Research Article

AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN

Volume: 11 Number: 2 June 28, 2023
TR EN

AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN

Abstract

The test system was created by adding flex and force sensors to this manikin to evaluate the results of heart massage applied on an artificial adult manikin, which has a spring that can give the chest stiffness of an adult human. This test system monitors and analyzes cardiopulmonary resuscitation (CPR) applications performed by the automated CPR device or manually by the person. The signals received from the sensor are transferred to the computer via the serial port with the Arduino Uno card and displayed in real-time in MATLAB graphical user interface (GUI). This GUI, designed with Matlab 2021a software, analyzes the sensor signals resulting from CPR. It gives a graph of the repeats of compression per minute, the depth of each compression, and the compressions rate variable performed by the user or the automatic CPR device during the CPR application. This created test system can evaluate the accuracy of both the automatic CPR device performed in this study and the manual CPR application. The test system designed in this respect can be used in the training and evaluation of cardiac massage applications, which is included in the first aid courses in secondary education, associate degree, undergraduate, and vocational education courses.

Keywords

Supporting Institution

Karabük Üniversitesi

Project Number

FYL-2020-2265

Thanks

By Karabuk University Scientific Research Projects Coordination Unit thank you for the support given to this work with project number FYL-2020-2265.

References

  1. Araç, S. , Zengin, Y. , İçer, M. , Gündüz, E. , Dursun, R. , Durgun, H. , Üstündağ, M. , Orak, M. , Kuyumcu, M. & Güloğlu, C., 2021. Acil Serviste Kardiyopulmoner Resüsitasyon Yapılan Hastaların Değerlendirilmesi; Retrospektif Çalışma. Abant Tıp Dergisi, 10 (1), 140–151.
  2. Berkom, P., Noordergraaf, G., Scheffer, G., Noordergraaf, A., 2008. Does use of the CPREzyTM involve more work than CPR without feedback?. Resuscitation, 78, 66–70.
  3. Carl, B., & Brook, S., 1985. United States patent. Geothermics, 14(4), 595–599.
  4. Çolak, İ. ve Bayındır R., 2005. PIC16F877 ile DA Motor Hız Kontrolü, Pamukkale Üniversitesi Mühendislik Bilimleri Dergisi, 11 (2), 277–285.
  5. Ikeno, F., 2003. Improved survival with a novel chest compression device during a porcine model of cardiac arrest. Circulation, 108 (17), 381–383.
  6. Ikram, U. H., Jai, P. U., Sharda, U., Douglas, W. T., Jonathan J. S., ve Arno, L. Z., 2008. Chest compression quality and rescuer fatigue with increased compression to ventilation ratio during single rescuer pediatric CPR. Resuscitation, 79, 82–89.
  7. İnal, M., Yıldız, U., Öğütçü, S., Pekdemir, M., Ünlüer, T., Çelikhası, C., Ayüz, E., Işık, E., 2021. Tıp Eğitiminde CPR Mankenlerinin Yapay Zeka Sınıflandırıcı ile Dijital Dönüşümü. Veri Bilimi Dergisi, 4(3), 81–88.
  8. Majer, J., Jaguszewski, M.J., Frass, M., Leskiewicz, M., Smereka, J., Ladny, J.R., Robak, O., Szarpak, L., 2019. Does the use of cardiopulmonary resuscitation feedback devices improve the quality of chest compressions performed by doctors? A prospective, randomized, cross-over simulation study. Cardiology Journal, 26, 5, 529–535.

Details

Primary Language

English

Subjects

Engineering

Journal Section

Research Article

Publication Date

June 28, 2023

Submission Date

August 31, 2022

Acceptance Date

December 22, 2022

Published in Issue

Year 2023 Volume: 11 Number: 2

APA
Kavsaoğlu, A. R., & Tamokur, M. A. (2023). AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN. Mühendislik Bilimleri Ve Tasarım Dergisi, 11(2), 474-485. https://doi.org/10.21923/jesd.1168807
AMA
1.Kavsaoğlu AR, Tamokur MA. AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN. JESD. 2023;11(2):474-485. doi:10.21923/jesd.1168807
Chicago
Kavsaoğlu, Ahmet Reşit, and Mehmet Ali Tamokur. 2023. “AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN”. Mühendislik Bilimleri Ve Tasarım Dergisi 11 (2): 474-85. https://doi.org/10.21923/jesd.1168807.
EndNote
Kavsaoğlu AR, Tamokur MA (June 1, 2023) AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN. Mühendislik Bilimleri ve Tasarım Dergisi 11 2 474–485.
IEEE
[1]A. R. Kavsaoğlu and M. A. Tamokur, “AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN”, JESD, vol. 11, no. 2, pp. 474–485, June 2023, doi: 10.21923/jesd.1168807.
ISNAD
Kavsaoğlu, Ahmet Reşit - Tamokur, Mehmet Ali. “AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN”. Mühendislik Bilimleri ve Tasarım Dergisi 11/2 (June 1, 2023): 474-485. https://doi.org/10.21923/jesd.1168807.
JAMA
1.Kavsaoğlu AR, Tamokur MA. AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN. JESD. 2023;11:474–485.
MLA
Kavsaoğlu, Ahmet Reşit, and Mehmet Ali Tamokur. “AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN”. Mühendislik Bilimleri Ve Tasarım Dergisi, vol. 11, no. 2, June 2023, pp. 474-85, doi:10.21923/jesd.1168807.
Vancouver
1.Ahmet Reşit Kavsaoğlu, Mehmet Ali Tamokur. AUTOMATIC BELT CARDIOPULMONARY RECISUTATION (CPR) DEVICE AND TEST SYSTEM DESIGN. JESD. 2023 Jun. 1;11(2):474-85. doi:10.21923/jesd.1168807