Abstract

According to ISO uncertainity is a parameter associated with the result of measurement that characterizes the dispersion of the values that could reasonably be attributed to the measurand. The aim of this study is to calculate measurement uncertainty of valproic acid by using internal quality control data and inter laboratory results to compare these calculated measurement uncertainties with total allowable error % (TEa %) value of CLIA'88.

Between October 2015 and March 2016, the internal quality control datas were screened. The estimation of valproic acid measurement uncertainity was calculated according to ISO21748 guide. Inter laboratory comparison was executed to estimate bias. Valproic acid levels were determined using the immunoturbidimetric method on the Roche Cobas Integra 800 analyzer.

Coefficient variations (CV) calculated from internal quality control analyses were 2.12, 2.46 and 3.01 at low, normal and high level of internal quality control sera values respectively. Total allowable error value of valproic acid was 9.94% and was not higher than desirable TEa% values of CLIA'88.

Valproic acid is a widely-used first-generation antiepileptic drug, therapeutic range between 50-100 µg/mL, prescribed predominantly in migraine prophylaxis, epilepsy and psychiatric disorders. Careful clinical monitoring should be performed during the first 6 months of therapy with valproic acid. For this purpose, using uncertainty value is a good practice to interpret of patient result.

Laboratories should calculate measurement uncertainity and evaluate results according to spesific criterias. 

Keywords

Kaynakça

  1. Gamze AVCİOGLU, Serpil ERDOGAN, Ceylan BAL, Orhan SEN, Pervin BARAN, Ozcan EREL
  2. Ataturk Training and Research Hospital, Department of Clinical Biochemistry, Ankara, Turkey Yildirim Beyazit University Faculty of Medicine, Department of Public Health, Ankara, Turkey

Ayrıntılar

Birincil Dil

İngilizce

Konular

-

Bölüm

-

Yazarlar

Gamze Avcioglu Bu kişi benim

Serpil Erdogan Bu kişi benim

Ceylan Bal Bu kişi benim

Pervin Baran Bu kişi benim

Ozcan Erel Bu kişi benim

Yayımlanma Tarihi

16 Şubat 2017

Gönderilme Tarihi

11 Mayıs 2017

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2017 Cilt: Volume 2 Sayı: İssue 1 (1)

Kaynak Göster

APA
Avcioglu, G., Erdogan, S., Bal, C., Sen, O., Baran, P., & Erel, O. (2017). MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS. The Turkish Journal Of Occupational / Environmental Medicine and Safety, Volume 2(İssue 1 (1), 258-258. https://izlik.org/JA83RR27PJ
AMA
1.Avcioglu G, Erdogan S, Bal C, Sen O, Baran P, Erel O. MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS. turjoem. 2017;Volume 2(İssue 1 (1):258-258. https://izlik.org/JA83RR27PJ
Chicago
Avcioglu, Gamze, Serpil Erdogan, Ceylan Bal, Orhan Sen, Pervin Baran, ve Ozcan Erel. 2017. “MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS”. The Turkish Journal Of Occupational / Environmental Medicine and Safety Volume 2 (İssue 1 (1): 258-58. https://izlik.org/JA83RR27PJ.
EndNote
Avcioglu G, Erdogan S, Bal C, Sen O, Baran P, Erel O (01 Şubat 2017) MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS. The Turkish Journal Of Occupational / Environmental Medicine and Safety Volume 2 İssue 1 (1) 258–258.
IEEE
[1]G. Avcioglu, S. Erdogan, C. Bal, O. Sen, P. Baran, ve O. Erel, “MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS”, turjoem, c. Volume 2, sy İssue 1 (1), ss. 258–258, Şub. 2017, [çevrimiçi]. Erişim adresi: https://izlik.org/JA83RR27PJ
ISNAD
Avcioglu, Gamze - Erdogan, Serpil - Bal, Ceylan - Sen, Orhan - Baran, Pervin - Erel, Ozcan. “MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS”. The Turkish Journal Of Occupational / Environmental Medicine and Safety VOLUME 2/İssue 1 (1) (01 Şubat 2017): 258-258. https://izlik.org/JA83RR27PJ.
JAMA
1.Avcioglu G, Erdogan S, Bal C, Sen O, Baran P, Erel O. MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS. turjoem. 2017;Volume 2:258–258.
MLA
Avcioglu, Gamze, vd. “MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS”. The Turkish Journal Of Occupational / Environmental Medicine and Safety, c. Volume 2, sy İssue 1 (1), Şubat 2017, ss. 258-, https://izlik.org/JA83RR27PJ.
Vancouver
1.Gamze Avcioglu, Serpil Erdogan, Ceylan Bal, Orhan Sen, Pervin Baran, Ozcan Erel. MEASUREMENT UNCERTAINITY IN VALPROIC ACID CONCENTRATIONS. turjoem [Internet]. 01 Şubat 2017;Volume 2(İssue 1 (1):258-. Erişim adresi: https://izlik.org/JA83RR27PJ