Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry

Cilt: 2011 Sayı: 3 1 Mart 2011
  • Baran Gencer
  • Vuslat Gürlü
PDF İndir
EN TR

Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry

Abstract

Objective: To evaluate factors affecting measurement quality of dynamic contour tonometry (DCT). Material and Methods: Three hundred patients took part in this prospective clinical study. For each patient, measurements for one eye were included in the study. Of the participants, 116 had ocular hypertension, 114 had glaucoma, and 70 had normal eyes. We used automatic refractokeratometry, ultrasonic pachimetry, Goldmann applanation tonometry, and DCT measurements to investigate the effects of age, corneal astigmatism, refractive error, central corneal thickness (CCT), intraocular pressures (IOP), and ocular pulse amplitude (OPA) on DCT measurement quality (Q). Logistic regression analysis was used for statistical analysis. Results: For 159 patients, the DCT Q score was 1 (group 1). For the remaining 141 patients, the Q score was 2–3 (group 2). The following values were determined for the two groups, respectively: mean age 58.9±9.9 and 61.4±10.6; mean corneal astigmatisms −0.78±0.67 and −0.92±0.77 diopter; mean spherical equivalents 0.33±1.87 and −0.01±2.02 diopter; mean CCT 547.0±35.2 and 549±38.7 μm. Measurements with DCT provided mean IOPs of 18.5±3.70 and 21.27±6.19 mm Hg and mean OPA 3.5±1.4 and 3.2±1.2 mm Hg. By logistic regression analysis, we determined that the parameters affecting measurement quality were age, DCT measurement level, and OPA (p=0.015, p=0.000, and p=0.000, respectively). Conclusion: High IOP values, advanced age, and low OPA values can lead to decreased DCT measurement quality.

Keywords

Kaynakça

  1. Whitacre MM, Stein R. Sources of error with use of Goldmann- type tonometers. Surv Ophthalmol 1993;38:1-30. [CrossRef]
  2. Bhan A, Browning AC, Shah S, Hamilton R, Dave D, Dua HS. Ef- fect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer and Tono-Pen. Invest Ophthalmol Vis Sci 2002;43:1389-92.
  3. Li Q, Li M, Fan Z, Wang N. The influence of central corneal thick- ness and corneal curvature and axial length on the measurement of intraocular pressure. Yan Ke Xue Bao 2002;18:176-80.
  4. Brandt JD, Beiser JA, Kass MA, Gordon MO. Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology 2001;108:1779-88. [CrossRef]
  5. Johnson M, Kass MA, Moses RA, Grodzki WJ. Increased corneal thickness simulating elevated intraocular pressure. Arch Ophthal- mol 1978;96:664- 5.
  6. Bron AM, Creuzot-Garcher C, Goudeau-Boutillon S, d’Athis P. Falsely elevated intraocular pressure due to increased central cor- neal thickness. Graefes Arch Clin Exp Ophthalmol 1999;237:220-4. [CrossRef]
  7. Kanngiester HE, Nee M, Kniestedt C, et al. Simulation of dynamic contour tonometry compared to in-vitro study revealing minimal influence of corneal radius and astigmatism. The theoretical foundations of dynamic contour tonometry. Invest Ophthalmol Vis Sci 2003;44:2641.
  8. Punjabi OS, Kniestedt C, Stamper RL, Lin SC. Dynamic con- tour tonometry: principle and use. Clin Experiment Ophthalmol 2006;34:837-40. [CrossRef]

Ayrıntılar

Birincil Dil

İngilizce

Konular

Sağlık Kurumları Yönetimi

Bölüm

Araştırma Makalesi

Yazarlar

Baran Gencer Bu kişi benim

Vuslat Gürlü Bu kişi benim

Yayımlanma Tarihi

1 Mart 2011

Gönderilme Tarihi

6 Ağustos 2014

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2011 Cilt: 2011 Sayı: 3

Kaynak Göster

APA
Gencer, B., & Gürlü, V. (2011). Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry. Balkan Medical Journal, 2011(3), 229-231. https://doi.org/10.5174/tutfd.2010.03504.2
AMA
1.Gencer B, Gürlü V. Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry. Balkan Medical Journal. 2011;2011(3):229-231. doi:10.5174/tutfd.2010.03504.2
Chicago
Gencer, Baran, ve Vuslat Gürlü. 2011. “Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry”. Balkan Medical Journal 2011 (3): 229-31. https://doi.org/10.5174/tutfd.2010.03504.2.
EndNote
Gencer B, Gürlü V (01 Mart 2011) Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry. Balkan Medical Journal 2011 3 229–231.
IEEE
[1]B. Gencer ve V. Gürlü, “Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry”, Balkan Medical Journal, c. 2011, sy 3, ss. 229–231, Mar. 2011, doi: 10.5174/tutfd.2010.03504.2.
ISNAD
Gencer, Baran - Gürlü, Vuslat. “Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry”. Balkan Medical Journal 2011/3 (01 Mart 2011): 229-231. https://doi.org/10.5174/tutfd.2010.03504.2.
JAMA
1.Gencer B, Gürlü V. Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry. Balkan Medical Journal. 2011;2011:229–231.
MLA
Gencer, Baran, ve Vuslat Gürlü. “Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry”. Balkan Medical Journal, c. 2011, sy 3, Mart 2011, ss. 229-31, doi:10.5174/tutfd.2010.03504.2.
Vancouver
1.Baran Gencer, Vuslat Gürlü. Evaluation of Parameters Affecting Measurement Quality of Dynamic Contour Tonometry. Balkan Medical Journal. 01 Mart 2011;2011(3):229-31. doi:10.5174/tutfd.2010.03504.2