Determination of Minimum Detectable Amount of Gold Using 241Am Be Isotopic Neutron Source by Neutron Activation Analysis
Abstract
In this study, using 241Am-Be neutron irradiation system, minimum detectable amounts of gold were determined by neutron activation analysis. The samples, used in the experiments, were irradiated in an isotropic neutron field of the 241Am-Be neutron source. The induced activities in the samples were measured by gamma spectrometry with a HpGe detector. The obtained result for minimum detectable amount of gold, determined by NAA, is 7,39 μg.
Keywords
NAA, Detection limit, Gold, Isotopic neutron source
241Am Be İzotopik Nötron Kaynağı Kullanılarak Nötron Aktivasyon Analizi ile Altın için Dedekte Edilebilir Minimum Miktarın Belirlenmesi
Abstract
Bu çalışmada, 241Am-Be nötron ışınlama sistemi kullanılarak, nötron aktivasyon analizi ile altının (Au) dedekte edilebilecek minimum miktarı belirlenmiştir. Deneylerde kullanılan numuneler, 241Am-Be nötron kaynağının, izotropik nötron alanında ışınlanmıştır. Numunelerde meydana gelen aktiviteler, HpGe dedektörlü gama spektrometresi kullanılarak ölçülmüştür. Altının nötron aktivasyon analizi ile ölçülebilir minimum miktarı için elde edilen sonuç; 7,39 μg\'dır.
Keywords
NAA, Dedeksiyon limiti, Altın, İzotopik nötron kaynağı