Research Article

Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates

Volume: 49 Number: 3 December 31, 2022
EN

Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates

Abstract

Purpose This study aims to compare the effect of light exposure in red, green and blue (RGB) colors prior to scanning of the PSP plates. Materials & Methods An Arduino-based system is produced for standardized light exposure to the irradiated PSP plates. The system consisted of an Arduino Mega 2560 developer board, 2 RGB LED light sources, a TSL2591 digital light sensor and a DHT11 temperature and humidity sensor. A light-tight platform is produced with additive manufacturing and electronical units are integrated into this platform. A two-step alloy was used to create contrast. PSP system (VistaScan, Dürr Dental, Germany) is irradiated with fixed parameters of 70 kV, 8 mA and 0.5 seconds. Scanning of the PSPs were delayed for 1-,3-,5-, and 10-minutes, and half of the active surfaces are exposed to RGB lights independently in full brightness (PWM) and calibrated with lux while the rest is protected. MGVs are measured in 6 regions per image. The MGV differences in regions between conditions were examined by Kruskal-Wallis test. A p-value<0.05 was considered statistically significant. Results In PWM setup, signal loss was higher in blue color at 1-,3-, and 5-minutes, but at 10-minutes, only the green light produced half-image. In LUX setup, signal loss was lower in green light. Contrast loss was lower in green light with LUX calibration. (p<0.05) Conclusion Among three colors compared, effect of exposure to green light prior to scanning of the irradiated PSP plates is found to be lower than the other colors.

Keywords

Thanks

The 3D manufactured platform is designed, produced, and combined in Cerrahpaşa Research, Simulation and Design Center (CAST).

References

  1. White SC, Michael PJ, editors. Oral Radiology. 8th ed. United States, Mosby; 2018.
  2. Brennan J. An introduction to digital radiography in dentistry. J Orthod. 2002;29:66–69.
  3. Noorsaeed AS, Almohammedsaleh AH, Alhayek MM, Alnajar AA, Kariri ON, Albakri AIM, et al. Overview on Updates on Digital Dental Radiography. J Pharm Res Int. 2021;33(59B):23-28
  4. Hassan AS, Bhateja S, Arora G, Prathyusha F. Digital Radiography. IJMI. 2020;6(2):33-36.
  5. Kaur J, Parganiha Y, Dubey V, Singh D. A review report on medical imaging phosphors. Res Chem Intermed. 2014;40(8):2837-2858.
  6. Parks ET, Williamson GF. Digital radiography: an overview. J Contemp Dent Pract. 2002;3(4):23-39.
  7. Marinho-Vieira LE, Martins LAC, Freitas DQ, Haiter-Neto F, Oliveira ML. Revisiting dynamic range and image enhancement ability of contemporary digital radiographic systems. Dentomaxillofac Radiol. 2002;51(4):20210404.
  8. Kondaveeti KH, Kumaravelu NK, Vanambathina SD, Mathe SE, Vappangi S. A systematic literature review on prototyping with Arduino: Applications, challenges, advantages, and limitations. Comput Sci Rev. 2021;40:100364.

Details

Primary Language

English

Subjects

Dentistry

Journal Section

Research Article

Publication Date

December 31, 2022

Submission Date

October 16, 2022

Acceptance Date

November 10, 2022

Published in Issue

Year 2022 Volume: 49 Number: 3

APA
Amasya, H., Orhan, K., & Alkan, A. (2022). Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates. European Annals of Dental Sciences, 49(3), 131-138. https://doi.org/10.52037/eads.2022.0040
AMA
1.Amasya H, Orhan K, Alkan A. Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates. EADS. 2022;49(3):131-138. doi:10.52037/eads.2022.0040
Chicago
Amasya, Hakan, Kaan Orhan, and Afra Alkan. 2022. “Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates”. European Annals of Dental Sciences 49 (3): 131-38. https://doi.org/10.52037/eads.2022.0040.
EndNote
Amasya H, Orhan K, Alkan A (December 1, 2022) Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates. European Annals of Dental Sciences 49 3 131–138.
IEEE
[1]H. Amasya, K. Orhan, and A. Alkan, “Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates”, EADS, vol. 49, no. 3, pp. 131–138, Dec. 2022, doi: 10.52037/eads.2022.0040.
ISNAD
Amasya, Hakan - Orhan, Kaan - Alkan, Afra. “Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates”. European Annals of Dental Sciences 49/3 (December 1, 2022): 131-138. https://doi.org/10.52037/eads.2022.0040.
JAMA
1.Amasya H, Orhan K, Alkan A. Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates. EADS. 2022;49:131–138.
MLA
Amasya, Hakan, et al. “Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates”. European Annals of Dental Sciences, vol. 49, no. 3, Dec. 2022, pp. 131-8, doi:10.52037/eads.2022.0040.
Vancouver
1.Hakan Amasya, Kaan Orhan, Afra Alkan. Comparison of the Effect of the Time Under the Three Primary Color Lighting of Led Production Before Scanning of Phosphorus Plates. EADS. 2022 Dec. 1;49(3):131-8. doi:10.52037/eads.2022.0040

Cited By