Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use
Öz
Aim: In endodontic treatment, nickel-titanium (NiTi) instruments, revolutionized root canal instrumentation owing to their superior flexibility, resistance to torsional fatigue, and shape memory. Among contemporary NiTi systems, ProTaper Ultimate (PU) (Dentsply Sirona, Ballaigues, Switzerland) and Fanta AF Rising (AF-R) (Fanta Dental, Shanghai, China) are widely utilized, yet differ consid-erably in terms of their metallurgical composition, design, and clinical performance. Although both systems are commonly used in clinical practice, there is a paucity of comparative data evaluating their canal centering ability, apical transportation, post-sterilization performance, and structural integrity after multiple uses. This study aimed to compare the shaping performance, apical transportation, and structural integrity of two thermally treated NiTi rotary file systems (AF-R and PU) after repeated use and sterilization cycles in standardized simulated curved canals.
Material and Methods: Sixty standardized resin blocks with 45° curvature were prepared using either AF-R or PU instruments. Each group was divided into 3 subgroups based on the number of uses and sterilizations (1st, 2nd, and 3rd cycle). Canal centering and transportation were evaluated at 5 reference levels using high-resolution digital overlays and ImageJ software. File deformation and separation were assessed under stereomicroscopy. Statistical analysis was performed using the Kruskal–Wallis and Bonferroni tests, with P < 0.05 considered statistically significant.
Results: AF-R demonstrated significantly superior canal centering (P = 0.008) and lower apical transportation (P = 0.004) across all refer- ence levels and usage cycles. No file separations occurred in the AF-R group, whereas PU exhibited progressive deformation and a 6.7% separation rate by the third cycle. AF-R instruments retained their original shape and mechanical behavior after autoclave sterilization.
Conclusion: AF-R exhibited enhanced shaping precision, structural durability, and resistance to apical transportation compared to PU, especially after multiple clinical uses and sterilization. These findings support its suitability for minimally invasive and cost-effec- tive endodontic practice, particularly in anatomically complex canals.
Anahtar Kelimeler
Destekleyen Kurum
None
Etik Beyan
The authors declare that this study does not carry the qualification of experiments on humans and animals, biological materials (biological fluids such as blood, urine, extracted human teeth and tissue samples, etc.) observational and descriptive research without intervention (questionnaire, scale file scans, system model development, audio and video recordings.)
Kaynakça
- Peters OA. Current challenges and concepts in the preparation of root canal systems: a review. J Endod. 2004;30:559–67.
- Lopes HP, Gambarra-Soares T, Elias CN, et al. Comparison of the mechanical properties of rotary instruments made of conventional nickel–titanium wire, M–Wire, or nickel–titanium alloy in R–phase. J Endod. 2013;39:516–20.
- Schäfer E, Schlingemann R. Efficiency of rotary nickel–titanium FlexMaster instruments compared with stainless steel hand K-Flexofile. Part 1. Shaping ability in simulated curved canals. Int Endod J. 2003;36:199–207.
- Al-Sudani D, Al-Shahrani S. A comparison of the canal centering ability of ProFile, K3, and RaCe nickel titanium rotary systems. J Endod. 2006;32:1198–201.
- Zhao D, Shen Y, Peng B, Haapasalo M. Micro–computed tomography evaluation of the preparation of mesiobuccal root canals in maxillary first molars with HyFlex CM, Twisted Files, and K3 instruments. J Endod. 2013;39:385–88.
- Gergi R, Osta N, Naaman A. Comparison of canal transpor- tation and centering ability of Twisted Files, Pathfile–ProTaper system, and stainless steel hand K–Files by using computed tomography. J Endod. 2010;36:904–7.
- Gambarini G. Cyclic fatigue of nickel–titanium rotary instruments after clinical use with low and high torque endodontic motors. J Endod. 2001;27:772–74.
- Hieawy A, Haapasalo M, Zhou H, et al. Phase transformation behavior of ProTaper Gold and ProTaper Universal instruments. J Endod. 2015;41:716–29.
Ayrıntılar
Birincil Dil
İngilizce
Konular
Endodonti
Bölüm
Klinik Araştırma
Yayımlanma Tarihi
22 Şubat 2026
Gönderilme Tarihi
25 Mayıs 2025
Kabul Tarihi
4 Ağustos 2025
Yayımlandığı Sayı
Yıl 2026 Sayı: 1
APA
Övsay, E., Atav, A., & Mustafa Pamuk, D. (2026). Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use. Medical Records, 1. https://doi.org/10.37990/medr.1705898
AMA
1.Övsay E, Atav A, Mustafa Pamuk D. Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use. Med Records. 2026;(1). doi:10.37990/medr.1705898
Chicago
Övsay, Emre, Ayfer Atav, ve Duygu Mustafa Pamuk. 2026. “Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use”. Medical Records, sy 1. https://doi.org/10.37990/medr.1705898.
EndNote
Övsay E, Atav A, Mustafa Pamuk D (01 Şubat 2026) Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use. Medical Records 1
IEEE
[1]E. Övsay, A. Atav, ve D. Mustafa Pamuk, “Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use”, Med Records, sy 1, Şub. 2026, doi: 10.37990/medr.1705898.
ISNAD
Övsay, Emre - Atav, Ayfer - Mustafa Pamuk, Duygu. “Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use”. Medical Records. 1 (01 Şubat 2026). https://doi.org/10.37990/medr.1705898.
JAMA
1.Övsay E, Atav A, Mustafa Pamuk D. Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use. Med Records. 2026. doi:10.37990/medr.1705898.
MLA
Övsay, Emre, vd. “Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use”. Medical Records, sy 1, Şubat 2026, doi:10.37990/medr.1705898.
Vancouver
1.Emre Övsay, Ayfer Atav, Duygu Mustafa Pamuk. Beyond the Curve: Evaluating Performance and Integrity of Modern NiTi Systems Under Repeated Use. Med Records. 01 Şubat 2026;(1). doi:10.37990/medr.1705898