Applying of external fixator in the distal radius (An anatomic study in cadaver forearms)

Cilt: 30 Sayı: 1 11 Eylül 2006
  • Ercan Olcay
  • Mehmet Asik
  • Murat Mert
  • Ayhan Kara
PDF İndir
EN TR

Applying of external fixator in the distal radius (An anatomic study in cadaver forearms)

Abstract

We performed an anatomic study in 16 cadaver forearms to determine the safety of percutaneous proximal pin placement for external fixation of distal radius fractures. Two 3mm. cortical pins were placed along the dorsoradial ridge 8 cm. proximal to the radial styloid process. Soft tissue injuries occured in 11 cases. Tendon injuries occured the extensor carpi radialis brevis in 3, extensor carpi radialis longus 4, abductor pollicis longus and brachioradialis in 1 case. Two pins transfixed the superficial branch of the radial nerve. A high degree of variability in the position of the superficial branch of the radial nerve was found and no safe zone for percutaneous pin placement could be found. We advice open pin placement for distal radius fractures because of percutaneous pin placement was not safe.

Keywords

Ayrıntılar

Birincil Dil

İngilizce

Konular

-

Bölüm

-

Yazarlar

Ercan Olcay Bu kişi benim

Mehmet Asik Bu kişi benim

Murat Mert Bu kişi benim

Ayhan Kara Bu kişi benim

Yayımlanma Tarihi

11 Eylül 2006

Gönderilme Tarihi

6 Mart 2014

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 1996 Cilt: 30 Sayı: 1

Kaynak Göster

APA
Olcay, E., Asik, M., Mert, M., & Kara, A. (2006). Applying of external fixator in the distal radius (An anatomic study in cadaver forearms). Acta Orthopaedica et Traumatologica Turcica, 30(1), 59-61. https://doi.org/10.3944/aott.v30i1.4151
AMA
1.Olcay E, Asik M, Mert M, Kara A. Applying of external fixator in the distal radius (An anatomic study in cadaver forearms). Acta Orthopaedica et Traumatologica Turcica. 2006;30(1):59-61. doi:10.3944/aott.v30i1.4151
Chicago
Olcay, Ercan, Mehmet Asik, Murat Mert, ve Ayhan Kara. 2006. “Applying of external fixator in the distal radius (An anatomic study in cadaver forearms)”. Acta Orthopaedica et Traumatologica Turcica 30 (1): 59-61. https://doi.org/10.3944/aott.v30i1.4151.
EndNote
Olcay E, Asik M, Mert M, Kara A (01 Eylül 2006) Applying of external fixator in the distal radius (An anatomic study in cadaver forearms). Acta Orthopaedica et Traumatologica Turcica 30 1 59–61.
IEEE
[1]E. Olcay, M. Asik, M. Mert, ve A. Kara, “Applying of external fixator in the distal radius (An anatomic study in cadaver forearms)”, Acta Orthopaedica et Traumatologica Turcica, c. 30, sy 1, ss. 59–61, Eyl. 2006, doi: 10.3944/aott.v30i1.4151.
ISNAD
Olcay, Ercan - Asik, Mehmet - Mert, Murat - Kara, Ayhan. “Applying of external fixator in the distal radius (An anatomic study in cadaver forearms)”. Acta Orthopaedica et Traumatologica Turcica 30/1 (01 Eylül 2006): 59-61. https://doi.org/10.3944/aott.v30i1.4151.
JAMA
1.Olcay E, Asik M, Mert M, Kara A. Applying of external fixator in the distal radius (An anatomic study in cadaver forearms). Acta Orthopaedica et Traumatologica Turcica. 2006;30:59–61.
MLA
Olcay, Ercan, vd. “Applying of external fixator in the distal radius (An anatomic study in cadaver forearms)”. Acta Orthopaedica et Traumatologica Turcica, c. 30, sy 1, Eylül 2006, ss. 59-61, doi:10.3944/aott.v30i1.4151.
Vancouver
1.Ercan Olcay, Mehmet Asik, Murat Mert, Ayhan Kara. Applying of external fixator in the distal radius (An anatomic study in cadaver forearms). Acta Orthopaedica et Traumatologica Turcica. 01 Eylül 2006;30(1):59-61. doi:10.3944/aott.v30i1.4151