Comparison of Central Corneal Thickness Measurements with Scheimpflug Photography, Optical Coherence Tomography and Ultrasonic Pachymetry Techniques

Year 2015, , 132 - 135, 15.07.2015
https://doi.org/10.18229/ktd.22397

Abstract

Objective: To compare Scheimpflug photography(SP),
optical coherence tomography(OCT) and ultrasonic pachymetry(USP) techniques in central corneal thickness(CCT) measurement.
Material and Methods: The CCT was measured in 43 healthy young adults(86 eyes in total) by using SP, OCT and
USP). Central corneal thickness, age, gender, intraocular
pressure and auto refractokeratometer values obtained
and used for comparison.
Results: The mean age of was 21,35 ± 0,75 (20-24) years. The mean CCT with USP was 558,42 ± 38,87 (477-
620) microns for right eye and 556,84 ± 39,83 (477-641)
microns for left eye. The mean CCT with SP was 555,02 ±
37,90 (486- 631) microns for right eye and 550,44 ± 42,30
(453-630) microns for left eye. The mean CCT with OCT
was 563, 52 ± 38,52 (492-644 microns for right eye and
550563,72 ± 39,84 (482-644) microns for left eye. Stataistically significant correlation found between OCT-USP
(P<0.003) and OCT-SP (P<0.001). OCT measurement
was significantly higher then other methods (about 7
micron). There was no statistically significant difference
between USP-SP.
Conclusion: CCT measurements with USP and SP were
relevant. OCT measurements were significantly higher
and others. This situation should be noted in regular CCT
measurements. 

References

  • Pflugfelder SC, Liu Z, Feuer W, et al. Corneal thickı- ness indices discriminate between keratoconus and contact lens-induced corneal thinning. Ophthalmo- logy 2002;109(12):2336-41.
  • Al-Mezaine HS, Al-Amro SA, Kangave D, et al. Com- parison between central corneal thickness measure- ments by oculus pentacam and ultrasonic pachymetry. International Ophthalmology. 2008;28(5):333–8.
  • Carrasco AJA, Vicent AD, Romín DM, et al. OCT tech- nique for evaluating diurnal variation in corneal thick- ness. Emmetropia 2013;4(4):197-201.
  • Şekil II: Optik kohorens tomografi Miglior S, Albe E, Guareschi M, et al. Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central cor- neal thickness. Br J Ophthalmol 2004;88(2):174–7.
  • Kawana K, Tokunaga T, Miyata K et al. Comparison of corneal thickness measurements using Orbscan II, noncontact specular microscopy, and ultrasonic pa- chymetry in eyes after laser in situ keratomileusis. Br J Ophthalmol 2004;88(4):466–8.
  • Piotrowiak I, Soldanska B, Burduk M, et al. Measu- ring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry. ISRN Ophthal- mol 2012:869319.
  • Leung DY, Lam DK, Yeung BY, Lam DS. Comparison between central corneal thickness measurements by ultrasound pachymetry and optical coherence tomog- raphy. Clin Experiment Ophthalmol 2006;34(8):751–4.
  • Grewal DS, Brar GS, Grewal SP. Assessment of central corneal thickness in normal, keratoconus, and post-la- ser in situ keratomileusis eyes using Scheimpflug ima- ging, spectral domain optical coherence tomography, and ultrasound pachymetry. J Cataract Refract Surg 2010;36(6):954–64
  • Fishman GR, Pons ME, Seedor JA, et al. Assessment of central corneal thickness using optical coherence tomography. J Cataract Refract Surg 2005; 31(4):707– 11
  • Buyuk K, Bozkurt B, Kamıs U, Ozkagnıcı A, Okudan S. Normal ve Keratokonuslu Gözlerde Ultrasonik Pa- kimetri ve OCULUS Pentacam ile Ölçülen Santral Kor- nea Kalınlıklarının Karşılaştırılması. Turk J Ophthalmol 2011;41(2):104-7.

Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması

Year 2015, , 132 - 135, 15.07.2015
https://doi.org/10.18229/ktd.22397

Abstract

Amaç: Santral kornea kalınlığının(SKK) ölçümünde ultrasonik pakimetri(UP), Scheimpflug fotograflama(SF) ve optik kohorens tomografi(OKT) yöntemleri ile elde edilen değerlerin karşılaştırılması.
Gereç ve Yöntem: 43 sağlıklı genç erişkinin sağ ve sol göz (toplam 86 göz) santral korneal kalınlığı UP, SF ve OKT yöntemleri ile ölçüldü. Santral kornea kalınlığı, yaş, cinsiyet, göz içi basıncı ve otorefraktokeratometre değerleri kaydedildi. Elde edilen değerler karşılaştırıldı.
Bulgular: Ortalama yaş 21,35 ± 0,75 (20-24) idi. UP ile ölçülen ortalama SKK sağ göz için 558,42 ± 38,87 (477-620) mikron, sol göz için 556,84 ± 39,83 (477-641) mikron idi.SF ile ölçülen ortalama SKK sağ göz için 555,02 ± 37,90 (486-631) mikron, sol göz için 550,44 ± 42,30 (453-630) mikron idi.OKT ile ölçülen ortalama SKK sağ göz için 563, 52 ± 38,52 (492-644) mikron, sol göz için 563,72 ± 39,84(482-644) mikron idi.Her üç yöntem ile ölçülen değerler karşılatırıldığında UP ve SF ile elde edilen değerler arasında istatistiki olarak anlamlı fark saptanmaz iken OKT ile elde edilen değerler ile arasında UP(P<0.003) ve SF(P<0.001) ile elde edilen değerler arasında istatistiki olarak anlamlı fark tespit edildi. OKT ile elde edilen değerler daha yüksek bulundu (yaklaşık 7 mikron).
Sonuç: Yapılan SKK ölçümünde UP ve SF ile elde edilen değerler birbirleri ile uyumludur. OKT ile elde edilen değerler daha yüksektir. Günlük uygulamada bu durum dikkate alınmalıdır.

 
Detect languageAfrikaansAlbanianArabicArmenianAzerbaijaniBasqueBelarusianBengaliBosnianBulgarianCatalanCebuanoChichewaChinese (Simplified)Chinese (Traditional)CroatianCzechDanishDutchEnglishEsperantoEstonianFilipinoFinnishFrenchGalicianGeorgianGermanGreekGujaratiHaitian CreoleHausaHebrewHindiHmongHungarianIcelandicIgboIndonesianIrishItalianJapaneseJavaneseKannadaKazakhKhmerKoreanLaoLatinLatvianLithuanianMacedonianMalagasyMalayMalayalamMalteseMaoriMarathiMongolianMyanmar (Burmese)NepaliNorwegianPersianPolishPortuguesePunjabiRomanianRussianSerbianSesothoSinhalaSlovakSlovenianSomaliSpanishSundaneseSwahiliSwedishTajikTamilTeluguThaiTurkishUkrainianUrduUzbekVietnameseWelshYiddishYorubaZulu AfrikaansAlbanianArabicArmenianAzerbaijaniBasqueBelarusianBengaliBosnianBulgarianCatalanCebuanoChichewaChinese (Simplified)Chinese (Traditional)CroatianCzechDanishDutchEnglishEsperantoEstonianFilipinoFinnishFrenchGalicianGeorgianGermanGreekGujaratiHaitian CreoleHausaHebrewHindiHmongHungarianIcelandicIgboIndonesianIrishItalianJapaneseJavaneseKannadaKazakhKhmerKoreanLaoLatinLatvianLithuanianMacedonianMalagasyMalayMalayalamMalteseMaoriMarathiMongolianMyanmar (Burmese)NepaliNorwegianPersianPolishPortuguesePunjabiRomanianRussianSerbianSesothoSinhalaSlovakSlovenianSomaliSpanishSundaneseSwahiliSwedishTajikTamilTeluguThaiTurkishUkrainianUrduUzbekVietnameseWelshYiddishYorubaZulu       
 
