We aimed to
assess to the analytical performance ofEureka LabDivison’s Drugs of Abuse in
urine by LC/MS kit (LC74010)
Imprecision,
linearity, LOD, LOQ, recovery, carryover studies were performed on 24 tests
with an Agilent 6420 mass spectrometer.
Imprecision was
comprised between 0.5 and 6.0%. The linearity for all analytes were excellent.
All r2 values were>0.995.LOD values were < 10 ng/mL and LOQ values were
< 20 ng/mL. Recovery was comprised 95% and 99%. No observed carryover
effects.
The results of
this study demonstrate that our findings were consistent with the kit
manufacturer's data. The absence of the comparison method is insufficient for
this study.
Although we were
not able to do in this study, further studies should be planned for method
comparison experiment.