Abstract

Measurement uncertainty is a quality parameter for the accuracy of the measurement results in clinical testing. It shows what extent the result represents the real value. Measurement uncertainty is very important for the clinician to provide medical decision, diagnosis and treatment. The purpose of this study is to calculate measurement uncertainty of phenytoin by using internal quality control datas, inter laboratory comparison results and calibrator uncertainty value to compare it with CLIA total error % (TEa%) value.

Between march 2015 and March 2016, the internal quality control datas were screened. The estimation of phenytoin measurement was calculated according to ISO21748 guide. Inter laboratory comparison and calibrator uncertainty values were used to estimate bias. Phenytoin levels were determined using the immunoturbidimetric method on the Roche Cobas Integra 800 analyzer.

TEa% values for phenytoin was 19,47. This value was not higher than CLIA (%25).

Discussion: Phenytoin is primary antiepileptic agent in all types of epilepsy except absence epilepsy Dosage adjustment is important for assessment treatment efficacy in therapeutic range and also protection from the negative health effects in toxic levels. Therefore, medical laboratories should produce analytically reliable results for the correct interpretation and use of results.

Laboratories should give results lower than TEa% limits for phenytoin.

Keywords

Kaynakça

  1. Cemil NURAL, Serpil ERDOGAN , Ceylan BAL , Orhan SEN , Pervin BARAN , Ozcan EREL
  2. Ataturk Training and Research Hospital, Department of Clinical Biochemistry, Ankara, Turkey Yildirim Beyazit University Faculty of Medicine, Department of Public Health, Ankara, Turkey

Ayrıntılar

Birincil Dil

İngilizce

Konular

-

Bölüm

-

Yazarlar

Cemil Nural Bu kişi benim

Serpil Erdogan Bu kişi benim

Ceylan Bal Bu kişi benim

Pervin Baran Bu kişi benim

Ozcan Erel Bu kişi benim

Yayımlanma Tarihi

16 Şubat 2017

Gönderilme Tarihi

18 Mayıs 2017

Kabul Tarihi

-

Yayımlandığı Sayı

Yıl 2017 Cilt: Volume 2 Sayı: İssue 1 (1)

Kaynak Göster

APA
Nural, C., Erdogan, S., Bal, C., Sen, O., Baran, P., & Erel, O. (2017). EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN. The Turkish Journal Of Occupational / Environmental Medicine and Safety, Volume 2(İssue 1 (1), 266-266. https://izlik.org/JA28FW27FW
AMA
1.Nural C, Erdogan S, Bal C, Sen O, Baran P, Erel O. EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN. turjoem. 2017;Volume 2(İssue 1 (1):266-266. https://izlik.org/JA28FW27FW
Chicago
Nural, Cemil, Serpil Erdogan, Ceylan Bal, Orhan Sen, Pervin Baran, ve Ozcan Erel. 2017. “EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN”. The Turkish Journal Of Occupational / Environmental Medicine and Safety Volume 2 (İssue 1 (1): 266-66. https://izlik.org/JA28FW27FW.
EndNote
Nural C, Erdogan S, Bal C, Sen O, Baran P, Erel O (01 Şubat 2017) EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN. The Turkish Journal Of Occupational / Environmental Medicine and Safety Volume 2 İssue 1 (1) 266–266.
IEEE
[1]C. Nural, S. Erdogan, C. Bal, O. Sen, P. Baran, ve O. Erel, “EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN”, turjoem, c. Volume 2, sy İssue 1 (1), ss. 266–266, Şub. 2017, [çevrimiçi]. Erişim adresi: https://izlik.org/JA28FW27FW
ISNAD
Nural, Cemil - Erdogan, Serpil - Bal, Ceylan - Sen, Orhan - Baran, Pervin - Erel, Ozcan. “EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN”. The Turkish Journal Of Occupational / Environmental Medicine and Safety VOLUME 2/İssue 1 (1) (01 Şubat 2017): 266-266. https://izlik.org/JA28FW27FW.
JAMA
1.Nural C, Erdogan S, Bal C, Sen O, Baran P, Erel O. EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN. turjoem. 2017;Volume 2:266–266.
MLA
Nural, Cemil, vd. “EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN”. The Turkish Journal Of Occupational / Environmental Medicine and Safety, c. Volume 2, sy İssue 1 (1), Şubat 2017, ss. 266-, https://izlik.org/JA28FW27FW.
Vancouver
1.Cemil Nural, Serpil Erdogan, Ceylan Bal, Orhan Sen, Pervin Baran, Ozcan Erel. EVALUATING MEASUREMENT UNCERTAINTY OF PHENYTOIN. turjoem [Internet]. 01 Şubat 2017;Volume 2(İssue 1 (1):266-. Erişim adresi: https://izlik.org/JA28FW27FW