In this study, control of tidal volume, respiration number and pressure value which are arrived to patient at mechanical ventilator device which is used in intensive care units were performed with fuzzy logic controller. The aim of this system is to reduce workload of aneshesiologist. By calculating tidal volume, respiration number and pressure value, the error Pe(k) between reference pressure value (Pref) and pressure of gas given ill person (Phasta) and error change rate δPe(k) were considered. Performed system is run with the pressure control ventilation (PCV) which is one of the most used modes among other ventilation modes. With the simulation study, it can be said that the obtained datas from the outputs of fuzzy controller are suitable for patients having lung disease.
Primary Language | Turkish |
---|---|
Subjects | Engineering |
Journal Section | Research Articles |
Authors | |
Publication Date | December 1, 2010 |
Submission Date | March 14, 2014 |
Published in Issue | Year 2010 Volume: 14 Issue: 2 |
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.