In this study, a sensitive, fast, economical, and practical voltammetric method was developed for analysing molnupiravir (MLP), an antiviral drug extensively used during the COVID-19 pandemic, without the need for modifying the glassy carbon (GC) electrode surface. During the electroanalysis, an oxidation peak for MLP was detected at 0.72555 V on the GC electrode. The method exhibited linearity across a concentration range of 1 - 250 µM for MLP with 0.33 µM of detection limit. Furthermore, with a 99.81% recovery from its pharmaceutical form, the method's effectiveness was validated. These results strongly indicate that this method is suitable for routine quality control of molnupiravir and may serve as a model for analysing other antiviral drugs.
Primary Language | English |
---|---|
Subjects | Pharmaceutical Analytical Chemistry |
Journal Section | Articles |
Authors | |
Publication Date | July 5, 2025 |
Submission Date | August 26, 2024 |
Acceptance Date | September 27, 2024 |
Published in Issue | Year 2025 Volume: 29 Issue: 4 |