Objectives: Nosocomial outbreaks of infection originating from ultrasound probes and contaminated coupling gels have been reported. It was reported that the ultrasound probe, if cultured after routine scanning of intact skin, may become colonized with skin flora.

Methods: Culture swabs from 22 probes of the 9 ultrasound machines and from the gels in the 10 gel folders were taken. All swabs taken from probe head, probe holder and the coupling gel in the folder at the beginning of the day were cultured. After fifth scanning and after wiping off the gel with a dry, nonsterile paper towel, cultures were again obtained from probe head and probe holder.

Results: A total of 98 culture results were included of which 42.8% were positive for bacterial growth. The rate of bacterial contamination from probes at morning before the start of examination and after scanning were 34.1% and 56.8%, respectively and this difference was statistically significant (p = 0.023).

Conclusions: We think that using nonsterile, dry, soft and absorbent paper towel after each procedure, could be inadequate for disinfection of probe head. Especially, good hand hygiene could decrease the rate of growth of bacterial colony at probe handle. 

Keywords

References

  1. [1] Ducell G, Fabry J, Nıcoll L. Prévention des infections nosocomiales, Guide pratique, 2nd ed., Organisation mondiale de la Santé. WHO. 2008.
  2. [2] Brachman PS. Epidemiology infections. In: Bennett JV, Brachmann PS, eds. Hospital infections, 3rd ed. Boston: Little, Brown, 1993:3-20.
  3. [3] Tikhomirov E. WHO Programme for the Control of Hospital Infections. Chemiotherapia 1987;3:148-51.
  4. [4] Bures S, Fishbain JT, Uyehara CF, Parker JM, Berg BW. Computer keyboards and faucet handles as reservoirs of nosocomial pathogens in the intensive care unit. Am J Infect Control 2000;28:465-71.
  5. [5] Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med 1993;118:117-28.
  6. [6] Steinberg PJ, de Hoop D. [The stethoscope as a vehicle of pathogenic microorganisms in the hospital]. Ned Tijdschr Geneeskd 1978;122:303-5. [Article in Dutch]
  7. [7] Livornese LL Jr, Dias S, Samel C, Romanowski B, Taylor S, May P, et al. Hospital-acquired infection with vancomycin-resistant Enterococcus faecium transmitted by electronic thermometers. Ann Intern Med 1992;117:112-6.
  8. [8] Gaillot O, Maruéjouls C, Abachin E, Lecuru F, Arlet G, Simonet M, et al. Nosocomial outbreak of Klebsiella pneumoniae producing SHV-5 extended-spectrum beta-lactamase, originating from a contaminated ultrasonography coupling gel. J Clin Microbiol 1998;36:1357-60.

Details

Primary Language

English

Subjects

Health Care Administration

Journal Section

Research Article

Authors

Gürkan Kıran This is me
0000-0002-6300-328X
Türkiye

Deniz Cemgil Arıkan This is me
0000-0001-8322-0209
Türkiye

Uğurkan Erkayıran
kahramanmaraş sütçü imam university
0000-0002-8519-1883
Türkiye

Hasan Çetin Ekerbiçer This is me
0000-0003-0064-3893
Türkiye

Publication Date

July 4, 2019

Submission Date

March 5, 2018

Acceptance Date

August 9, 2018

Published in Issue

Year 1970 Volume: 5 Number: 4

AMA
1.Kıran H, Aral M, Kıran G, et al. Bacterial contamination of ultrasound probes and coupling gels in a university hospital in Turkey. Eur Res J. 2019;5(4):582-587. doi:10.18621/eurj.401327

Cited By