Ion implanted - silicon detectors are widely used in the alpha radiation, especially for quantitative analysis in environmental applications. For quantitative measurements of the alpha radiation, their characterization and electronic calibration is required. We have tested the silicon detector and CSA (Charge sensitive amplifier) at the room temperature with test setup, using a Am-241 source with a characteristic energy of 5.485 MeV. The purpose of these experiments was to calibrate and characterize the response of the radon system to charged particle of the same types and energies as those encountered in soil. To meet these requirements, discriminating with pulse width of 400 ns is adjusted. The experimental count results demonstrated that this system can be used in online radon measurements.
İyon implantlı silikon dedektörler özellikle çevresel uygulamalardaki nicel analizler için alfa radyasyonu tespitinde kullanılmaktadırlar. Alfa radyasyonunun nicel ölçümleri için, bu tip dedektörlerin karakterizasyonu ve kalibrasyonu mutlaka gerekmektedir. Bu çalışmada, oda sıcaklığında 5.485 MeV'luk karakteristik enerjiye sahip olan Am-241 kaynağı kullanarak silikon dedektör ve CSA (yüke duyarlı amplifikatör) test edilmiştir. Bu deneylerin amacı ise toprak içinde karşılaşılan radyasyon enerjileri ve aynı tipteki parçacıklara karşı dedektörün tepkisini ortaya koymak ve dedektörün kalibrasyonu olmuştur. Bunu sağlamak için ise 400 ns'lik puls genişliği diskriminasyonu ayarlanmıştır. Deneysel sayım sonuçları bu sistemin gerçek zamanlı radon ölçümlerinde kullanılabileceğini göstermiştir.