HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY

Abstract

In this study, it was aimed that evaluate the hemolysis effect on ethanol assay.

In this study, about 10 and 40 mg/dL concentration ethanol containing samples was prepared with ethanol calibrator (92.7 mg/dL) and evaluated the effect of hemolysis. For providing similar matrix, 10 ethanol-free normal patients’ serums were pooled and used for diluations. Hemoglobin was isolated from fresh blood. Evaluation was performed in 5 levels of hemolysis (about 0, 250, 500, 750, 1000 mg/dL).

It has been shown to vary that about 10 mg/dL ethanol at 5 hemolysis levels 10.3, 10.5, 12.3, 14, 16.6 respectively. (According to the first reading percent differences are 1.84, 19.42, 35.92, 61,17 respectively). It has been shown to vary that about 10 mg/dL ethanol at 5 hemolysis levels 41.6, 41.6, 41.2, 43.9, 44.8 respectively (according to the first reading percent differences are 0, -0,96, 5.53, 7.69 respectively).

Most of the methods currently on the market use the same enzymatic reaction (alcohol dehydrogenase). Therefore, a similar interference may occur in devices from other brands. High rate of interference, especially in low level of ethanol, may be associated with the analytical sensitivity.

Hemolysis samples should be carefully considered to not experience legal troubles.

Keywords

Kaynakça

  1. Metin GENÇ, Özlem DOĞAN, Zihni KARAEREN, Aslıhan AVCI
  2. Ankara University, Faculty of Medicine, Department of Medical Biochemistry Ankara University, Ibni Sina Hospital, Central Laboratory

Ayrıntılar

Birincil Dil

İngilizce

Konular

-

Bölüm

-

Yazarlar

Metin Genç Bu kişi benim

Özlem Doğan Bu kişi benim

Zihni Karaeren Bu kişi benim

Aslıhan Avcı Bu kişi benim

Yayımlanma Tarihi

16 Şubat 2017

Gönderilme Tarihi

17 Nisan 2017

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2017 Cilt: Volume 2 Sayı: İssue 1 (1)

Kaynak Göster

APA
Genç, M., Doğan, Ö., Karaeren, Z., & Avcı, A. (2017). HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY. The Turkish Journal Of Occupational / Environmental Medicine and Safety, Volume 2(İssue 1 (1), 216-216. https://izlik.org/JA85NW29RM
AMA
1.Genç M, Doğan Ö, Karaeren Z, Avcı A. HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY. turjoem. 2017;Volume 2(İssue 1 (1):216-216. https://izlik.org/JA85NW29RM
Chicago
Genç, Metin, Özlem Doğan, Zihni Karaeren, ve Aslıhan Avcı. 2017. “HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY”. The Turkish Journal Of Occupational / Environmental Medicine and Safety Volume 2 (İssue 1 (1): 216-16. https://izlik.org/JA85NW29RM.
EndNote
Genç M, Doğan Ö, Karaeren Z, Avcı A (01 Şubat 2017) HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY. The Turkish Journal Of Occupational / Environmental Medicine and Safety Volume 2 İssue 1 (1) 216–216.
IEEE
[1]M. Genç, Ö. Doğan, Z. Karaeren, ve A. Avcı, “HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY”, turjoem, c. Volume 2, sy İssue 1 (1), ss. 216–216, Şub. 2017, [çevrimiçi]. Erişim adresi: https://izlik.org/JA85NW29RM
ISNAD
Genç, Metin - Doğan, Özlem - Karaeren, Zihni - Avcı, Aslıhan. “HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY”. The Turkish Journal Of Occupational / Environmental Medicine and Safety VOLUME 2/İssue 1 (1) (01 Şubat 2017): 216-216. https://izlik.org/JA85NW29RM.
JAMA
1.Genç M, Doğan Ö, Karaeren Z, Avcı A. HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY. turjoem. 2017;Volume 2:216–216.
MLA
Genç, Metin, vd. “HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY”. The Turkish Journal Of Occupational / Environmental Medicine and Safety, c. Volume 2, sy İssue 1 (1), Şubat 2017, ss. 216-, https://izlik.org/JA85NW29RM.
Vancouver
1.Metin Genç, Özlem Doğan, Zihni Karaeren, Aslıhan Avcı. HEMOLYSIS INTERFERENCE ON ETHANOL ASSAY. turjoem [Internet]. 01 Şubat 2017;Volume 2(İssue 1 (1):216-. Erişim adresi: https://izlik.org/JA85NW29RM