Glove perforation time and frequency in total hip arthroplasty procedures

Öz

Objective: The aim of the present study was to investigate glove perforation rate and time and evaluate the factors affecting glove perforation in total hip arthroplasty (THA).
Methods: Nine hundred seventy-nine gloves used in 57 THA procedures were assessed according to the perforation. Forty-four (77.2%) procedures were primary THA and 13 (22.8%) were revision THA. Gloves were changed when perforated, become dirty with blood or blood products, and before bone cementing. All gloves were filled with water at the end of the operation and controlled for perforation. Two hundred and one surgical gloves used during scrubbing and removed after draping the patient were examined as the control group. The location (which finger), number and time of the perforation, surgery type and duration, and distribution of the perforation location according to the surgical team were assessed.
Results: Patients’ mean age was 62.9±14.6 (range: 33 to 97) years and the mean surgery duration was 162.9±32.0 minutes. Thirty-two glove perforations were noted in 19 of the operations. Of these perforations, 28 belonged to the surgeons and first assistants. There was no significant difference between the dominant or non-dominant hand according to the location of perforations. Perforations in the first and second fingers of the gloves accounted for 81.3% of all perforations. There was no significant difference in terms of number of gloves used, perforation numbers and operation duration between the primary and revision THA procedures. Two perforated gloves (0.99%) were found in the control group and the difference between the number of perforations in the control and study groups was significant (p=0.048).
Conclusion: We recommend the use of two pairs of gloves to avoid the risk of contamination and protect the surgical team from infectious disease in major surgeries like THA. Surgical gloves should be changed when they are excessively contaminated with surgical fluids and the surgeon and first assistant should also change their outer gloves at an average of every 90 minutes.

Anahtar Kelimeler

Ayrıntılar

Birincil Dil

İngilizce

Konular

Sağlık Kurumları Yönetimi

Bölüm

Araştırma Makalesi

Yazarlar

İbrahim Kaya Bu kişi benim

İbrahim Sungur Bu kişi benim

Murat Yilmaz Bu kişi benim

Musa Korkmaz Bu kişi benim

Ercan Cetinus Bu kişi benim

Yayımlanma Tarihi

23 Mart 2012

Gönderilme Tarihi

7 Mart 2014

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2012 Cilt: 46 Sayı: 1

Kaynak Göster

APA
Kaya, İ., Ugras, A., Sungur, İ., Yilmaz, M., Korkmaz, M., & Cetinus, E. (2012). Glove perforation time and frequency in total hip arthroplasty procedures. Acta Orthopaedica et Traumatologica Turcica, 46(1), 57-60. https://izlik.org/JA89PU98SA
AMA
1.Kaya İ, Ugras A, Sungur İ, Yilmaz M, Korkmaz M, Cetinus E. Glove perforation time and frequency in total hip arthroplasty procedures. Acta Orthopaedica et Traumatologica Turcica. 2012;46(1):57-60. https://izlik.org/JA89PU98SA
Chicago
Kaya, İbrahim, A. Ugras, İbrahim Sungur, Murat Yilmaz, Musa Korkmaz, ve Ercan Cetinus. 2012. “Glove perforation time and frequency in total hip arthroplasty procedures”. Acta Orthopaedica et Traumatologica Turcica 46 (1): 57-60. https://izlik.org/JA89PU98SA.
EndNote
Kaya İ, Ugras A, Sungur İ, Yilmaz M, Korkmaz M, Cetinus E (01 Mart 2012) Glove perforation time and frequency in total hip arthroplasty procedures. Acta Orthopaedica et Traumatologica Turcica 46 1 57–60.
IEEE
[1]İ. Kaya, A. Ugras, İ. Sungur, M. Yilmaz, M. Korkmaz, ve E. Cetinus, “Glove perforation time and frequency in total hip arthroplasty procedures”, Acta Orthopaedica et Traumatologica Turcica, c. 46, sy 1, ss. 57–60, Mar. 2012, [çevrimiçi]. Erişim adresi: https://izlik.org/JA89PU98SA
ISNAD
Kaya, İbrahim - Ugras, A. - Sungur, İbrahim - Yilmaz, Murat - Korkmaz, Musa - Cetinus, Ercan. “Glove perforation time and frequency in total hip arthroplasty procedures”. Acta Orthopaedica et Traumatologica Turcica 46/1 (01 Mart 2012): 57-60. https://izlik.org/JA89PU98SA.
JAMA
1.Kaya İ, Ugras A, Sungur İ, Yilmaz M, Korkmaz M, Cetinus E. Glove perforation time and frequency in total hip arthroplasty procedures. Acta Orthopaedica et Traumatologica Turcica. 2012;46:57–60.
MLA
Kaya, İbrahim, vd. “Glove perforation time and frequency in total hip arthroplasty procedures”. Acta Orthopaedica et Traumatologica Turcica, c. 46, sy 1, Mart 2012, ss. 57-60, https://izlik.org/JA89PU98SA.
Vancouver
1.İbrahim Kaya, A. Ugras, İbrahim Sungur, Murat Yilmaz, Musa Korkmaz, Ercan Cetinus. Glove perforation time and frequency in total hip arthroplasty procedures. Acta Orthopaedica et Traumatologica Turcica [Internet]. 01 Mart 2012;46(1):57-60. Erişim adresi: https://izlik.org/JA89PU98SA