Araştırma Makalesi

Investigation From Different Calculation Solutions In Vertebra Plans And The Reliability Of Different Dose Rate Dose

Cilt: 5 Sayı: 3 30 Kasım 2019
PDF İndir
EN

Investigation From Different Calculation Solutions In Vertebra Plans And The Reliability Of Different Dose Rate Dose

Abstract

In this study, it was aimed to investigate the effect of different dose rate and different computational resolutions on stereotactic vertebrae irradiation with the results of the verification schemes on Gamma analysis and the effect on dose for critical organ and target volume. Ten patients with vertebrae participated in the study. Patients participating in the study were investigated for their effect on doses for critical organ and target volume by recalculating treatment plans at different dose rates and different computational resolutions without changing the optimization parameters using Intensity Modulated Radiation Therapy (IMRT) technique in the Eclipse treatment planning system with Varian Trilogy Linear Accelerator located in Şişli Hamidiye Etfal Training and Research Hospital , Radiation Oncology Department. Verification schedules of all the treatment plans The results were compared by using the Electronic Portal Imaging Device (EPID). Using different dose rates did not affect our treatment planning accuracy. However, it was seen that the high resolution calculated plans represented the target and critical organ doses better.

Keywords

Kaynakça

  1. [1] M.J. Murphy, et. al., The management of imaging dose during image-guided radiotherapy: Report of the AAPM Task Group 75(2007)
  2. [2] D. Verellen, M. De Ridder, G. Storme, A history of image guided radiotherapy. 4-13(2008), DOI:10.1016/j.radonc.2007.11.023
  3. [3] S.P Collins, N.D. Coppa,Y. Zhang, B.T. Collins, D.A. McRae, W.C. Jean, CK Radiosurgery in the treatment of Complex Skull Base Tumors: Analysis of Planning Parameters.Radiation Oncology. 1(46):1-10(2006)
  4. [4] J.M. Galvin, G. Bednarz, Quality Assurance Procedures for Stereotactic Body Radiation Therapy. Int J Radiat Oncol Biol Phys. 71(1): 122-125(2008), DOI:10.1016/j.ijrobp.2007.10.002
  5. [5] P.C. Gerszten, C. Ozhasoglu, S.A. Burton, et al., CyberKnife frameless stereotactic radiosurgery for spinal lesions: clinical experience in 125 cases. Neurosurgery. 55(1):89–98(2004).
  6. [6] J.W. Degen, G.J. Gagnon, J.M. Voyadzis, et al., CyberKnife stereotactic radiosurgical treatment of spinal tumors for pain control and quality of life. J Neurosurgery Spine. 2(5):540–49(2005), DOI:10.3171/spi.2005.2.5.0540
  7. [7] Y. Lemoigne, Caner A. Radiotherapy and Brachytherapy. Springer, Netherlands, 2009
  8. [8] F.M. Khan,. The Physics of Radiation Therapy 3rd Edition. Lippincott Williams & Wilkins Company, USA, 2010.

Ayrıntılar

Birincil Dil

İngilizce

Konular

Klinik Tıp Bilimleri

Bölüm

Araştırma Makalesi

Yayımlanma Tarihi

30 Kasım 2019

Gönderilme Tarihi

29 Temmuz 2019

Kabul Tarihi

26 Eylül 2019

Yayımlandığı Sayı

Yıl 2019 Cilt: 5 Sayı: 3

Kaynak Göster

APA
Çelen, Y. Y., & Kızılkaya, H. O. (2019). Investigation From Different Calculation Solutions In Vertebra Plans And The Reliability Of Different Dose Rate Dose. International Journal of Computational and Experimental Science and Engineering, 5(3), 135-138. https://doi.org/10.22399/ijcesen.597819