Internal fixation of bilateral sacroiliac dislocation with transiliac locked plate: a biomechanical study on pelvic models
Abstract
Methods: Bilateral sacroiliac dislocations were created in 21 pelvic models and divided into three different fixation method groups. Group 1 was fixed using posterior tension band plating with a 3.5 mm locked plate combined with fixed-angle locked 3.5 mm screw fixation of bilateral S1 vertebra pedicles through suitable holes of the plate. Group 2 underwent posterior tension band plating with a 3.5 mm locked plate combined with bilateral spongious iliosacral screw fixation and Group 3 bilateral iliosacral spongious screw fixation alone. The ultimate load to failure and load for 10 mm of displacement for all three groups were compared.
Results: The average loads to failure for Groups 1, 2 and 3 were 1775, 2084 and 2230 N, respectively, and average loads for 10 mm of displacement were 1033, 1884 and 2013 N, respectively. Group 2 and 3 had the strongest fixation constructs although there was no statistically significant difference between these two groups (p=0.452). Group 2 and 3 were superior to Group 1 in terms of loads for 10 mm of displacement. There was no significant difference between Group 2 and 3 in this regard (p=0.397).
Conclusion: Iliosacral screws are superior to bilateral S1 pedicle fixation through posterior tension band plating. However, the combination of tension band plating with iliosacral screw fixation does not improve the stability of the posterior pelvic ring.
Keywords
Kaynakça
- Routt ML Jr, Nork SE, Mills WJ. High-energy pelvic ring disruptions. Orthop Clin North Am 2002;33:59-72.
- Routt ML Jr, Nork SE, Mills WJ. Percutaneous fixation of pelvic ring disruptions. Clin Orthop Relat Res 2000;(375):15
- Esenkaya, I. A morphologic evaluation of the iliosacral joint and plate fixation on a pelvic model using a S1 pedicular screw, transiliosacral screws, and a compression rod for iliosacral joint injuries. [Article in Turkish] Acta Orthop Traumatol Turc 2002;36:432-41.
- Ayvaz M, Yilmaz G, Akpinar E, Acaro¤lu RE. Percutaneous iliolumbar screw fixation in a patient with unstable pelvic fracture and transitional lumbar vertebrae and cadaveric illustration of the fixation. [Article in Turkish] Acta Orthop Traumatol Turc 2009;43:444-7.
- Humphrey CA, Liu Q, Templeman DC, Ellis TJ. Locked plates reduce displacement of vertically unstable pelvic fractures in a Mechanical Testing Model. J Trauma 2010;69: 1230Lebwohl NH, Cunningham BW, Dmitriev A, Shimamoto N, Gooch L, Devlin V, et al. Biomechanical comparison of lumbosacral fixation techniques in a calf spine model. Spine (Phila Pa 1976) 2002;27:2312-20.
- Sar C, Kilicoglu O. S1 pediculoiliac screw fixation in instabilities of the iliosacral complex: biomechanical study and report of two cases. J Orthop Trauma 2003;17:262-70. van Zwienen CM, van den Bosch EW, Snijders CJ, Kleinrensink GJ, van Vugt AB. Biomechanical comparison of iliosacral screw techniques for unstable pelvic ring fractures. J Orthop Trauma 2004;18:589-95.
- Stocks GW, Gabel GT, Noble PC, Hanson GW, Tullos HS. Anterior and posterior internal fixation of vertical shear fractures of the pelvis. J Orthop Res 1991;9:237-45.
- Yinger K, Scalise J, Olson SA, Bay BK, Finkemeier CG. Biomechanical comparison of posterior pelvic ring fixation. J Orthop Trauma 2003;17:481-7.
Ayrıntılar
Birincil Dil
İngilizce
Konular
Sağlık Kurumları Yönetimi
Bölüm
Araştırma Makalesi
Yazarlar
Orcun Sahin
Bu kişi benim
Huseyin Demirors
Bu kişi benim
Rahmi Akgun
Bu kişi benim
Ismail Tuncay
Bu kişi benim
Yayımlanma Tarihi
7 Şubat 2014
Gönderilme Tarihi
7 Mart 2014
Kabul Tarihi
-
Yayımlandığı Sayı
Yıl 2013 Cilt: 47 Sayı: 6