Effect of Acquisition Parameters of SPECT Gamma Camera System on Image Quality
Öz
OBJECTIVE: Quality control testing of a SPECT gamma camera is crucial in assessing the suitability of the camera for use in nuclear medicine department. The aim of the present study was to investigate the effect of gamma camera acquisition parameters on image quality.
METHODS: Camera scanning was carried out using a double-headed gamma cameraon a total of 48 patients (29 female and 19 male, mean age: 47.4±11.1) referred to our department for myocardial perfusion scintigraphy and thyroid scintigraphy. Then, camera acquisition parameters were changed (for myocardial perfusion scintigraphy from 64x64 to 128x128 matrix change and for thyroid scintigraphy from 5 cm to 10 cm distance change), and scanning was repeated and images were analyzed.
RESULTS: Left ventricle ejection fraction (EF) value in 64x64 matrix was calculated to be 62.7±8.8%. Smaller EF value of 48.9±10.3% was obtained in 128x128 matrix for the same patients. 99mTc-pertechnetate uptake percentage was 3.8±2.3% in measurements carried out at a distance of 10 cm. On the other hand, a higher uptake percentage of 6.2±3.6% was found for the same patients measured at a distance of 5 cm.
CONCLUSION:
In order to obtain proper imaging in SPECT gamma camera system, correct
acquisition parameters should be used along with quality control tests for
intrinsic flood-field uniformity and relative sensitivity.
Anahtar Kelimeler
Kaynakça
- Fontana M, Dauvergne D, Letang JM, Ley JL, Testa E. Compton camera study for high efficiency SPECT and benchmark with Anger system. Phys Med Biol 2017;62(23):8794-8812.
- Jha AK, Mithun S, Chauhan MH, Purandare N, Shah S, Agrawal A, et al. A Novel 141Ce-Based Flood Field Phantom: Assessment of Suitability for Daily Uniformity Testing in a Clinical Nuclear Medicine Department. J Nucl Med Technol 2017;45(3):225-9.
- Tindale WB. Specifying dual-detector gamma cameras and associated computer systems. Nucl Med Commun 1995;16(7):534-8.
- Elkamhawy AA, Rothenbach JR, Damaraju S, Badruddin SM. Intrinsic uniformity and relative sensitivity quality control tests for single-head gamma cameras. J Nucl Med Technol 2000;28(4):252-6.
- O’Connor MK. Instrument- and computer-related problems and artifacts in nuclear medicine. Semin Nucl Med 1996;26(4):256-77.
- Blokland JA, Camps JA, Pauwels EK. Aspects of performance assessment of whole body imaging systems. Eur J Nucl Med 1997;24(10):1273-83.
- Kappadath SC, Erwin WD, Wendt RE. Observed inter-camera variability of clinically relevant performance characteristics for Siemens Symbia gamma cameras. J Appl Clin Med Phys 2006;7(4):74-80.
- Peterson TE, Furenlid LR. SPECT detectors: the Anger Camera and beyond. Phys Med Biol 2011;56(17):145-82.
Ayrıntılar
Birincil Dil
İngilizce
Konular
Sağlık Kurumları Yönetimi
Bölüm
Araştırma Makalesi
Yazarlar
Yayımlanma Tarihi
30 Eylül 2018
Gönderilme Tarihi
30 Mart 2018
Kabul Tarihi
21 Haziran 2018
Yayımlandığı Sayı
Yıl 2018 Cilt: 8 Sayı: 3