Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution

Volume: 16 Number: 2 June 1, 2026
EN TR

Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution

Abstract

In this study, the biomechanical performance of a newly developed spherical connection interface design for angled abutment systems used in all-on-four implant-supported bar-retained overdenture prostheses in completely edentulous patients was investigated. It is well known that bending moments generated in cantilever-extended prostheses lead to high stress concentrations in the implant, abutment, and surrounding bone tissue. In order to mitigate these adverse effects, novel abutment designs featuring a spherical connection surface were developed as an alternative to conventional 15° and 30° angled abutment models, and were evaluated in combination with two different bar cross-sections (I and UCG). The three-dimensional mandibular model was generated based on CT data, and the mechanical behavior of the prosthesis-implant-bone assembly was analyzed using the finite element method in the ANSYS Workbench environment. The models were assumed to be homogeneous, isotropic, and linearly elastic, and a 100 N occlusal load was applied in a 45° oblique direction. The results obtained according to the von Mises criterion demonstrated that the spherical connection designs provided a significant reduction in stress, particularly in the posterior region, compared with conventional models. In the I-section spherical model with 15° angulation, stress reduction of up to 23.2% was observed in the cortical bone surrounding the posterior implant, whereas the spherical models with UCG section demonstrated an improvement ranging between 24 - 27% in the posterior implant body. The findings indicate that the spherical interface design has the potential to reduce biomechanical risks associated with the cantilever effect and to enhance the service life of implant and superstructure components. This approach offers an innovative and clinically applicable solution for preventing complications related to the use of angled abutments.

Keywords

Supporting Institution

Scientific Research Projects Coordination Unit of Iğdır University

Project Number

MUF0422Y14

Ethical Statement

No experiments on humans or animals were conducted in this study. Therefore, ethical committee approval was not required.

Thanks

This study was supported by the Scientific Research Projects Coordination Unit of Iğdır University under Project No. MUF0422Y14

References

  1. Abdeen, R. M., Elsyad, M. A., & Kortam, S. A. (2025). Occlusal contact area, occlusal forces, and marginal bone loss around implants supporting overdentures and fixed implant-supported prostheses in patients with an atrophied mandible: A 1-year randomized clinical trial. The Journal of Prosthetic Dentistry, 134(6), 2320-2327.
  2. Adam, N. M., Attia, O. H., Al-Sulttani, A. O., Mahmood, H. A., As’arry, A., & Rezali, K. A. M. (2020). Numerical analysis for solar panel subjected with an external force to overcome adhesive force in desert areas. CFD Letters, 12(9), 60-75.
  3. Akgün, G. & Kayacan, R. (2023). Abutment-Bar Structure Connection Geometry: An Important Design Parameter for Implant-Supported Bar-Retained Overdentures With Cantilever Extension. Journal of Oral Implantology, 49(3), 330–339.
  4. Almutairi, M., Byun, S., Chee, W., & Park, C. (2024). Conversion of a failing fixed implant prosthesis into an implant-retained overdenture: A clinical report. The Journal of Prosthetic Dentistry. 134(6), 2053-2057
  5. Al-Matrafi, T., Ahmed, K., Basunbul, G., Ismail, M., & Melaibari, A. (2025). Enhanced implant longevity: Finite element optimization of all-on-four implant designs to minimize jawbone stresses. The Saudi Dental Journal, 37(10), 90.
  6. Ausiello, P., Tribst, J. P. M., Ventre, M., Salvati, E., Di Lauro, A. E., Martorelli, M., ... & Watts, D. C. (2021). The role of cortical zone level and prosthetic platform angle in dental implant mechanical response: A 3D finite element analysis. Dental Materials, 37(11), 1688-1697.
  7. Baki, N., Avcu, G. Y., & Tuzlali, M. (2025). Biomechanical evaluation of three full-arch immediate loading protocols in the mandible via finite element analysis: All-on-4, Trefoil, and Five-implant design. BMC Oral Health, 25, 1995.
  8. Bekhit, S. S., Abdelhamid, A. M., & Metwally, N. A. (2025). Evaluation of prosthetic screw loosening in all-on-four implant-supported restorations with two different connection systems. The Journal of Prosthetic Dentistry, 133(4), 1070-1077.

Details

Primary Language

English

Subjects

Biomechanic

Journal Section

Research Article

Publication Date

June 1, 2026

Submission Date

January 27, 2026

Acceptance Date

March 9, 2026

Published in Issue

Year 2026 Volume: 16 Number: 2

APA
Akgün, G. (2026). Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution. Journal of the Institute of Science and Technology, 16(2), 663-677. https://doi.org/10.21597/jist.1872679
AMA
1.Akgün G. Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution. J. Inst. Sci. and Tech. 2026;16(2):663-677. doi:10.21597/jist.1872679
Chicago
Akgün, Gökçen. 2026. “Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution”. Journal of the Institute of Science and Technology 16 (2): 663-77. https://doi.org/10.21597/jist.1872679.
EndNote
Akgün G (June 1, 2026) Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution. Journal of the Institute of Science and Technology 16 2 663–677.
IEEE
[1]G. Akgün, “Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution”, J. Inst. Sci. and Tech., vol. 16, no. 2, pp. 663–677, June 2026, doi: 10.21597/jist.1872679.
ISNAD
Akgün, Gökçen. “Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution”. Journal of the Institute of Science and Technology 16/2 (June 1, 2026): 663-677. https://doi.org/10.21597/jist.1872679.
JAMA
1.Akgün G. Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution. J. Inst. Sci. and Tech. 2026;16:663–677.
MLA
Akgün, Gökçen. “Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution”. Journal of the Institute of Science and Technology, vol. 16, no. 2, June 2026, pp. 663-77, doi:10.21597/jist.1872679.
Vancouver
1.Gökçen Akgün. Influence of a Novel Spherical Connection Interface for Angled Abutments in Cantilever-Extended All-on-Four Bar Prostheses on Stress Distribution. J. Inst. Sci. and Tech. 2026 Jun. 1;16(2):663-77. doi:10.21597/jist.1872679