Year 2015, Volume 16, Issue 2, Pages 132 - 135 2015-07-15

Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması

Mustafa DOĞAN [1] , Nurgül AVCI [2] , Mehmet DURMAZ [3] , Fatih KARTAL [4] , Mehmet AKAY [5] , Sibel İNAN [6]

270 763

Objective: To compare Scheimpflug photography(SP), optical coherence tomography(OCT) and ultrasonic pachymetry(USP) techniques in central corneal thickness(CCT) measurement.Material and Methods: The CCT was measured in 43 healthy young adults(86 eyes in total) by using SP, OCT and USP). Central corneal thickness, age, gender, intraocular pressure and auto refractokeratometer values obtained and used for comparison.Results: The mean age of was 21,35 ± 0,75 years. The mean CCT with USP was 558,42 ± 38,87 (477620) microns for right eye and 556,84 ± 39,83 microns for left eye. The mean CCT with SP was 555,02 ±

Amaç: Santral kornea kalınlığının(SKK) ölçümünde ultrasonik pakimetri(UP), Scheimpflug fotograflama(SF) ve optik kohorens tomografi(OKT) yöntemleri ile elde edilen değerlerin karşılaştırılması.
Gereç ve Yöntem: 43 sağlıklı genç erişkinin sağ ve sol göz (toplam 86 göz) santral korneal kalınlığı UP, SF ve OKT yöntemleri ile ölçüldü. Santral kornea kalınlığı, yaş, cinsiyet, göz içi basıncı ve otorefraktokeratometre değerleri kaydedildi. Elde edilen değerler karşılaştırıldı.
Bulgular: Ortalama yaş 21,35 ± 0,75 (20-24) idi. UP ile ölçülen ortalama SKK sağ göz için 558,42 ± 38,87 (477-620) mikron, sol göz için 556,84 ± 39,83 (477-641) mikron idi.SF ile ölçülen ortalama SKK sağ göz için 555,02 ± 37,90 (486-631) mikron, sol göz için 550,44 ± 42,30 (453-630) mikron idi.OKT ile ölçülen ortalama SKK sağ göz için 563, 52 ± 38,52 (492-644) mikron, sol göz için 563,72 ± 39,84(482-644) mikron idi.Her üç yöntem ile ölçülen değerler karşılatırıldığında UP ve SF ile elde edilen değerler arasında istatistiki olarak anlamlı fark saptanmaz iken OKT ile elde edilen değerler ile arasında UP(P<0.003) ve SF(P<0.001) ile elde edilen değerler arasında istatistiki olarak anlamlı fark tespit edildi. OKT ile elde edilen değerler daha yüksek bulundu (yaklaşık 7 mikron).
Sonuç: Yapılan SKK ölçümünde UP ve SF ile elde edilen değerler birbirleri ile uyumludur. OKT ile elde edilen değerler daha yüksektir. Günlük uygulamada bu durum dikkate alınmalıdır.

        
 
 
Text-to-speech function is limited to 100 characters
 
  • Pflugfelder SC, Liu Z, Feuer W, et al. Corneal thickı- ness indices discriminate between keratoconus and contact lens-induced corneal thinning. Ophthalmo- logy 2002;109(12):2336-41.
  • Al-Mezaine HS, Al-Amro SA, Kangave D, et al. Com- parison between central corneal thickness measure- ments by oculus pentacam and ultrasonic pachymetry. International Ophthalmology. 2008;28(5):333–8.
  • Carrasco AJA, Vicent AD, Romín DM, et al. OCT tech- nique for evaluating diurnal variation in corneal thick- ness. Emmetropia 2013;4(4):197-201.
  • Şekil II: Optik kohorens tomografi Miglior S, Albe E, Guareschi M, et al. Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central cor- neal thickness. Br J Ophthalmol 2004;88(2):174–7.
  • Kawana K, Tokunaga T, Miyata K et al. Comparison of corneal thickness measurements using Orbscan II, noncontact specular microscopy, and ultrasonic pa- chymetry in eyes after laser in situ keratomileusis. Br J Ophthalmol 2004;88(4):466–8.
  • Piotrowiak I, Soldanska B, Burduk M, et al. Measu- ring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry. ISRN Ophthal- mol 2012:869319.
  • Leung DY, Lam DK, Yeung BY, Lam DS. Comparison between central corneal thickness measurements by ultrasound pachymetry and optical coherence tomog- raphy. Clin Experiment Ophthalmol 2006;34(8):751–4.
  • Grewal DS, Brar GS, Grewal SP. Assessment of central corneal thickness in normal, keratoconus, and post-la- ser in situ keratomileusis eyes using Scheimpflug ima- ging, spectral domain optical coherence tomography, and ultrasound pachymetry. J Cataract Refract Surg 2010;36(6):954–64
  • Fishman GR, Pons ME, Seedor JA, et al. Assessment of central corneal thickness using optical coherence tomography. J Cataract Refract Surg 2005; 31(4):707– 11
  • Buyuk K, Bozkurt B, Kamıs U, Ozkagnıcı A, Okudan S. Normal ve Keratokonuslu Gözlerde Ultrasonik Pa- kimetri ve OCULUS Pentacam ile Ölçülen Santral Kor- nea Kalınlıklarının Karşılaştırılması. Turk J Ophthalmol 2011;41(2):104-7.
Primary Language en
Journal Section Articles
Authors

Author: Mustafa DOĞAN

Author: Nurgül AVCI

Author: Mehmet DURMAZ

Author: Fatih KARTAL

Author: Mehmet AKAY

Author: Sibel İNAN

Bibtex
APA DOĞAN, M , AVCI, N , DURMAZ, M , KARTAL, F , AKAY, M , İNAN, S . (). . , 16 (2), 132-135. Retrieved from http://dergipark.org.tr/kocatepetip/issue/17437/182657
MLA DOĞAN, M , AVCI, N , DURMAZ, M , KARTAL, F , AKAY, M , İNAN, S . "". 16 (): 132-135 <http://dergipark.org.tr/kocatepetip/issue/17437/182657>
Chicago DOĞAN, M , AVCI, N , DURMAZ, M , KARTAL, F , AKAY, M , İNAN, S . "". 16 (): 132-135
RIS TY - JOUR T1 - AU - Mustafa DOĞAN , Nurgül AVCI , Mehmet DURMAZ , Fatih KARTAL , Mehmet AKAY , Sibel İNAN Y1 - 2015 PY - 2015 N1 - DO - T2 - Kocatepe Tıp Dergisi JF - Journal JO - JOR SP - 132 EP - 135 VL - 16 IS - 2 SN - 1302-4612-2149-7869 M3 - UR - Y2 - 2019 ER -
EndNote %0 Kocatepe Medical Journal %A Mustafa DOĞAN , Nurgül AVCI , Mehmet DURMAZ , Fatih KARTAL , Mehmet AKAY , Sibel İNAN %T %D 2015 %J Kocatepe Tıp Dergisi %P 1302-4612-2149-7869 %V 16 %N 2 %R %U
ISNAD DOĞAN, Mustafa , AVCI, Nurgül , DURMAZ, Mehmet , KARTAL, Fatih , AKAY, Mehmet , İNAN, Sibel . "". Kocatepe Tıp Dergisi 16 / 2 (July 2015): 132-135.