Research Article

The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device within the Scope of TS EN 50160

Volume: 12 Number: 2 June 30, 2020
Emre Metin , Özgür Karagülle , Kardelen Kamışlı , Ertuğrul Çam *
EN

The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device within the Scope of TS EN 50160

Abstract

Today, the increase in demand for electrical energy has revealed a more reliable and higher quality energy concept. This subject has taken its place in electrical engineering as "Power Quality" or "Electric Energy Quality" and has made a point for its importance with the studies on it. Especially in the process of working with devices containing electronic circuits with direct current and using the alternating current offered from the network by using rectifiers, the devices may fail or become disabled due to self-protection circuits in case of power quality problems arising from the mains or the rectifiers of the device. In this article, a detailed analysis is carried out considering the problem may be due to the power quality due to the fact that the MRI device in Kırıkkale University Faculty of Medicine is in continuous failure mode. In addition, the analyzes made in accordance with TS EN 50160 standard within the scope of the research were analyzed in computer environment and the results were shared in our article. In this article, the relationship between power quality problems and grounding is also examined.

Keywords

energy analysis, , en50160, power qualty

References

  1. Anderson, P.M., FOUAD, A.A., (1994), Power System Control and Stability, Revised Printing, IEEE Press, 1994.
  2. Arrillaga, J., Bradley, D.A. ve Bodger, P.S., (1985), Power System Harmonics, John Wiley & Sons, Norwich.
  3. Arrillaga, J. ve Arnold, C.P., (1990), Computer Analysis of Power Systems, John Wiley & Sons, Essex.
  4. Arrillaga, J., Smith, B.C., Watson, N.R. ve Wood, A.R., (1997), Power System Harmonic Analysis, John Wiley & Sons.
  5. BERG G.J., “Assestment of critical voltage and load margins in VAR compensated power transmission systems”, electric power system research, 12,pp.63-69,1978
  6. Cenelec, E. N. (2014). 50160: Voltage characteristics of electricity supplied by public distribution systems. Cenelec, Brüssel.
  7. Chapman, D. 2001. Güç Kalitesi Uygulama Kılavuzu, Copper Development Association, Sarkuysan Elektrolitik Bakır Sanayi ve Ticaret A.Ş. , İstanbul.
  8. Domijan, A., & Embriz-Santander, E. (1992). A novel electric power laboratory for power quality and energy studies: training aspects. IEEE transactions on power systems, 7(4), 1571-1578.
  9. Dommel, H. W. (1969). Digital computer solution of electromagnetic transients in single-and multiphase networks. IEEE transactions on power apparatus and systems, (4), 388-399.
  10. Dugan, R.C., McGranaghan, M.F., Beaty, H.W., “Electrical Power Systems Quality”, McGraw-Hill, 1996.
APA
Metin, E., Karagülle, Ö., Kamışlı, K., & Çam, E. (2020). The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device within the Scope of TS EN 50160. International Journal of Engineering Research and Development, 12(2), 700-710. https://doi.org/10.29137/umagd.747710
AMA
1.Metin E, Karagülle Ö, Kamışlı K, Çam E. The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device within the Scope of TS EN 50160. IJERAD. 2020;12(2):700-710. doi:10.29137/umagd.747710
Chicago
Metin, Emre, Özgür Karagülle, Kardelen Kamışlı, and Ertuğrul Çam. 2020. “The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device Within the Scope of TS EN 50160”. International Journal of Engineering Research and Development 12 (2): 700-710. https://doi.org/10.29137/umagd.747710.
EndNote
Metin E, Karagülle Ö, Kamışlı K, Çam E (June 1, 2020) The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device within the Scope of TS EN 50160. International Journal of Engineering Research and Development 12 2 700–710.
IEEE
[1]E. Metin, Ö. Karagülle, K. Kamışlı, and E. Çam, “The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device within the Scope of TS EN 50160”, IJERAD, vol. 12, no. 2, pp. 700–710, June 2020, doi: 10.29137/umagd.747710.
ISNAD
Metin, Emre - Karagülle, Özgür - Kamışlı, Kardelen - Çam, Ertuğrul. “The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device Within the Scope of TS EN 50160”. International Journal of Engineering Research and Development 12/2 (June 1, 2020): 700-710. https://doi.org/10.29137/umagd.747710.
JAMA
1.Metin E, Karagülle Ö, Kamışlı K, Çam E. The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device within the Scope of TS EN 50160. IJERAD. 2020;12:700–710.
MLA
Metin, Emre, et al. “The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device Within the Scope of TS EN 50160”. International Journal of Engineering Research and Development, vol. 12, no. 2, June 2020, pp. 700-1, doi:10.29137/umagd.747710.
Vancouver
1.Emre Metin, Özgür Karagülle, Kardelen Kamışlı, Ertuğrul Çam. The Importance of Energy Quality in Medical Devices and Evaluation of Measurements Made in Kırıkkale University Medical Faculty MRI Device within the Scope of TS EN 50160. IJERAD. 2020 Jun. 1;12(2):700-1. doi:10.29137/umagd.747710