A Fully Automatic Novel Method to Determine QT Interval Based on Continuous Wavelet Transform
Year 2017,
Volume: 17 Issue: 1, 3103 - 3111, 27.03.2017
Cüneyt Yılmaz
,
Mehmet İşcan
,
Abdurrahman Yılmaz
Abstract
Nowadays, automatic recognition algorithm
is being frequently utilized to extract the information concerning cardiac
abnormalities. In this study, a fully automatic novel method based on the continuous
wavelet transform (CWT) was developed for QT intervals in various ECG signals.
Especially, the determination of T-wave end is the paramount problem to be
solved.The developed method was performed to find the beginning of QRS
complexes and the end of T-wave. The proposed algorithm was tested on
MIT-BIH-NSR database given by QT database, then, it yielded the scores 15.17
milliseconds and root-mean-square error of 17.19 milliseconds at silver
standard, 19.22 milliseconds and 20.22 milliseconds at gold standard,
respectively. In conclusion, the proposed algorithm is a fully automatic method
to attain a high performance in the calculation of QT intervals at various ECG
signals.
References
- Goldberg, R. J., Bengtson, J., Chen, Z., Anderson, K. M., Locati, E., & Levy, D.,"Duration of the QT interval and total and cardiovascular mortality in healthy persons (The Framingham Heart Study experience)".,The American journal of cardiology, 67(1), 55-58.,1991
- Zhang, Q., Manriquez, A. I., Médigue, C., Papelier, Y., & Sorine, M.,“An algorithm for robust and efficient location of T-wave ends in electrocardiograms”,IEEE Transactions on Biomedical Engineering, 53(12), 2544-2552,2006.
- Goldenberg, I. L. A. N., Moss, A. J., & Zareba, W. QT interval: how to measure it and what is" normal". Journal of cardiovascular electrophysiology, 17(3), 333-336, (2006).
- Helfenbein, E. D., Zhou, S. H., Lindauer, J. M., Field, D. Q., Gregg, R. E., Wang, J. J., ... & Michaud, F. P. An algorithm for continuous real-time QT interval monitoring. Journal of electrocardiology, 39(4), S123-S127, (2006).
- Murray, A., McLaughlin, N. B., Bourke, J. P., Doig, J. C., Furniss, S. S., & Campbell, R. W.,“Errors in manual measurement of QT intervals”,British Heart Journal, 71(4), 386-390, (1994).
- McLaughlin, N. B., Campbell, R. W. F., & Murray, A.,“Accuracy of automatic QT measurement techniques”. Computers in Cardiology, Proceedings. 863-866. IEEE, 1993.
- Hayn, D., Kollmann, A., & Schreier, G.,“Automated QT interval measurement from multilead ECG signals”,Computers in Cardiology, 381-384, IEEE, 2006.
- Clifford, G. D., & Villarroel, M. C.,“Model-based determination of QT intervals”,Computers in Cardiology, 357-360, IEEE, 2006.
- Xue, Q., & Reddy, S. Algorithms for computerized QT analysis. Journal of electrocardiology, 30, 181-186. (1998).
- Alexander A. Suarez Leon et al, "Neural Network Approach for T wave End Detection: A comparison of Architectures", Comput. Cardiol., vol. 42, pp. 589-592, 2015.
- Mehmet İşcan; Cüneyt Yilmaz; Faruk Yi̇ği̇t, " T-wave end pattern classification based on Gaussian mixture model", DOI: 10.1109/SIU.2016.7496149, Pages 1953 - 1956, 2016.
- Mehmet İşcan, Faruk Yi̇ği̇t, Cüneyt Yilmaz," Heartbeat Pattern Classification Algorithm Based on Gaussian Mixture Model", Inter. Symp. on Med. Meas. and App., Italy,2016.
- Chesnokov, Y. C., Nerukh, D., & Glen, R. C.,“Individually adaptable automatic QT detector”. Computers in Cardiology,337-340, IEEE, 2006.
- Li, C., Zheng, C., & Tai, C. Detection of ECG characteristic points using wavelet transforms. IEEE Transactions on biomedical Engineering, 42(1), 21-28., (1995).
