Optimization of RPLC Conditions for Quantitative Analysis of Atorvastatin and Rosuvastatin in Pharmaceutical Dosage Form
Abstract
In this study, a simple reverse phase liquid chromatographic (RPLC) method has been developed and subsequently validated for simultaneous determination of atorvastatin (ATV) and
rosuvastatin (RSV). The column used was X Terra C18, (250 mm x 4.6 mm I.D., 5µm) with flow rate of 1 mL min-1 using photodiode array detection at 244 nm. The described method was linear over a concentration range of 3-13 µg mL-1 and 4-14 µg mL-1 for the assay of ATV and RSV respectively. Losartan was used as internal standard (IS) in the experiment. The limit of detection (LOD) values for ATV and RSV were found to be 0.133 and 0.221 µg mL-1 respectively. Limit of quantification (LOQ) for ATV and RSV were found to be 0.473 and 0.670 µg mL-1
, respectively. The results of the study showed that the proposed RPLC method is useful for the routine determination of ATV and RSV and in their pharmaceutical dosage form.
Keywords
References
- [1] Lennernas, H., Foger, G.; Pharmacodynamics and pharmacokinetics of the HMG-CoA reductase inhibitors; Clinical Pharmacokinetics, (1997); 32: 403-425.
- [2] Sultana, N., Arayne, M.S., Shahzad, W.; Simultaneous determination of ceftriaxone sodium and statin drugs in pharmaceutical formulations and human serum by rp-hplc; Journal of the Chilean Chemical Society, (2010); 55; 193-198.
- [3]Chen, X., Xiong, F., Zheng, C., Li, J., Chen, F.; Synthetic studies on statins Part 3: A facile synthesis of rosuvastatin calcium through catalytic enantioselective allylation strategy; Tetrahedron, (2014); 70: 5794-5799.
- [4] Atorvastatin: Lipitor, www.medicinenet.com (accessed May 20, 2015).
- [5] Talay, A., Demiralay, E.Ç., Daldal, Y.D., Üstün, Z.; Investigation of thermodynamic acidity constants of some statins with RPLC method; Journal of Molecular Liquids, (2015); 208: 286-290.
- [6] Shah, Y., Iqbal, Z., Ahmad, L., Khan, A., Khan, M.I., Nazir, S., Nasir, F.; Simultaneous determination of rosuvastatin and atorvastatin in human serum using RP-HPLC/UV detection: Method development, validation and optimization of various experimental parameters; Journal of Chromatography B, (2011); 879: 557-563.
- [7] Venkata Ramana, G., Pavani, G., Supriya, P., Madhavi Latha, N.; Reversed Phase-High Performance liquid chromatography method development and validation of atorvastatin in bulk drug and formulation; Asian Journal of Pharmaceutical and Clinical Research, (2015); 8: 84-87.
- [8] Donthula, S., Kumar, M.K., Teja, G.S., Kumar, Y.M., Krishna, J.Y., Ramesh, D.; A new validated RP-HPLC method for determination of rosuvastatin calcium in bulk and pharmaceutical dosage form; Der Pharmacia Lettre, (2011); 3: 350-356.
Details
Primary Language
English
Subjects
Structural Biology, Chemical Engineering
Journal Section
Research Article
Publication Date
May 24, 2016
Submission Date
May 24, 2016
Acceptance Date
-
Published in Issue
Year 2016 Volume: 11 Number: 1