 
Text-to-speech function is limited to 100 characters
 

References

  • Pflugfelder SC, Liu Z, Feuer W, et al. Corneal thickı- ness indices discriminate between keratoconus and contact lens-induced corneal thinning. Ophthalmo- logy 2002;109(12):2336-41.
  • Al-Mezaine HS, Al-Amro SA, Kangave D, et al. Com- parison between central corneal thickness measure- ments by oculus pentacam and ultrasonic pachymetry. International Ophthalmology. 2008;28(5):333–8.
  • Carrasco AJA, Vicent AD, Romín DM, et al. OCT tech- nique for evaluating diurnal variation in corneal thick- ness. Emmetropia 2013;4(4):197-201.
  • Şekil II: Optik kohorens tomografi Miglior S, Albe E, Guareschi M, et al. Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central cor- neal thickness. Br J Ophthalmol 2004;88(2):174–7.
  • Kawana K, Tokunaga T, Miyata K et al. Comparison of corneal thickness measurements using Orbscan II, noncontact specular microscopy, and ultrasonic pa- chymetry in eyes after laser in situ keratomileusis. Br J Ophthalmol 2004;88(4):466–8.
  • Piotrowiak I, Soldanska B, Burduk M, et al. Measu- ring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry. ISRN Ophthal- mol 2012:869319.
  • Leung DY, Lam DK, Yeung BY, Lam DS. Comparison between central corneal thickness measurements by ultrasound pachymetry and optical coherence tomog- raphy. Clin Experiment Ophthalmol 2006;34(8):751–4.
  • Grewal DS, Brar GS, Grewal SP. Assessment of central corneal thickness in normal, keratoconus, and post-la- ser in situ keratomileusis eyes using Scheimpflug ima- ging, spectral domain optical coherence tomography, and ultrasound pachymetry. J Cataract Refract Surg 2010;36(6):954–64
  • Fishman GR, Pons ME, Seedor JA, et al. Assessment of central corneal thickness using optical coherence tomography. J Cataract Refract Surg 2005; 31(4):707– 11
  • Buyuk K, Bozkurt B, Kamıs U, Ozkagnıcı A, Okudan S. Normal ve Keratokonuslu Gözlerde Ultrasonik Pa- kimetri ve OCULUS Pentacam ile Ölçülen Santral Kor- nea Kalınlıklarının Karşılaştırılması. Turk J Ophthalmol 2011;41(2):104-7.
There are 10 citations in total.

Details

Primary Language Turkish
Subjects Clinical Sciences
Journal Section Articles
Authors

Mustafa Doğan

Nurgül Avcı This is me

Mehmet Durmaz This is me

Fatih Kartal This is me

Mehmet Akay This is me

Sibel İnan This is me

Publication Date July 15, 2015
Published in Issue Year 2015

Cite

APA Doğan, M., Avcı, N., Durmaz, M., Kartal, F., et al. (2015). Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması. Kocatepe Tıp Dergisi, 16(2), 132-135. https://doi.org/10.18229/ktd.22397
AMA Doğan M, Avcı N, Durmaz M, Kartal F, Akay M, İnan S. Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması. KTD. July 2015;16(2):132-135. doi:10.18229/ktd.22397
Chicago Doğan, Mustafa, Nurgül Avcı, Mehmet Durmaz, Fatih Kartal, Mehmet Akay, and Sibel İnan. “Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi Ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması”. Kocatepe Tıp Dergisi 16, no. 2 (July 2015): 132-35. https://doi.org/10.18229/ktd.22397.
EndNote Doğan M, Avcı N, Durmaz M, Kartal F, Akay M, İnan S (July 1, 2015) Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması. Kocatepe Tıp Dergisi 16 2 132–135.
IEEE M. Doğan, N. Avcı, M. Durmaz, F. Kartal, M. Akay, and S. İnan, “Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması”, KTD, vol. 16, no. 2, pp. 132–135, 2015, doi: 10.18229/ktd.22397.
ISNAD Doğan, Mustafa et al. “Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi Ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması”. Kocatepe Tıp Dergisi 16/2 (July 2015), 132-135. https://doi.org/10.18229/ktd.22397.
JAMA Doğan M, Avcı N, Durmaz M, Kartal F, Akay M, İnan S. Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması. KTD. 2015;16:132–135.
MLA Doğan, Mustafa et al. “Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi Ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması”. Kocatepe Tıp Dergisi, vol. 16, no. 2, 2015, pp. 132-5, doi:10.18229/ktd.22397.
Vancouver Doğan M, Avcı N, Durmaz M, Kartal F, Akay M, İnan S. Santral Kornea Kalınlığının Ölçümünde Scheimpflug Fotoğraflama, Optik Kohorens Tomografi ve Ultrasonik Pakimetri Yöntemlerinin Karşılaştırılması. KTD. 2015;16(2):132-5.

88x31.png
Bu Dergi Creative Commons Atıf-GayriTicari-AynıLisanslaPaylaş 4.0 Uluslararası Lisansı ile lisanslanmıştır.