Research Article

Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation

Volume: 48 Number: 2 April 21, 2014
  • Umit Ozgur Guler
  • Alihan Derincek
  • Murat Ali Hersekli
  • Metin Ozalay
  • Bekir Murat Cinar
  • Emre Acaroglu
TR EN

Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation

Abstract

Objective: The aim of the present study was to compare calcium sulfate (CAS) and polymethylmethacrylate (PMMA) bone cements used for the augmentation of a failed pedicle screw with biomechanical pull-out strength (POS) testing.

Methods: Thirty lumbar vertebrae were harvested from 6 calves and bone mineral densities (BMD) were measured. Primary polyaxial pedicle screws were randomly inserted and pulled out and the POSs of the specimen were recorded. For revision, specimens were randomly assigned to the CAS-augmented pedicle screws group (Group 1) or PMMA-augmented pedicle screw group (Group 2). Pull-out tests were repeated to compare both groups.

Results: Mean BMD of the specimens was 1.006±0.116 g/cm2. There were no statistically significant differences between BMD results of the two groups (p=0.116). For Group 1, mean POS of primary screws was 2,441.3±936.4 N and was 2,499.5±1,425.1 N after CAS augmentation, demonstrating no statistically significant difference (p=0.865). In Group 2, mean POS of the primary screws was 2,876.6±926.6 N and significantly increased to 3,745.5±1,299.2 N after PMMA augmentation (p=0.047). There was also a significant difference in mean POS between the CAS and PMMA groups (p=0.026).

Conclusion: Although CAS augmentation facilitates a revision screw POS as strong as that of primary screws, it is not as strong as PMMA augmentation.

Keywords

References

  1. Lonstein JE, Denis F, Perra JH, Pinto MR, Smith MD, Winter RB. Complications associated with pedicle screws. J Bone Joint Surg Am 1999;81:1519-28.
  2. Chang MC, Liu CL, Chen TH. Polymethylmethacrylate augmentation of pedicle screw for osteoporotic spinal sur- gery: a novel technique. Spine 2008;33:E317-24. CrossRef
  3. Lotz JC, Hu SS, Chiu DF, Yu M, Colliou O, Poser RD. Carbonated apatite cement augmentation of pedicle screw fixation in the lumbar spine. Spine 1997;22:2716-23. CrossRef
  4. Moore DC, Maitra RS, Farjo LA, Graziano GP, Goldstein SA. Restoration of pedicle screw fixation with an in situ setting calcium phosphate cement. Spine 1997;22:1696- 705. CrossRef
  5. Motzkin NE, Chao EY, An KN, Wikenheiser MA, Lewallen DG. Pull-out strength of screws from polymeth- ylmethacrylate cement. J Bone Joint Surg Br 1994;76:320- 3.
  6. Rohmiller MT, Schwalm D, Glattes RC, Elalayli TG, Spengler DM. Evaluation of calcium sulfate paste for aug- mentation of lumbar pedicle screw pullout strength. Spine J 2002;2:255-60. CrossRef
  7. Spivak JM, Neuwirth MG, Labiak JJ, Kummer FJ, Ricci JL. Hydroxyapatite enhancement of posterior spinal in- strumentation fixation. Spine 1994;19:955-64. CrossRef
  8. Wilkes RA, Mackinnon JG, Thomas WG. Neurological deterioration after cement injection into a vertebral body. J Bone Joint Surg Br 1994;76:155.

Details

Primary Language

English

Subjects

Health Care Administration

Journal Section

Research Article

Authors

Umit Ozgur Guler This is me

Alihan Derincek This is me

Murat Ali Hersekli This is me

Metin Ozalay This is me

Bekir Murat Cinar This is me

Emre Acaroglu This is me

Publication Date

April 21, 2014

Submission Date

April 21, 2014

Acceptance Date

-

Published in Issue

Year 2014 Volume: 48 Number: 2

APA
Guler, U. O., Derincek, A., Hersekli, M. A., Ozalay, M., Murat Cinar, B., & Acaroglu, E. (2014). Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation. Acta Orthopaedica et Traumatologica Turcica, 48(2), 202-206. https://doi.org/10.3944/AOTT.2014.3193
AMA
1.Guler UO, Derincek A, Hersekli MA, Ozalay M, Murat Cinar B, Acaroglu E. Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation. Acta Orthopaedica et Traumatologica Turcica. 2014;48(2):202-206. doi:10.3944/AOTT.2014.3193
Chicago
Guler, Umit Ozgur, Alihan Derincek, Murat Ali Hersekli, Metin Ozalay, Bekir Murat Cinar, and Emre Acaroglu. 2014. “Restoration of Pull-Out Strength of the Failed Pedicle Screw: Biomechanical Comparison of Calcium Sulfate Vs Polymethylmethacrylate Augmentation”. Acta Orthopaedica et Traumatologica Turcica 48 (2): 202-6. https://doi.org/10.3944/AOTT.2014.3193.
EndNote
Guler UO, Derincek A, Hersekli MA, Ozalay M, Murat Cinar B, Acaroglu E (April 1, 2014) Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation. Acta Orthopaedica et Traumatologica Turcica 48 2 202–206.
IEEE
[1]U. O. Guler, A. Derincek, M. A. Hersekli, M. Ozalay, B. Murat Cinar, and E. Acaroglu, “Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation”, Acta Orthopaedica et Traumatologica Turcica, vol. 48, no. 2, pp. 202–206, Apr. 2014, doi: 10.3944/AOTT.2014.3193.
ISNAD
Guler, Umit Ozgur - Derincek, Alihan - Hersekli, Murat Ali - Ozalay, Metin - Murat Cinar, Bekir - Acaroglu, Emre. “Restoration of Pull-Out Strength of the Failed Pedicle Screw: Biomechanical Comparison of Calcium Sulfate Vs Polymethylmethacrylate Augmentation”. Acta Orthopaedica et Traumatologica Turcica 48/2 (April 1, 2014): 202-206. https://doi.org/10.3944/AOTT.2014.3193.
JAMA
1.Guler UO, Derincek A, Hersekli MA, Ozalay M, Murat Cinar B, Acaroglu E. Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation. Acta Orthopaedica et Traumatologica Turcica. 2014;48:202–206.
MLA
Guler, Umit Ozgur, et al. “Restoration of Pull-Out Strength of the Failed Pedicle Screw: Biomechanical Comparison of Calcium Sulfate Vs Polymethylmethacrylate Augmentation”. Acta Orthopaedica et Traumatologica Turcica, vol. 48, no. 2, Apr. 2014, pp. 202-6, doi:10.3944/AOTT.2014.3193.
Vancouver
1.Umit Ozgur Guler, Alihan Derincek, Murat Ali Hersekli, Metin Ozalay, Bekir Murat Cinar, Emre Acaroglu. Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation. Acta Orthopaedica et Traumatologica Turcica. 2014 Apr. 1;48(2):202-6. doi:10.3944/AOTT.2014.3193