Research Article

Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems

Volume: 50 Number: 1 April 30, 2023
EN

Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems

Abstract

Purpose: This study aims to investigate the effect of 4 different scanning protocols offered by the VistaScan Mini Easy scanner on image quality at different exposure times. Materials & Methods: Four number size-2 photostimulable phosphor plates were exposed with 5 different exposure times while keeping other parameters constant. The exposed plates were scanned without delay using 4 different scanning protocols. 10 lp/mm, 20 lp/mm, 25 lp/mm, and 40 lp/mm are offered by the VistaScan Mini Easy scanner. The mean gray value was calculated using the ImageJ program by identifying three non-overlapping regions of interest from the background and each step in the obtained images. The mean of all mean gray values determined for the background and steps on a plate was also considered the mean gray value of that plate. Results: When plate mean gray values at 0.20 s and 0.40 s were examined, a statistically significant difference was observed between the scanning protocols (p<0.001, p=0.001 respectively). It was determined that the plate mean gray value at 40 lp/mm in 0.20 s was lower than that of other scanning protocols. The plate mean gray value at 20 lp/mm in 0.20 s was higher than that at 25 lp/mm. It was determined that the plate mean gray value at 10 lp/mm in 0.40 s was lower than that of the other groups. Conclusion: The effect of spatial resolution on diagnostics in digital imaging per se is a subject under investigation and still not agreed upon. Therefore, more studies on spatial resolution are needed.

Keywords

Thanks

The authors wish to express their sincere thanks to Ph.D. Student Aylin GÖÇOĞLU, Dokuz Eylül University Faculty of Educational Sciences, İzmir, Turkey for helping with the statistical analyses, and Assistant Professor Umut PAMUKÇU, Department of Oral and Maxillofacial Radiology, Gazi University Faculty of Dentistry, Ankara, Turkey, contribution in the study.

References

  1. 1. Farman AG, Farman TT. A comparison of 18 different x-ray detectors currently used in dentistry. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005 Apr;99(4):485-9.
  2. 2. Vandenberghe B, Jacobs R, Bosmans H. Modern dental imaging: a review of the current technology and clinical applications in dental practice. Eur Radiol. 2010 Nov;20(11):2637-55.
  3. 3. Akarslan Z. Dijital intraoral radyografinin diş hekimliği uygulamalarındaki yeri dental patolojilerde teşhis etkinliği avantaj ve dezavantajları tercih edilme durumu. Türkiye Klinikleri J Oral Maxillofac Radiol-Special Topics. 2016;2(2):29-34
  4. 4. Korner M, Weber CH, Wirth S, Pfeifer KJ, Reiser MF, Treitl MJR. Advances in digital radiography: physical principles and system overview. Radiographics. 2007 May-Jun;27(3):675-86.
  5. 5. Abubekir H. Ağız Diş ve Çene Radyolojisi, 2. Baskı. İstanbul: Nobel Tıp Kitabevleri; 2014.
  6. 6. Kurt H, Nalçacı R. İntraoral dijital görüntüleme sistemleri: Direkt sistemler, CCD, CMOS, düz panel dedektörler, indirekt sistemler, yarı direkt dijital görüntüleme, fosfor plak taramaları. Türkiye Klinikleri J Oral Maxillofac Radiol-Special Topics. 2016;2(2):p 4-9.
  7. 7. Parks ET. Digital radiographic imaging: is the dental practice ready? J Am Dent Assoc. 2008 Apr;139(4):477-81.
  8. 8. Borg E, Gröndahl HG. On the dynamic range of different X-ray photon detectors in intra-oral radiography. A comparison of image quality in film, charge-coupled device and storage phosphor systems. Dentomaxillofac Radiol. 1996 Apr;25(2):82-8.

Details

Primary Language

English

Subjects

Dentistry

Journal Section

Research Article

Early Pub Date

May 1, 2023

Publication Date

April 30, 2023

Submission Date

October 10, 2022

Acceptance Date

April 8, 2023

Published in Issue

Year 2023 Volume: 50 Number: 1

APA
Topal, C. G., Tetik, H., & Üçok, Ö. (2023). Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems. European Annals of Dental Sciences, 50(1), 28-34. https://doi.org/10.52037/eads.2023.0007
AMA
1.Topal CG, Tetik H, Üçok Ö. Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems. EADS. 2023;50(1):28-34. doi:10.52037/eads.2023.0007
Chicago
Topal, Ceyda Gizem, Hatice Tetik, and Özlem Üçok. 2023. “Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems”. European Annals of Dental Sciences 50 (1): 28-34. https://doi.org/10.52037/eads.2023.0007.
EndNote
Topal CG, Tetik H, Üçok Ö (April 1, 2023) Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems. European Annals of Dental Sciences 50 1 28–34.
IEEE
[1]C. G. Topal, H. Tetik, and Ö. Üçok, “Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems”, EADS, vol. 50, no. 1, pp. 28–34, Apr. 2023, doi: 10.52037/eads.2023.0007.
ISNAD
Topal, Ceyda Gizem - Tetik, Hatice - Üçok, Özlem. “Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems”. European Annals of Dental Sciences 50/1 (April 1, 2023): 28-34. https://doi.org/10.52037/eads.2023.0007.
JAMA
1.Topal CG, Tetik H, Üçok Ö. Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems. EADS. 2023;50:28–34.
MLA
Topal, Ceyda Gizem, et al. “Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems”. European Annals of Dental Sciences, vol. 50, no. 1, Apr. 2023, pp. 28-34, doi:10.52037/eads.2023.0007.
Vancouver
1.Ceyda Gizem Topal, Hatice Tetik, Özlem Üçok. Evaluation of the Effect of Spatial Resolution on Image Quality in Phosphor Plate Systems. EADS. 2023 Apr. 1;50(1):28-34. doi:10.52037/eads.2023.0007