Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate?

Abstract

Objective: Current literature describes improved clinical outcomes and a minor rate of pseudoarthrosis following operatively treated clavicular fractures. We investigated the feasibility of using a standard 3.5 mm AO locking compression plate (LCP) of adequate length for the stabilisation of mid-shaft fractures of the clavicle.
Methods: The length and acromial and diaphyseal curvature depths were measured in 49 cadaveric clavicles. We then assessed how well the 6, 7, 8 and 9-hole plates fit on the clavicles.
Results: The mean clavicular length was 155±12 mm, with a mean acromial curvature of 18.1± 3.7 mm and a mean diaphyseal curvature of 12 mm±4 mm. The optimum plate for the clavicle was a 7-hole LCP, providing adequate fixation in 48 of the 49 clavicles.
Conclusion: The described technique for plate osteosynthesis of the clavicle with AO locking compression plate is feasible and results in a biomechanically strong construct for mid-shaft fractures. With the use of a locking plate, comminuted fractures may be bridged without a reduction in the strength of the construct.

Keywords

Ayrıntılar

Birincil Dil

İngilizce

Konular

Sağlık Kurumları Yönetimi

Bölüm

Araştırma Makalesi

Yazarlar

Wolfgang Grechenig Bu kişi benim

Nima Heidari Bu kişi benim

Ottmal Leitgoeb Bu kişi benim

Walter Prager Bu kişi benim

Wolfgang Pichler Bu kişi benim

Annelie M. Weinber Bu kişi benim

Yayımlanma Tarihi

3 Mayıs 2011

Gönderilme Tarihi

11 Mart 2014

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2011 Cilt: 45 Sayı: 2

Kaynak Göster

APA
Grechenig, W., Heidari, N., Leitgoeb, O., Prager, W., Pichler, W., & M. Weinber, A. (2011). Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate? Acta Orthopaedica et Traumatologica Turcica, 45(2), 115-119. https://izlik.org/JA84BD89GJ
AMA
1.Grechenig W, Heidari N, Leitgoeb O, Prager W, Pichler W, M. Weinber A. Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate? Acta Orthopaedica et Traumatologica Turcica. 2011;45(2):115-119. https://izlik.org/JA84BD89GJ
Chicago
Grechenig, Wolfgang, Nima Heidari, Ottmal Leitgoeb, Walter Prager, Wolfgang Pichler, ve Annelie M. Weinber. 2011. “Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate?”. Acta Orthopaedica et Traumatologica Turcica 45 (2): 115-19. https://izlik.org/JA84BD89GJ.
EndNote
Grechenig W, Heidari N, Leitgoeb O, Prager W, Pichler W, M. Weinber A (01 Mayıs 2011) Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate? Acta Orthopaedica et Traumatologica Turcica 45 2 115–119.
IEEE
[1]W. Grechenig, N. Heidari, O. Leitgoeb, W. Prager, W. Pichler, ve A. M. Weinber, “Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate?”, Acta Orthopaedica et Traumatologica Turcica, c. 45, sy 2, ss. 115–119, May. 2011, [çevrimiçi]. Erişim adresi: https://izlik.org/JA84BD89GJ
ISNAD
Grechenig, Wolfgang - Heidari, Nima - Leitgoeb, Ottmal - Prager, Walter - Pichler, Wolfgang - M. Weinber, Annelie. “Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate?”. Acta Orthopaedica et Traumatologica Turcica 45/2 (01 Mayıs 2011): 115-119. https://izlik.org/JA84BD89GJ.
JAMA
1.Grechenig W, Heidari N, Leitgoeb O, Prager W, Pichler W, M. Weinber A. Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate? Acta Orthopaedica et Traumatologica Turcica. 2011;45:115–119.
MLA
Grechenig, Wolfgang, vd. “Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate?”. Acta Orthopaedica et Traumatologica Turcica, c. 45, sy 2, Mayıs 2011, ss. 115-9, https://izlik.org/JA84BD89GJ.
Vancouver
1.Wolfgang Grechenig, Nima Heidari, Ottmal Leitgoeb, Walter Prager, Wolfgang Pichler, Annelie M. Weinber. Is plating of mid-shaft clavicular fractures possible with a conventional straight 3.5 millimeter locking compression plate? Acta Orthopaedica et Traumatologica Turcica [Internet]. 01 Mayıs 2011;45(2):115-9. Erişim adresi: https://izlik.org/JA84BD89GJ