- Chudacek, V., Huptych, M., Novak, D., & Lhotska, L.. Automatic QT interval measurement using rule-based gradient method. In Computers in Cardiology, 2006 (pp. 349-351). IEEE, (2006, September)
- Laguna P. R. G. Mark, Goldberger A., Moody G. B., “A database for evaluation of algorithms for measurement of QT and other waveform intervals in the ECG”, Comput.in Card., 1997.
- Willems, J. L., Abreu-Lima, C., Arnaud, P., van Bemmel, J. H., Brohet, C., Degani, R., ... & Machado, H. The diagnostic performance of computer programs for the interpretation of electrocardiograms. New England Journal of Medicine, 325(25), 1767-1773., (1991).
- Jané, R., Blasi, A., García, J., & Laguna, P.. Evaluation of an automatic threshold based detector of waveform limits in Holter ECG with the QT database. In Computers in Cardiology 1997 (pp. 295-298). IEEE., (1997, September)
- Cesari, M., Mehlsen, J., Mehlsen, A. B., & Sorensen, H. B. D. Application of a new robust ECG T-wave delineation algorithm for the evaluation of the autonomic innervation of the myocardium. In Engineering in Medicine and Biology Society (EMBC), 2016 IEEE 38th Annual International Conference of the (pp. 3801-3804). IEEE, (2016, August).
- Madeiro, J. P., Nicolson, W. B., Cortez, P. C., Marques, J. A., Vázquez-Seisdedos, C. R., Elangovan, N., ... & Schlindwein, F. S. New approach for T-wave peak detection and T-wave end location in 12-lead paced ECG signals based on a mathematical model. Medical engineering & physics, 35(8), 1105-1115., (2013).
- Manriquez, A. I., Zhang, Q., Medigue, C., Papelier, Y., & Sorine, M. Multi-lead T wave end detection based on statistical hypothesis testing. IFAC Proceedings Volumes, 39(18), 93-98., (2006).
- Martínez, J. P., Almeida, R., Olmos, S., Rocha, A. P., & Laguna, P. A wavelet-based ECG delineator: evaluation on standard databases. IEEE Transactions on biomedical engineering, 51(4), 570-581., (2004).
- Vila, J. A., Gang, Y., Presedo, J. M. R., Fernández-Delgado, M., Barro, S., & Malik, M. A new approach for TU complex characterization. IEEE Transactions on Biomedical Engineering, 47(6), 764-772., (2000).
Year 2017,
Volume: 17 Issue: 1, 3103 - 3111, 27.03.2017
Cüneyt Yılmaz
,
Mehmet İşcan
,
Abdurrahman Yılmaz
References
- Goldberg, R. J., Bengtson, J., Chen, Z., Anderson, K. M., Locati, E., & Levy, D.,"Duration of the QT interval and total and cardiovascular mortality in healthy persons (The Framingham Heart Study experience)".,The American journal of cardiology, 67(1), 55-58.,1991
- Zhang, Q., Manriquez, A. I., Médigue, C., Papelier, Y., & Sorine, M.,“An algorithm for robust and efficient location of T-wave ends in electrocardiograms”,IEEE Transactions on Biomedical Engineering, 53(12), 2544-2552,2006.
- Goldenberg, I. L. A. N., Moss, A. J., & Zareba, W. QT interval: how to measure it and what is" normal". Journal of cardiovascular electrophysiology, 17(3), 333-336, (2006).
- Helfenbein, E. D., Zhou, S. H., Lindauer, J. M., Field, D. Q., Gregg, R. E., Wang, J. J., ... & Michaud, F. P. An algorithm for continuous real-time QT interval monitoring. Journal of electrocardiology, 39(4), S123-S127, (2006).
- Murray, A., McLaughlin, N. B., Bourke, J. P., Doig, J. C., Furniss, S. S., & Campbell, R. W.,“Errors in manual measurement of QT intervals”,British Heart Journal, 71(4), 386-390, (1994).
