Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis with Thermoluminescence Dosimeters

Number: 22 April 30, 2024
EN TR

Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis with Thermoluminescence Dosimeters

Abstract

Aim: This study investigated the dose calculation accuracy of different treatment planning algorithms used in radiotherapy patients with hip prostheses. Method: The current research produced a tissue-equivalent cylindrical phantom that imitates a leg using a 3D printer. Co-Cr-Mo alloy and Ti-6AI-4V alloy prostheses were placed in the centre of the phantom, respectively. Both prostheses' dose measurements were taken with thermoluminescent dosimeters (TLD) at 92 points. The dose calculation accuracy of the Analytical Anisotropic Algorithm (AAA) and Pencil Beam Convolution (PBC) algorithms, widely used in radiotherapy, were compared with the measurement results. Results: Since the Co-Cr-Mo hip prosthesis has a high density, the number of backscattered photons around it was higher than the Ti-6AI-4V hip prosthesis. The average surface dose of the Co-Cr-Mo alloy was 364.05 cGy, while the average surface dose of the Ti-6AI-4V alloy was 347.79 cGy. Conclusion: It was observed that the dose estimation abilities of the AAA and PBC algorithms decreased as the density of the hip replacement increased. In addition, the AAA algorithm predicted the surface dose in the phantom better than the PBC algorithm.

Keywords

Supporting Institution

Selcuk University

Project Number

21401102

Ethical Statement

This study was supported by the Scientific Research Projects (BAP) Grants Unit, Selcuk University, Konya, TURKEY [Grant Number: 21401102].

References

  1. 1. Antapur P, Mahomed N, Gandhi R. Fractures in the elderly: When is hip replacement a necessity? Clin Interv Aging. 2011;6:1-7. doi: 10.2147/CIA.S10204.
  2. 2. Wang Y. Current concepts in developmental dysplasia of the hip and total hip arthroplasty. Arthroplasty. 2019;1(1). doi: 10.1186/s42836-019-0004-6.
  3. 3. Rahmouni K, Besnard A, Oulmi K, Nouveau C, Hidoussi A, et al. In vitro corrosion response of CoCrMo and Ti-6Al-4V orthopedic implants with Zr columnar thin films. Surface and Coatings Technology. 2022;436-444. doi: 10.1016/j.surfcoat.2022.128310.
  4. 4. Abdul Aziz MZ, Mohd Kamarulzaman FN, Mohd Termizi NAS, Abdul Raof N, Tajuddin AA. Effects of density from various hip prosthesis materials on 6 MV photon beam: A Monte Carlo study. Journal of Radiotherapy in Practice. 2017;16(2):155-160. doi: 10.1017/s1460396917000012.
  5. 5. Palleri F, Baruffaldi F, Angelini AL, Ferri A, Spezi E. Monte Carlo characterization of materials for prosthetic implants and dosimetric validation of Pinnacle3 TPS. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 2008;266(23):5001-5006. doi: 10.1016/j.nimb.2008.08.013.
  6. 6. Wellenberg RHH, Boomsma MF, Van Osch JAC, et al. Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses. European Journal of Radiology. 2017;88:61-70. doi: 10.1016/j.ejrad.2017.01.002.
  7. 7. Çatlı S, Tanır G. Experimental and Monte Carlo evaluation of eclipse treatment planning system for effects on dose distribution of the hip prostheses. Medical Dosimetry. 2013;38(3):332-336. doi: 10.1016/j.meddos.2013.03.005.
  8. 8. De Martino F, Clemente S, Graeff C, Palma G, Cella L. Dose calculation algorithms for external radiation therapy: An overview for practitioners. Applied Sciences. 2021;11(15). doi: 10.3390/app11156806.

Details

Primary Language

English

Subjects

Medical Physics

Journal Section

Research Article

Early Pub Date

April 27, 2024

Publication Date

April 30, 2024

Submission Date

December 6, 2023

Acceptance Date

March 13, 2024

Published in Issue

Year 2024 Number: 22

APA
Gül, O. V., Başaran, H., Yıldırım, A., & İnan, G. (2024). Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis with Thermoluminescence Dosimeters. Istanbul Gelisim University Journal of Health Sciences, 22, 15-28. https://doi.org/10.38079/igusabder.1401159
AMA
1.Gül OV, Başaran H, Yıldırım A, İnan G. Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis with Thermoluminescence Dosimeters. IGUSABDER. 2024;(22):15-28. doi:10.38079/igusabder.1401159
Chicago
Gül, Osman Vefa, Hamit Başaran, Ahmet Yıldırım, and Gökçen İnan. 2024. “Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis With Thermoluminescence Dosimeters”. Istanbul Gelisim University Journal of Health Sciences, nos. 22: 15-28. https://doi.org/10.38079/igusabder.1401159.
EndNote
Gül OV, Başaran H, Yıldırım A, İnan G (April 1, 2024) Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis with Thermoluminescence Dosimeters. Istanbul Gelisim University Journal of Health Sciences 22 15–28.
IEEE
[1]O. V. Gül, H. Başaran, A. Yıldırım, and G. İnan, “Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis with Thermoluminescence Dosimeters”, IGUSABDER, no. 22, pp. 15–28, Apr. 2024, doi: 10.38079/igusabder.1401159.
ISNAD
Gül, Osman Vefa - Başaran, Hamit - Yıldırım, Ahmet - İnan, Gökçen. “Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis With Thermoluminescence Dosimeters”. Istanbul Gelisim University Journal of Health Sciences. 22 (April 1, 2024): 15-28. https://doi.org/10.38079/igusabder.1401159.
JAMA
1.Gül OV, Başaran H, Yıldırım A, İnan G. Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis with Thermoluminescence Dosimeters. IGUSABDER. 2024;:15–28.
MLA
Gül, Osman Vefa, et al. “Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis With Thermoluminescence Dosimeters”. Istanbul Gelisim University Journal of Health Sciences, no. 22, Apr. 2024, pp. 15-28, doi:10.38079/igusabder.1401159.
Vancouver
1.Osman Vefa Gül, Hamit Başaran, Ahmet Yıldırım, Gökçen İnan. Investigation of Dose Calculation Accuracy of Eclipse Treatment Planning System in the Presence of Metal Hip Prosthesis with Thermoluminescence Dosimeters. IGUSABDER. 2024 Apr. 1;(22):15-28. doi:10.38079/igusabder.1401159

 Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)