Literature Review on Pacemaker Lead Dislodgement Complexes: The experience of a Tertiary Center with Eight Cases
Öz
Cardiac pacemakers
(CPM) and implantable cardioverter defibrillators (ICD) are indispensable
treatment option for the cardiac arrhythmias. The most common complication of
pacemaker implantation is lead
dislodgement. We will present eight patients with lead dislodgement about
Twiddler, Reel and Ratchet Syndromes. The patients who need lead repositioning
for lead dislodgement were collected retrospectively at the Cardiology Clinic
of İzmir Katip Celebi University Ataturk Training and Research Hospital from 2016
to 2017. Lead dislodgement is mostly detected at the device interrogation, with
worsening insensing voltage or pacing thresholds. The dislodgement of a
ventricular lead incidence in a CPM is known as 0.5 to 2.5%. Our center's
experience is lower than these rates (<%0.03). Patient education, opening
appropriate pocket for generator size,
fixing the sleeves tight enough as required,
hanging battery with a strong suture can be used to avoid lead
dislodgement.
Anahtar Kelimeler
Kaynakça
- 1. Link MS, Estes NA 3rd, Griffin JJ, et al. Complications of dual chamber pacemaker implantation in the elderly. Pacemaker Selection in the Elderly (PASE) Investigators. J Interv Card Electrophysiol. 1998;2(2):175–9.
- 2. Chauhan A, Grace AA, Newell SA et al. Early complications after dual chamber versus single chamber pacemaker implantation. Pacing Clin Electrophysiol. 1994;17(2):2012–5.
- 3. Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK, Alexander ME. Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease. Heart Rhythm. 2004;1(2):150–9.
- 4. Bayliss CE, Beanlands DS, Baird RJ. The pacemaker-Twiddler’s syndrome: a new complication of implantable transvenous pacemakers. Can Med Assoc J. 1968;99(8):371–3.
- 5. Carnero-Varo A, Perez-Paredes M, Ruiz-Ros JA et al. “Reel Syndrome”: a new form of Twiddler’s syndrome? Circulation. 1999;100(8):e45– 6.
- 6. Von Bergen NH. "Ratchet" syndrome, another etiology for pacemaker lead dislodgement: a case report. Heart Rhythm. 2007;4(6):788-9.
- 7. Alvarez-Acosta L, Romero Garrido R, Farrais-Villalba M, Hernández Afonso J.Reel syndrome: a rare cause of pacemaker malfunction. BMJ case reports. 2014.
- 8. Ghani A, Delnoy PP, Ramdat Misier AR et al. Incidence of lead dislodgement, malfunction and perforation during the first year following device implantation. Neth Heart J. 2014;22(6):286-91.