- McLaughlin, N. B., Campbell, R. W. F., & Murray, A.,“Accuracy of automatic QT measurement techniques”. Computers in Cardiology, Proceedings. 863-866. IEEE, 1993.
- Hayn, D., Kollmann, A., & Schreier, G.,“Automated QT interval measurement from multilead ECG signals”,Computers in Cardiology, 381-384, IEEE, 2006.
- Clifford, G. D., & Villarroel, M. C.,“Model-based determination of QT intervals”,Computers in Cardiology, 357-360, IEEE, 2006.
- Xue, Q., & Reddy, S. Algorithms for computerized QT analysis. Journal of electrocardiology, 30, 181-186. (1998).
- Alexander A. Suarez Leon et al, "Neural Network Approach for T wave End Detection: A comparison of Architectures", Comput. Cardiol., vol. 42, pp. 589-592, 2015.
- Mehmet İşcan; Cüneyt Yilmaz; Faruk Yi̇ği̇t, " T-wave end pattern classification based on Gaussian mixture model", DOI: 10.1109/SIU.2016.7496149, Pages 1953 - 1956, 2016.
- Mehmet İşcan, Faruk Yi̇ği̇t, Cüneyt Yilmaz," Heartbeat Pattern Classification Algorithm Based on Gaussian Mixture Model", Inter. Symp. on Med. Meas. and App., Italy,2016.
- Chesnokov, Y. C., Nerukh, D., & Glen, R. C.,“Individually adaptable automatic QT detector”. Computers in Cardiology,337-340, IEEE, 2006.
- Li, C., Zheng, C., & Tai, C. Detection of ECG characteristic points using wavelet transforms. IEEE Transactions on biomedical Engineering, 42(1), 21-28., (1995).
- Chudacek, V., Huptych, M., Novak, D., & Lhotska, L.. Automatic QT interval measurement using rule-based gradient method. In Computers in Cardiology, 2006 (pp. 349-351). IEEE, (2006, September)
- Laguna P. R. G. Mark, Goldberger A., Moody G. B., “A database for evaluation of algorithms for measurement of QT and other waveform intervals in the ECG”, Comput.in Card., 1997.
- Willems, J. L., Abreu-Lima, C., Arnaud, P., van Bemmel, J. H., Brohet, C., Degani, R., ... & Machado, H. The diagnostic performance of computer programs for the interpretation of electrocardiograms. New England Journal of Medicine, 325(25), 1767-1773., (1991).
- Jané, R., Blasi, A., García, J., & Laguna, P.. Evaluation of an automatic threshold based detector of waveform limits in Holter ECG with the QT database. In Computers in Cardiology 1997 (pp. 295-298). IEEE., (1997, September)
- Cesari, M., Mehlsen, J., Mehlsen, A. B., & Sorensen, H. B. D. Application of a new robust ECG T-wave delineation algorithm for the evaluation of the autonomic innervation of the myocardium. In Engineering in Medicine and Biology Society (EMBC), 2016 IEEE 38th Annual International Conference of the (pp. 3801-3804). IEEE, (2016, August).
- Madeiro, J. P., Nicolson, W. B., Cortez, P. C., Marques, J. A., Vázquez-Seisdedos, C. R., Elangovan, N., ... & Schlindwein, F. S. New approach for T-wave peak detection and T-wave end location in 12-lead paced ECG signals based on a mathematical model. Medical engineering & physics, 35(8), 1105-1115., (2013).
- Manriquez, A. I., Zhang, Q., Medigue, C., Papelier, Y., & Sorine, M. Multi-lead T wave end detection based on statistical hypothesis testing. IFAC Proceedings Volumes, 39(18), 93-98., (2006).
- Martínez, J. P., Almeida, R., Olmos, S., Rocha, A. P., & Laguna, P. A wavelet-based ECG delineator: evaluation on standard databases. IEEE Transactions on biomedical engineering, 51(4), 570-581., (2004).
- Vila, J. A., Gang, Y., Presedo, J. M. R., Fernández-Delgado, M., Barro, S., & Malik, M. A new approach for TU complex characterization. IEEE Transactions on Biomedical Engineering, 47(6), 764-772., (2000).