Case of Malleable Penile Prosthesis Fracture Possibly Due to Wrong Surgical Technique

Yıl 2013, Cilt: 20 Sayı: 3, - , 01.06.2013
https://izlik.org/JA73RN58FE

Öz

Mechanical fracture is an unusual complication of malleable penile prosthesis. In the present case, we reported that a 29-year-old man, using the AMS 650 malleable penile prosthesis since 2004, was presented. The patient described that there was deformity at the right side of the penis. Pain and difficulty during sexual intercourse was described. Right root of the penile prosthesis was found to be fractured in the physical examination and then it was removed by surgical exploration. It was determined that the right root of the penile prosthesis was misplaced on front-backward plane. A new AMS 650 malleable penile prosthesis was implanted. Because of previous prosthesis was shorter than needed, a new prosthesis was used 3 cm longer than the previous. The patient was discharged a day after the operation without any complication. Key words: Penile Prosthesis; Mechanical Fracture; Surgically Misplacement.

Kaynakça

  • For citing/Atıf için: Correspondence/İletişim

Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu

Yıl 2013, Cilt: 20 Sayı: 3, - , 01.06.2013
https://izlik.org/JA73RN58FE

Öz

Malleable penil protezin mekanik olarak kırılması çok nadir görülen ve beklenmedik bir komplikasyondur. Burada 2004 yılından beri komplikasyonsuz AMS 650 malleable penil protez kullanan 29 yaşında bir erkek hasta sunuldu. Hasta penis sağ yan kısımda deformite, cinsel ilişkide ağrı ve güçlük sikayeti ile başvurdu. Fizik muayenede penil protezin sağ kolunun kırık olduğu tespit edildi. Penil protez cerrahi işlemle çıkarıldı. Penil protezin sağ kolunun ön-arka düzlemde yanlış yerleştirilmiş olduğu fark edildi. Hastaya yeni AMS 650 malleable penil protez takıldı. Çıkarılan protez gereğinden daha kısa olduğu için yeni takılan protez 3 cm daha uzundu. Hastada hiçbir komplikasyon olmaksızın ameliyat sonrası gün taburcu edildi. Anahtar kelimeler: Penil Protez; Mekanik Fraktür; Cerrahi Yerleşim Kusurları.

Kaynakça

  • For citing/Atıf için: Correspondence/İletişim
Toplam 1 adet kaynakça vardır.

Ayrıntılar

Birincil Dil Türkçe
Yazarlar

Fatih Oğuz Bu kişi benim

Ahmet Soylu Bu kişi benim

Ramazan Altıntaş Bu kişi benim

Ali Beytur Bu kişi benim

Aliseydi Bozkurt Bu kişi benim

Yayımlanma Tarihi 1 Haziran 2013
IZ https://izlik.org/JA73RN58FE
Yayımlandığı Sayı Yıl 2013 Cilt: 20 Sayı: 3

Kaynak Göster

APA Oğuz, F., Soylu, A., Altıntaş, R., Beytur, A., & Bozkurt, A. (2013). Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu. Journal of Turgut Ozal Medical Center, 20(3). https://izlik.org/JA73RN58FE
AMA 1.Oğuz F, Soylu A, Altıntaş R, Beytur A, Bozkurt A. Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu. J Turgut Ozal Med Cent. 2013;20(3). https://izlik.org/JA73RN58FE
Chicago Oğuz, Fatih, Ahmet Soylu, Ramazan Altıntaş, Ali Beytur, ve Aliseydi Bozkurt. 2013. “Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu”. Journal of Turgut Ozal Medical Center 20 (3). https://izlik.org/JA73RN58FE.
EndNote Oğuz F, Soylu A, Altıntaş R, Beytur A, Bozkurt A (01 Haziran 2013) Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu. Journal of Turgut Ozal Medical Center 20 3
IEEE [1]F. Oğuz, A. Soylu, R. Altıntaş, A. Beytur, ve A. Bozkurt, “Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu”, J Turgut Ozal Med Cent, c. 20, sy 3, Haz. 2013, [çevrimiçi]. Erişim adresi: https://izlik.org/JA73RN58FE
ISNAD Oğuz, Fatih - Soylu, Ahmet - Altıntaş, Ramazan - Beytur, Ali - Bozkurt, Aliseydi. “Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu”. Journal of Turgut Ozal Medical Center 20/3 (01 Haziran 2013). https://izlik.org/JA73RN58FE.
JAMA 1.Oğuz F, Soylu A, Altıntaş R, Beytur A, Bozkurt A. Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu. J Turgut Ozal Med Cent. 2013;20. Available at https://izlik.org/JA73RN58FE.
MLA Oğuz, Fatih, vd. “Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu”. Journal of Turgut Ozal Medical Center, c. 20, sy 3, Haziran 2013, https://izlik.org/JA73RN58FE.
Vancouver 1.Fatih Oğuz, Ahmet Soylu, Ramazan Altıntaş, Ali Beytur, Aliseydi Bozkurt. Yanlış Cerrahi Uygulama Nedeniyle Malleable Penil Protezin Kırılması: Olgu Sunumu. J Turgut Ozal Med Cent [Internet]. 01 Haziran 2013;20(3). Erişim adresi: https://izlik.org/JA73RN58FE