Research Article

Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors

Volume: 12 Number: 2 August 20, 2021
EN TR

Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors

Abstract

Aim: An effective immobilization is warranted to minimize set up deviations while treating head & neck (HN) and brain tumors with intensity modulated radiotherapy (IMRT). In this study we evaluated the effectiveness of thermoformable vacuum head cushion (TVHC) on immobilization of these patients via weekly electronic digital port films (EDPF). Material and Method: Twenty patients treated with IMRT for HN and Brain tumors were enrolled in the study. Thermoplastic mask with additional standard plastic head rest (n:10) in Group A or TVHC supporting the head and neck region (n:10) in Group B were used for immobilization. Weekly EDPF images were obtained on the linear accelerator. Set-up displacements in latero-medial, anteroposterior and cranio-caudal directions in these images were compared with variation data of first treatment. Results: When each group was analyzed in terms of deviations in all 3 directions, deviations calculated in anteroposterior and craniocaudal directions were significantly higher for TVHC group (p:0.039 and p:0.002 respectively). Deviations >4mm in all three directions were also analyzed and similarly, higher rates of >4mm deviations were recorded for TVHC group in anteroposterior and craniocaudal directions (p:0.023 and p:<0.001 respectively) which of both were also statistically significant. Conclusion: We found that TVHC utilization in HN and brain IMRT for immobilization leads to significantly higher set-up deviations so future investigations with larger patient populations and evaluating set up accuracy via more sensitive alignments are needed to suggest TVHC for utilization in daily clinical practice.

Keywords

References

  1. 1. Webb S. Intensity-modulated radiation therapy. Bristol (UK): Institute of Physics Publishing; 2001.
  2. 2. Palta JR, Mackie TR, editors. Intensity-modulated radiation therapy—the state of the art. Madison (WI): Medical Physics Publishing; 2003.
  3. 3. Boda-Heggemann J, Walter C, Rahn A, et al. Repositioning accuracy of two different mask systems-3D revisited: Comparison using true 3D/3D matching with cone-beam CT. Int J Radiat Oncol Biol Phys 2006;66:1568–1575.
  4. 4. Hong TS, Tome WA, Chappell RJ, et al. The impact of Daily setup variations on head-and neck intensity modulated radiation therapy. Int J Radiat Oncol Biol Phys 2005;61:779–788.
  5. 5. Prisciandaro JI, Frechette CM, Herman MG, et al. A methodology to determine margins by EPID measurements of patient setup variation and motion as applied to immobilization devices. Med Phys 2004;31:2978–2988.
  6. 6. Hatherly KE, Smylie JC, Rodger A, et al. A double exposed portal image comparison between electronic portal imaging hard copies and port films in radiation therapy treatment setup confirmation to determine its clinical application in radiotherapy center. Int J Radiat Oncol Biol Phys 2001;49:191–198.
  7. 7. Brock KK, McShan DL, Balter JM. A comparison of computer controlled versus manual on-line patient setup adjustment. J Appl Med Phys 2002;3:241–247.
  8. 8. de Boer HC, van Sornsen de Koste JR, et al. Electronic portal image assisted reduction of systematic set-up errors in head and neck irradiation. Radiother Oncol 2001;61:299–308.

Details

Primary Language

English

Subjects

Health Care Administration

Journal Section

Research Article

Publication Date

August 20, 2021

Submission Date

October 14, 2020

Acceptance Date

November 25, 2020

Published in Issue

Year 2021 Volume: 12 Number: 2

APA
Özkan, E. E., Özseven, A., & Kaymak, Z. A. (2021). Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors. Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi, 12(2), 131-139. https://doi.org/10.22312/sdusbed.810408
AMA
1.Özkan EE, Özseven A, Kaymak ZA. Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors. Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi. 2021;12(2):131-139. doi:10.22312/sdusbed.810408
Chicago
Özkan, Emine Elif, Alper Özseven, and Zümrüt Arda Kaymak. 2021. “Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors”. Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi 12 (2): 131-39. https://doi.org/10.22312/sdusbed.810408.
EndNote
Özkan EE, Özseven A, Kaymak ZA (August 1, 2021) Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors. Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi 12 2 131–139.
IEEE
[1]E. E. Özkan, A. Özseven, and Z. A. Kaymak, “Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors”, Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi, vol. 12, no. 2, pp. 131–139, Aug. 2021, doi: 10.22312/sdusbed.810408.
ISNAD
Özkan, Emine Elif - Özseven, Alper - Kaymak, Zümrüt Arda. “Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors”. Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi 12/2 (August 1, 2021): 131-139. https://doi.org/10.22312/sdusbed.810408.
JAMA
1.Özkan EE, Özseven A, Kaymak ZA. Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors. Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi. 2021;12:131–139.
MLA
Özkan, Emine Elif, et al. “Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors”. Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi, vol. 12, no. 2, Aug. 2021, pp. 131-9, doi:10.22312/sdusbed.810408.
Vancouver
1.Emine Elif Özkan, Alper Özseven, Zümrüt Arda Kaymak. Efficacy of Thermoformable Vacuum Head Cushion on Set-up Accuracy in IMRT of Head & Neck and Brain Tumors. Süleyman Demirel Üniversitesi Sağlık Bilimleri Dergisi. 2021 Aug. 1;12(2):131-9. doi:10.22312/sdusbed.810408