Araştırma Makalesi

Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı

Cilt: 31 Sayı: ÖS2 15 Ekim 2016
PDF İndir
EN TR

Pull out Performance of Medical Screws used in Orthopaedic Surgeries

Öz

In this study, a new medical bone screw was designed and manufactured by Ti6Al4VELI GR23 material and tested in accordance with the ATSM F 543-02 test procedure experimentally. This test is used to measure the axial tensile force required to fail or remove a bone screw from a defined material. Concurrently, 3-D Finite Element Analysis (FEA) models of these medical bone screws were developed to simulate and biomechanically evaluate the pull-out forces and stiffness of the screw samples using FEA software ANSYS. Obtained results from the FEA models are in reasonable agreement with the experimental data on the pull-out strength of the medical bone screw. Results from FE models are experimentally verified.

Anahtar Kelimeler

Kaynakça

  1. 1. Battula, S., Schoenfeld, A., Vrabec, G., Njus, G. O., 2006. Experimental Evaluation of the Holding Power/Stiffness of the Self-tapping Bone Screws in Normal and Osteoporotic Bone Material, Clinical Biomechanics, 21, pp. 533–537.
  2. 2. Hsu, C.C., Chao, C.K., Wang, J.L., Hou, S.M., Tsai, Y.T., Lin, J., 2005. Increase of Pullout Strength of Spinal Pedicle Screws with Conical Core: Biomechanical Tests and Finite Element Analyses, Journal of Orthopaedic Research, 23, pp. 788–794.
  3. 3. Chapman, J.R., Harrington, R.M., Lee, K.M., Anderson, P.A., Tencer, A.F., Kowalski, D., 1996. Factors Affecting the Pullout Strength of Cancellous Bone Screws, Journal of Biomechanical Engineering, August, 118, pp. 391-398.
  4. 4. Kim, Y., Choi, W., Rhyu, K., 2012. Assessment of Pedicle Screw Pullout Strength Based on Various Screw Designs and Bone Densities-an Ex Vivo Biomechanical Study, The Spine Journal, 12, pp. 164–168.
  5. 5. Amaritsakul, Y., Chao, C., Jin, C., 2014. Comparison Study of the Pullout Strength of Conventional Spinal Pedicle Screws and a Novel Design in Full and Backed-out Insertions using Mechanical Tests, Proc IMechE Part H:J Engineering in Medicine, 228(3) pp. 250–257.
  6. 6. Inceoglu, S., Kılınçer, C., McLain, R.F., 2008. Screw Design Alters the Effects of Stress Relaxation on Pullout, Bio-Medical Materials and Engineering, 18, pp. 53–60.
  7. 7. Yaman, O., Demir, T., Arslan, A.K., İyidiker, M.E., Tolunay, T., Camuscu, N., Ulutas, M., 2015. The Comparison of Pullout Strengths of Various Pedicle Screw Designs on Synthetic Foams and Ovine Vertebrae, Turk Neurosurg, 25, 4, pp. 532-538.
  8. 8. Perren, S.M., 2002. Evolution of the Internal Fixation of Long Bone Fractures, The Scientific Basis of Biological Internal Fixation: choosing a new balance between stability and biology, J Bone Joint Surg., 84B, pp. 1093-1110.

Ayrıntılar

Birincil Dil

Türkçe

Konular

-

Bölüm

Araştırma Makalesi

Yazarlar

Ahmet Çetin Bu kişi benim
Türkiye

Yayımlanma Tarihi

15 Ekim 2016

Gönderilme Tarihi

29 Mayıs 2017

Kabul Tarihi

12 Ekim 2016

Yayımlandığı Sayı

Yıl 2016 Cilt: 31 Sayı: ÖS2

Kaynak Göster

APA
Bircan, D. A., & Çetin, A. (2016). Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı. Çukurova Üniversitesi Mühendislik-Mimarlık Fakültesi Dergisi, 31(ÖS2), 221-228. https://doi.org/10.21605/cukurovaummfd.316776
AMA
1.Bircan DA, Çetin A. Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı. cukurovaummfd. 2016;31(ÖS2):221-228. doi:10.21605/cukurovaummfd.316776
Chicago
Bircan, Durmuş Ali, ve Ahmet Çetin. 2016. “Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı”. Çukurova Üniversitesi Mühendislik-Mimarlık Fakültesi Dergisi 31 (ÖS2): 221-28. https://doi.org/10.21605/cukurovaummfd.316776.
EndNote
Bircan DA, Çetin A (01 Eylül 2016) Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı. Çukurova Üniversitesi Mühendislik-Mimarlık Fakültesi Dergisi 31 ÖS2 221–228.
IEEE
[1]D. A. Bircan ve A. Çetin, “Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı”, cukurovaummfd, c. 31, sy ÖS2, ss. 221–228, Eyl. 2016, doi: 10.21605/cukurovaummfd.316776.
ISNAD
Bircan, Durmuş Ali - Çetin, Ahmet. “Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı”. Çukurova Üniversitesi Mühendislik-Mimarlık Fakültesi Dergisi 31/ÖS2 (01 Eylül 2016): 221-228. https://doi.org/10.21605/cukurovaummfd.316776.
JAMA
1.Bircan DA, Çetin A. Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı. cukurovaummfd. 2016;31:221–228.
MLA
Bircan, Durmuş Ali, ve Ahmet Çetin. “Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı”. Çukurova Üniversitesi Mühendislik-Mimarlık Fakültesi Dergisi, c. 31, sy ÖS2, Eylül 2016, ss. 221-8, doi:10.21605/cukurovaummfd.316776.
Vancouver
1.Durmuş Ali Bircan, Ahmet Çetin. Ortopedik Operasyonlarda Kullanılan Medikal Vidaların Sıyırma Performansı. cukurovaummfd. 01 Eylül 2016;31(ÖS2):221-8. doi:10.21605/cukurovaummfd.316776