Research Article

The effects of technical factors on the fractal dimension in different dental radiographic images

Volume: 57 Number: 2 May 31, 2023
EN

The effects of technical factors on the fractal dimension in different dental radiographic images

Abstract

Purpose The aim of this study was to assess the impact of exposure parameters and image formats on fractal dimension (FD) values in periapical, panoramic, and CBCT images. Materials and Methods Seven dry male mandibles were selected, and a Gutta-Percha was used to identify identical regions of interest. A periapical radiograph was taken with 60 kVp/7 mA and exported in DICOM, JPEG, TIFF, and PNG formats. Nine periapical radiographs (60, 65, 70 kVp; 4, 5, 6 mA) were taken from seven dry human mandibles. Additionally, 12 panoramic radiographs (60, 70, 81, 90 kVp; 5, 8, 13 mA) and 10 CBCT images (with different scanning options and FOVs) were taken from each mandible. FDs were measured from a standard area. Results The intra-class correlation coefficient demonstrated a high degree of agreement between observers. No significant difference was found between TIFF and PNG formats (p > 0.05). The highest FD mean was found in TIFF format, while the lowest FD mean was found in JPEG format (p < 0.001). There was no significant difference between kVp and mA settings in periapical images. In panoramic images, a significant difference was found at 90 kVp (p = 0.001) and 13 mA (p < 0.001), with lower FD values observed at these settings. There was no significant difference between FOV and resolution in CBCT images (p > 0.05). Conclusion The format of the image can influence FD. For periapical and panoramic radiographs, kVp and mA settings do not have a significant impact on FD. However, fractal analysis may not be an ideal method for evaluating three-dimensional images, such as those obtained with CBCT.

Keywords

Supporting Institution

none

Project Number

none

Thanks

Yeliz Kaşko Arıcı

References

  1. Reference1: Güngör E, Yildirim D, Çevik R. Evaluation of osteoporosis in jaw bones using cone beam CT and dual-energy X-ray absorptiometry. Journal of oral science. 2016;58(2):185-94.
  2. Reference2: Tolga Suer B, Yaman Z, Buyuksarac B. Correlation of Fractal Dimension Values with Implant Insertion Torque and Resonance Frequency Values at Implant Recipient Sites. International Journal of Oral & Maxillofacial Implants. 2016;31(1).
  3. Reference3:Geraets W, Van Der Stelt P. Fractal properties of bone. Dentomaxillofacial Radiology. 2000;29(3):144-53.
  4. Reference4: Kato CN, Barra SG, Tavares NP, Amaral TM, Brasileiro CB, Mesquita RA, et al. Use of fractal analysis in dental images: a systematic review. Dentomaxillofacial Radiology. 2020;49(2):20180457.
  5. Reference5: White SC, Rudolph DJ. Alterations of the trabecular pattern of the jaws in patients with osteoporosis. Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology. 1999;88(5):628-35.
  6. Reference6: Bechara B, McMahan CA, Moore WS, Noujeim M, Geha H, Teixeira FB. Contrast-to-noise ratio difference in small field of view cone beam computed tomography machines. Journal of oral science. 2012;54(3):227-32.
  7. Reference7: Ibrahim N, Parsa A, Hassan B, van der Stelt P, Aartman IH, Wismeijer D. The effect of scan parameters on cone beam CT trabecular bone microstructural measurements of the human mandible. Dentomaxillofacial Radiology. 2013;42(10):20130206.
  8. Reference8: Jolley L, Majumdar S, Kapila S. Technical factors in fractal analysis of periapical radiographs. Dentomaxillofacial Radiology. 2006;35(6):393-7.

Details

Primary Language

English

Subjects

Dentistry

Journal Section

Research Article

Publication Date

May 31, 2023

Submission Date

August 18, 2021

Acceptance Date

June 24, 2022

Published in Issue

Year 2023 Volume: 57 Number: 2

APA
Amuk, M., Şirin Sarıbal, G., Ersu, N., & Yılmaz, S. (2023). The effects of technical factors on the fractal dimension in different dental radiographic images. European Oral Research, 57(2), 68-74. https://doi.org/10.26650/eor.2023984422
AMA
1.Amuk M, Şirin Sarıbal G, Ersu N, Yılmaz S. The effects of technical factors on the fractal dimension in different dental radiographic images. EOR. 2023;57(2):68-74. doi:10.26650/eor.2023984422
Chicago
Amuk, Mehmet, Gamze Şirin Sarıbal, Nihal Ersu, and Serkan Yılmaz. 2023. “The Effects of Technical Factors on the Fractal Dimension in Different Dental Radiographic Images”. European Oral Research 57 (2): 68-74. https://doi.org/10.26650/eor.2023984422.
EndNote
Amuk M, Şirin Sarıbal G, Ersu N, Yılmaz S (May 1, 2023) The effects of technical factors on the fractal dimension in different dental radiographic images. European Oral Research 57 2 68–74.
IEEE
[1]M. Amuk, G. Şirin Sarıbal, N. Ersu, and S. Yılmaz, “The effects of technical factors on the fractal dimension in different dental radiographic images”, EOR, vol. 57, no. 2, pp. 68–74, May 2023, doi: 10.26650/eor.2023984422.
ISNAD
Amuk, Mehmet - Şirin Sarıbal, Gamze - Ersu, Nihal - Yılmaz, Serkan. “The Effects of Technical Factors on the Fractal Dimension in Different Dental Radiographic Images”. European Oral Research 57/2 (May 1, 2023): 68-74. https://doi.org/10.26650/eor.2023984422.
JAMA
1.Amuk M, Şirin Sarıbal G, Ersu N, Yılmaz S. The effects of technical factors on the fractal dimension in different dental radiographic images. EOR. 2023;57:68–74.
MLA
Amuk, Mehmet, et al. “The Effects of Technical Factors on the Fractal Dimension in Different Dental Radiographic Images”. European Oral Research, vol. 57, no. 2, May 2023, pp. 68-74, doi:10.26650/eor.2023984422.
Vancouver
1.Mehmet Amuk, Gamze Şirin Sarıbal, Nihal Ersu, Serkan Yılmaz. The effects of technical factors on the fractal dimension in different dental radiographic images. EOR. 2023 May 1;57(2):68-74. doi:10.26650/eor.2023984422