandona R, Dandona L: Refractive error blindness. Bull World Health Organ. 2001, 79: 237-243.
  2. 2. Resnikoff S, Pascolini D, Mariotti SP, Pokharel GP. Global magnitude of visual impairment caused by uncorrected refractive errors in 2004. Bull World Health Organ. 2008; 86:63-70.
  3. 3. Rassow B, Wesemann W. Automatic infrared refractors--1985. Ophthalmology 1985; 92(8 Suppl):20-33.
  4. 4. Schimitzek T, Haase W. Efficiency of a video-autorefractometer used as a screening device for amblyogenic factors. Graefes Arch Clin Exp Ophthalmol 2002; 240:710–6
  5. 5. . Choi M, Weiss S, Schaeffel F, Seidemann A, Howland HC, Wilhelm B, et al. Laboratory, clinical, and kindergarten test of a new eccentric infrared photorefractor (PowerRefractor). Optom Vis Sci 2000; 77(10):537-48.
  6. 6. Abrahamsson M, Ohlsson J, Björndahl M, Abrahamsson H. Clinical evaluation of an eccentric infrared photorefractor: the PowerRefractor. Acta Ophthalmol Scand 2003; 81(6): 605-10.
  7. 7. Allen PM, Radhakrishnan H, O’Leary DJ. Repeatability and validity of the PowerRefractor and the Nidek AR600-A in an adult population with healthy eyes. Optom Vis Sci 2003; 80(3): 245-51.
  8. 8. Demirel S. Bilak S, Yuvacı I, Cumurcu T, Colak C. Objective measurement of refractive errors: Comparison of Plusoptix S08 with a standard autorefratometer. J Clin Exp Invest 2013; 4(1):40-6.

Details

Primary Language

English

Subjects

Health Care Administration

Journal Section

Research Article

Early Pub Date

August 1, 2023

Publication Date

July 19, 2023

Submission Date

October 4, 2022

Acceptance Date

March 20, 2023

Published in Issue

Year 2023 Volume: 40 Number: 2

APA
Duran, M., Yakar, K., & Kan, E. (2023). Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults. Deneysel Ve Klinik Tıp Dergisi, 40(2), 290-293. https://izlik.org/JA75HF84MK
AMA
1.Duran M, Yakar K, Kan E. Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults. J. Exp. Clin. Med. 2023;40(2):290-293. https://izlik.org/JA75HF84MK
Chicago
Duran, Mustafa, Konuralp Yakar, and Emrah Kan. 2023. “Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults”. Deneysel Ve Klinik Tıp Dergisi 40 (2): 290-93. https://izlik.org/JA75HF84MK.
EndNote
Duran M, Yakar K, Kan E (July 1, 2023) Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults. Deneysel ve Klinik Tıp Dergisi 40 2 290–293.
IEEE
[1]M. Duran, K. Yakar, and E. Kan, “Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults”, J. Exp. Clin. Med., vol. 40, no. 2, pp. 290–293, July 2023, [Online]. Available: https://izlik.org/JA75HF84MK
ISNAD
Duran, Mustafa - Yakar, Konuralp - Kan, Emrah. “Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults”. Deneysel ve Klinik Tıp Dergisi 40/2 (July 1, 2023): 290-293. https://izlik.org/JA75HF84MK.
JAMA
1.Duran M, Yakar K, Kan E. Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults. J. Exp. Clin. Med. 2023;40:290–293.
MLA
Duran, Mustafa, et al. “Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults”. Deneysel Ve Klinik Tıp Dergisi, vol. 40, no. 2, July 2023, pp. 290-3, https://izlik.org/JA75HF84MK.
Vancouver
1.Mustafa Duran, Konuralp Yakar, Emrah Kan. Comparison of Refractive Values Measured by Plusoptix A09 Photorefractometer and Standard Autorefractometer in Adults. J. Exp. Clin. Med. [Internet]. 2023 Jul. 1;40(2):290-3. Available from: https://izlik.org/JA75HF84MK