Research Article
BibTex RIS Cite

Comparison of sterilization times in membrane filter devices with microbial suspensions at different concentrations.

Year 2017, Volume: 2 Issue: 1(3) - Volume 2, Issue 1(3), 328 - 334, 01.06.2017

Abstract

Introduction. Sterilization
of the membrane filter device used in water microbiology analysis is very
important. Aim of the study. To
compare the efficacy of sterilization by burning with alcohol and flaming with
a Bunsen burner. Material and methods. Bacterial
suspensions were prepared at different concentrations and for each concentration,
each analyst studied 10 culture media for each time periods. The contaminated
membrane filter funnels with the prepared samples were sterilized by flaming
for 10, 20, 40 and 60 seconds. Then, the membrane filter was placed and 100 ml
of sterile distilled water was filtered. Membrane filter was placed on
appropriate growth medium and it was incubated. The same procedure was also
carried out by burning with alcohol. Results.
At 102 concentration, bacterial growth was not observed. At 103
concentration, bacterial growth was observed 6,66% of growth medium at 10
seconds. At 104 concentration, bacterial growth was observed 30% of
growth medium at 10 seconds and 23,3% of growth medium at 20 seconds. At 105,
106 and 107 concentrations, bacterial growth was observed
in different percentages at all the time periods. Bacterial growth was not
observed in sterilization using alcohol burning technique at all the time
periods. Conclusions. Effective
sterilization can be achieved by flame with Bunsen burner with low microbial
concentration suspensions and low time periods. At high concentration
suspensions, membrane filter device can be effectively sterilized by burning
with alcohol.



Key words: sterilization,
sanitation, E.coli, Pseudomonas
aeruginosa,
membrane filter device, water quality.

References

  • NO*, Büşra Sarım*, Burcu Aslaner*, Fatih Sultan Kazdal*, Ayfer Kına*, Necla Erdi*, Gülcan Turan*, Gani Orhan* *Istanbul Public Health Laboratory Number 3, Istanbul, Turkey.
Year 2017, Volume: 2 Issue: 1(3) - Volume 2, Issue 1(3), 328 - 334, 01.06.2017

Abstract

References

  • NO*, Büşra Sarım*, Burcu Aslaner*, Fatih Sultan Kazdal*, Ayfer Kına*, Necla Erdi*, Gülcan Turan*, Gani Orhan* *Istanbul Public Health Laboratory Number 3, Istanbul, Turkey.
There are 1 citations in total.

Details

Journal Section Articles
Authors

Özge Gündoğdu Güngör This is me

Burcu Aslaner This is me

Fatih Sultan Kazdal This is me

Ayfer Kına This is me

Necla Erdi This is me

Gülcan Turan This is me

Gani Orhan This is me

BÜŞRA Sarım This is me

Publication Date June 1, 2017
Published in Issue Year 2017 Volume: 2 Issue: 1(3) - Volume 2, Issue 1(3)

Cite

APA Gündoğdu Güngör, Ö., Aslaner, B., Kazdal, F. S., Kına, A., et al. (2017). Comparison of sterilization times in membrane filter devices with microbial suspensions at different concentrations. The Turkish Journal Of Occupational / Environmental Medicine and Safety, 2(1(3), 328-334.
AMA Gündoğdu Güngör Ö, Aslaner B, Kazdal FS, Kına A, Erdi N, Turan G, Orhan G, Sarım B. Comparison of sterilization times in membrane filter devices with microbial suspensions at different concentrations. turjoem. June 2017;2(1(3):328-334.
Chicago Gündoğdu Güngör, Özge, Burcu Aslaner, Fatih Sultan Kazdal, Ayfer Kına, Necla Erdi, Gülcan Turan, Gani Orhan, and BÜŞRA Sarım. “Comparison of Sterilization Times in Membrane Filter Devices With Microbial Suspensions at Different Concentrations”. The Turkish Journal Of Occupational / Environmental Medicine and Safety 2, no. 1(3) (June 2017): 328-34.
EndNote Gündoğdu Güngör Ö, Aslaner B, Kazdal FS, Kına A, Erdi N, Turan G, Orhan G, Sarım B (June 1, 2017) Comparison of sterilization times in membrane filter devices with microbial suspensions at different concentrations. The Turkish Journal Of Occupational / Environmental Medicine and Safety 2 1(3) 328–334.
IEEE Ö. Gündoğdu Güngör, B. Aslaner, F. S. Kazdal, A. Kına, N. Erdi, G. Turan, G. Orhan, and B. Sarım, “Comparison of sterilization times in membrane filter devices with microbial suspensions at different concentrations”., turjoem, vol. 2, no. 1(3), pp. 328–334, 2017.
ISNAD Gündoğdu Güngör, Özge et al. “Comparison of Sterilization Times in Membrane Filter Devices With Microbial Suspensions at Different Concentrations”. The Turkish Journal Of Occupational / Environmental Medicine and Safety 2/1(3) (June 2017), 328-334.
JAMA Gündoğdu Güngör Ö, Aslaner B, Kazdal FS, Kına A, Erdi N, Turan G, Orhan G, Sarım B. Comparison of sterilization times in membrane filter devices with microbial suspensions at different concentrations. turjoem. 2017;2:328–334.
MLA Gündoğdu Güngör, Özge et al. “Comparison of Sterilization Times in Membrane Filter Devices With Microbial Suspensions at Different Concentrations”. The Turkish Journal Of Occupational / Environmental Medicine and Safety, vol. 2, no. 1(3), 2017, pp. 328-34.
Vancouver Gündoğdu Güngör Ö, Aslaner B, Kazdal FS, Kına A, Erdi N, Turan G, Orhan G, Sarım B. Comparison of sterilization times in membrane filter devices with microbial suspensions at different concentrations. turjoem. 2017;2(1(3):328-34.