A new femoral nail with single distal locking screw for maximum interfragment rotational stability
Öz
Methods: We used a total of 30 composite femurs, ten for each group. We determined interfragment rotational displacement of composite femurs with three types of single distal screw nails (10 interlocking nails, 10 compression nails and 10 Mehmet anti-rotation nails compressed by tube) at 6-Nm external− 6 Nm internal and 10 Nm extern– 6 Nm internal torques, which imitating respectively the level walking and descending stairs, using an axial distraction testing machine and a custom designed rotation apparatus.
Results: Between 10 Nm external and 6 Nm internal torques with single distal locking screw, the interfragment rotational displacement in the Mehmet nail compressed by the 8 Nm torque wrench was mean 1.14 mm and 540% less than mean 7.31 mm in interlocking nails and 400% less than 5.72 mm in compression nails compressed by the 2.5 Nm torque wrench.
Conclusion: The single distal screw Mehmet nail is superior than other single (or even some double) distal screw nails for maximum rotational stability, with no interfragment distraction gap, no superior nail migration, decreased operation time, and less radiation exposure in axially stable transverse and short oblique femur fractures in daily activities like stair descending or level walking.
Anahtar Kelimeler
Ayrıntılar
Birincil Dil
İngilizce
Konular
-
Bölüm
-
Yazarlar
Ahmet Adnan Karaarslan
Bu kişi benim
Hakan Aycan
Bu kişi benim
Yaman Örgen
Bu kişi benim
Ahmet Karakaşlı
Bu kişi benim
Fatih Ertem
Bu kişi benim
Erhan Sesli
Bu kişi benim
Yayımlanma Tarihi
16 Ocak 2016
Gönderilme Tarihi
8 Temmuz 2015
Kabul Tarihi
-
Yayımlandığı Sayı
Yıl 2016 Cilt: 50 Sayı: 1