Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements with Anterior Segment Optical Coherence Tomography

Volume: 12 Number: 1 January 15, 2022
EN TR

Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements with Anterior Segment Optical Coherence Tomography

Abstract

Abstract Aim: To compare corneal thickness measurements made at different times and to evaluate their dependence on individuals. Materials and Methods: The central corneal thickness of the right eyes of 30 healthy young adults was measured by the first investigator at three different times by the Anterior Segment Optical Coherence Tomography (AS-OCT) method. In addition, the fourth measurements were made by the second researcher at a different time. The obtained values were compared. Results: The mean age was 29.8 ± 9.1 years. The mean CCT was 561.90 ± 37.6 µ in the first measurement, 562.20 ± 37.9 µ in the second measurement, and 562.80 ± 37.7 µ in the third measurement. In the measurement of CCT by the second investigator, the mean was 562.20 ± 38.0 µ. In statistical analysis, no significant difference was found between CCT values performed at different times (p<0.05). In addition, when the first measurement of the first researcher was compared with the first measurement of the second researcher, no significant difference was found between the CCT values of the two researchers (p<0.05). Conclusion: Canon HS-100 AS-OCT device had very good reproducibility in the evaluation of CCT. There was no significant difference between the shots of different people. The use of OCT in daily practice is beneficial because it is both a non-contact method and has very good reproducibility in daily practice. Keywords: Central corneal thickness, reproducibility, measurement

Keywords

References

  1. 1. Wang Z, Chen J, Yang B. Posterior corneal surface topographic changes after laser in situ keratomileusis are related to residual corneal bed thickness. Ophthalmology 1999;106(2):406-9; discussion 9-10.
  2. 2. Ucakhan OO, Ozkan M, Kanpolat A. Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry. Journal of cataract and refractive surgery 2006;32(6):970-7.
  3. 3. Holden BA, Mertz GW, McNally JJ. Corneal swelling response to contact lenses worn under extended wear conditions. Investigative ophthalmology & visual science 1983;24(2):218-26.
  4. 4. Auffarth GU, Wang L, Volcker HE. Keratoconus evaluation using the Orbscan Topography System. Journal of cataract and refractive surgery 2000;26(2):222-8.
  5. 5. Barkana Y, Gerber Y, Elbaz U, Schwartz S, Ken-Dror G, Avni I, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. Journal of cataract and refractive surgery 2005;31(9):1729-35.
  6. 6. Swarbrick HA, Wong G, O'Leary DJ. Corneal response to orthokeratology. Optometry and vision science : official publication of the American Academy of Optometry 1998;75(11):791-9.
  7. 7. Kawana K, Tokunaga T, Miyata K, Okamoto F, Kiuchi T, Oshika T. Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis. The British journal of ophthalmology 2004;88(4):466-8.
  8. 8. Piotrowiak I, Soldanska B, Burduk M, Kaluzny BJ, Kaluzny J. Measuring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry. ISRN ophthalmology 2012;2012:869319.

Details

Primary Language

English

Subjects

Health Care Administration

Journal Section

Research Article

Publication Date

January 15, 2022

Submission Date

September 19, 2021

Acceptance Date

October 11, 2021

Published in Issue

Year 2022 Volume: 12 Number: 1

APA
Çelik, F., & Ulucan, P. B. (2022). Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements with Anterior Segment Optical Coherence Tomography. Journal of Contemporary Medicine, 12(1), 10-13. https://doi.org/10.16899/jcm.997505
AMA
1.Çelik F, Ulucan PB. Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements with Anterior Segment Optical Coherence Tomography. J Contemp Med. 2022;12(1):10-13. doi:10.16899/jcm.997505
Chicago
Çelik, Fatih, and Pamuk Betül Ulucan. 2022. “Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements With Anterior Segment Optical Coherence Tomography”. Journal of Contemporary Medicine 12 (1): 10-13. https://doi.org/10.16899/jcm.997505.
EndNote
Çelik F, Ulucan PB (January 1, 2022) Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements with Anterior Segment Optical Coherence Tomography. Journal of Contemporary Medicine 12 1 10–13.
IEEE
[1]F. Çelik and P. B. Ulucan, “Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements with Anterior Segment Optical Coherence Tomography”, J Contemp Med, vol. 12, no. 1, pp. 10–13, Jan. 2022, doi: 10.16899/jcm.997505.
ISNAD
Çelik, Fatih - Ulucan, Pamuk Betül. “Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements With Anterior Segment Optical Coherence Tomography”. Journal of Contemporary Medicine 12/1 (January 1, 2022): 10-13. https://doi.org/10.16899/jcm.997505.
JAMA
1.Çelik F, Ulucan PB. Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements with Anterior Segment Optical Coherence Tomography. J Contemp Med. 2022;12:10–13.
MLA
Çelik, Fatih, and Pamuk Betül Ulucan. “Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements With Anterior Segment Optical Coherence Tomography”. Journal of Contemporary Medicine, vol. 12, no. 1, Jan. 2022, pp. 10-13, doi:10.16899/jcm.997505.
Vancouver
1.Fatih Çelik, Pamuk Betül Ulucan. Evaluation of Individual Dependence and Reproducibility of Central Corneal Thickness Measurements with Anterior Segment Optical Coherence Tomography. J Contemp Med. 2022 Jan. 1;12(1):10-3. doi:10.16899/jcm.997505