BibTex RIS Cite

Topographic Features of Extracorporeal Circulation Circuit Surface Observed by Atomic Force Microscopy; Coated or Non-coated?

Year 2016, Volume: 23 Issue: 2, 0 - 0, 30.06.2016

Abstract

Abstract
Objective
Polymer based materials used in the manufacture of extracorporeal circulation (ECC) circuit that is the main part of the heart lung machine used in cardiac surgery, poses the unavoidable inflammatory and coagulation reactions because of the contact with the tissue and biological fluids particularly the blood. The reason for these reactions may include chemical properties of the material used but the surface properties of the contact surface are also important. Atomic force microscopy (AFM) is useful method to analyze the surface morphology of a thin line and obtain information about the thickness, roughness and height asymmetries values such as the skewness and kurtosis of the line. The aim of the study is demonstrating the ECC circuits’ topographic features that hypothesized as one of the main reason for side effect reactions as inflammation and coagulation.
Methods
For this purpose, non-coated (n=3), heparine coated (n=3) and phosphorylcholin coated (n=3) ECC lines imaged by AFM and surface photographs, 3d topography, skewness and kurtosis of the sectional images and surface properties were evaluated at preoperative and postoperative.
Results
Atomic force microscopy images of the non-coated (n=3), heparine coated (n=3) and phosphorylcholin coated (n=3) ECC lines revealed that none of the circuit surface morphologies are enough for a biocompatible device.
Conclusion
Considering the biomaterial design approaches to date, although the influence of the chemistry of the substrate is certainly of high importance, micro/nanoscale topography on biomaterial surfaces is a promising new strategy and shows potential for improving blood compatibility.

Year 2016, Volume: 23 Issue: 2, 0 - 0, 30.06.2016

Abstract

There are 0 citations in total.

Details

Journal Section Research Articles
Authors

Kadir Çeviker

Şahin Şahinalp This is me

Rasih Yazkan This is me

Publication Date June 30, 2016
Submission Date April 19, 2016
Published in Issue Year 2016 Volume: 23 Issue: 2

Cite

Vancouver Çeviker K, Şahinalp Ş, Yazkan R. Topographic Features of Extracorporeal Circulation Circuit Surface Observed by Atomic Force Microscopy; Coated or Non-coated?. Med J SDU. 2016;23(2).

                                                                                               14791 


Süleyman Demirel Üniversitesi Tıp Fakültesi Dergisi/Medical Journal of Süleyman Demirel University is licensed under Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International.