Heavy metal pollution
has become a serious threat for human health, living resources and ecological
systems. Therefore, the determination of heavy metal concentrations is
important. In this study; metal sensor based nanoparticles was prepared to
determine heavy metal ions. For this purpose, solid surface electrodes were
prepared using the new synthesized calixarene functionalized multiwalled carbon
nanotube as nanopartule and developed a new method to determine metal on
prepared surface.
To determine the
performance of the prepared CNT-Calix-Cra/GCE electrode, SWASV (square wave
anodic stripping voltammetry) voltamogramme calibration graphs were obtained
using certain range of standards. When the
sensitivity and operating range of the modified electrode examined it was
observed that a higher Cd(II) sensitivity. The optimum pH, Calix [4] arene
immobilized multi-walled carbon nanotube quantity, buffer concentration and
work potential for the modified electrode, respectively; pH-5,5µL, 0.1 M
and -1.3 V were found. Deposition time
120 s and linear operating range 3,23.10-7 - 4,81.10-5molL-1
for CNT-Calix-Cra/GCE and correlation coefficient was recorded as 0,9905.
Ağır metal kirliliği insan
sağlığı, canlı kaynaklara ve ekolojik sistemleri için ciddi bir tehdit haline
gelmiştir. Bu nedenle ağır metallerin tayini oldukça önemlidir. Bu çalışmada;
ağır metal iyonlarının kolayca tayin edilebilmesi için, nanopartikül temelli
metal sensör hazırlanmıştır. Nanopartikül olarak yeni sentezlenen
kaliksarenlefonksiyonlandırılmış çok duvarlı karbon nanotüp kullanılarak katı
yüzey elektrotlar hazırlanmış, hazırlanan yüzeylerde metal tayinin
yapılabilmesi için yeni bir yöntem geliştirilmiştir.
Hazırlanan CNT-Calix-Cra/GCE elektrodun çalışmada analizi yapılması
istenilen Cd(II) iyonuna karşı performansını belirlemek için belli aralıkta
hazırlanan standart Cd(II) iyonuna karşı SWASV( kare dalda anodik sıyırma)
voltammogram grafikleri elde edilmiştir. Duyarlılıkları ve çalışma aralıkları
incelendiğinde modifiye edilmiş olan elektrodun Cd(II) duyarlılığının daha
yüksek olduğu gözlemlenmiştir. Modifiye elektrot için optimumpH,
kaliks[4]arenimmobilize çok duvarlı karbon nanotüp miktarı, tampon derişimi ve
çalışma potansiyeli sırasıyla; pH-5,5µL, 0,1 M ve -1,3 V olarak bulunmuştur.
Modifiyeelektrodun biriktirme süresi 120 s ve CNT-Calix-Cra/GCE için doğrusal
çalışma aralığı 3,23.10-7 - 4,81.10-5 molL-1ve
korelasyon sabiti 0,9905 olarak kaydedilmiştir.
Journal Section | Research Articles |
---|---|
Authors | |
Publication Date | April 28, 2017 |
Submission Date | April 28, 2017 |
Published in Issue | Year 2017 Volume: 43 Issue: 1 |
Journal Owner: On behalf of Selçuk University Faculty of Science, Rector Prof. Dr. Hüseyin YILMAZ
Selcuk University Journal of Science Faculty accepts articles in Turkish and English with original results in basic sciences and other applied sciences. The journal may also include compilations containing current innovations.
It was first published in 1981 as "S.Ü. Fen-Edebiyat Fakültesi Dergisi" and was published under this name until 1984 (Number 1-4).
In 1984, its name was changed to "S.Ü. Fen-Edeb. Fak. Fen Dergisi" and it was published under this name as of the 5th issue.
When the Faculty of Letters and Sciences was separated into the Faculty of Science and the Faculty of Letters with the decision of the Council of Ministers numbered 2008/4344 published in the Official Gazette dated 3 December 2008 and numbered 27073, it has been published as "Selcuk University Journal of Science Faculty" since 2009.
It has been scanned in DergiPark since 2016.
Selcuk University Journal of Science Faculty is licensed under a Creative Commons Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) License.