A concept of the integration of the interpolative and extrapolative
calibration methods - commonly used in analytical chemistry - is presented,
i.e. it is proposed to perform both methods according to a single calibration
procedure. Such an approach allows one to obtain two estimations of the
analyte concentration in a sample and to verify it in terms of accuracy. It is
suggested to execute the integrated calibration method by the flow injection
technique in accordance with two different procedures. The instrumental
flow systems designed for this purpose are shown. The principle of both
procedural versions is revealed and the specificity of them is discussed. The
reasons are explained why the method is worth to be brought into the
analytical practice.
Other ID | JA22AF26ED |
---|---|
Journal Section | Article |
Authors | |
Publication Date | June 1, 2000 |
Submission Date | June 1, 2000 |
Published in Issue | Year 2000 Volume: 2 Issue: 2 |