Conference Paper

Applicability of a Gage R&R Study on a Home Blood Glucose Meter

Volume: 23 September 30, 2023
  • Zoltan Varadı *
  • Gyozo Attila Szılagyı
EN

Applicability of a Gage R&R Study on a Home Blood Glucose Meter

Abstract

Home glucose meters have become widely accessible and, in addition to self-monitoring applications, are now commonly used to support biohacking. Accu-Chek Instant is an easy-to-use and cheaply available device, and thus, it is recommended to use in a known lifestyle change program. Participants and users do not require special knowledge, though have due or undue concerns about changes in blood glucose levels. The study aims to find the minimal magnitude of change in blood glucose level, i.e. the effective resolution of the fore-mentioned device. In addition, we are seeking applicability of the standard Gage R&R study method which is well known in the automotive industry. As Gage R&R requires to incorporate several variance factors into the study, we reproduced typical home applications and daily routines of laymen. Our results on measurement uncertainty turned to be in range with the accuracy specified by the manufacturer, and the Gage R&R method helped us to formulate answer to home user concerns on a layman’s language.

Keywords

References

  1. Automotive Industry Action Group. (2010). Measurement systems analysis. Reference manual (4th ed.). Chrysler Group LLC, Ford Motor Company, General Motors Corporation
  2. Erbach, M., Freckmann, G., Hinzmann, R., Kulzer, B., Ziegler, R., Heinemann, L., & Schnell, O. (2016). Interferences and limitations in blood glucose self-testing: An overview of the current knowledge. Journal of Diabetes Science and Technology. 10(5), 1161-1168.
  3. Freckmann, G., Schmid, C., Baumstark, A., Rutschmann, M., Haug, C., & Heinemann, L. (2015). Analytical performance requirements for systems for self-monitoring of blood glucose with focus on system accuracy. Journal of Diabetes Science and Technology. 9(4), 885–894.

Details

Primary Language

English

Subjects

Environmental and Sustainable Processes

Journal Section

Conference Paper

Authors

Zoltan Varadı * This is me
Hungary

Gyozo Attila Szılagyı This is me
Hungary

Early Pub Date

September 29, 2023

Publication Date

September 30, 2023

Submission Date

April 16, 2023

Acceptance Date

September 3, 2023

Published in Issue

Year 2023 Volume: 23

APA
Varadı, Z., & Szılagyı, G. A. (2023). Applicability of a Gage R&R Study on a Home Blood Glucose Meter. The Eurasia Proceedings of Science Technology Engineering and Mathematics, 23, 292-299. https://doi.org/10.55549/epstem.1368281