In this study, radon background fluctuation effect on the measurement of 214Pb, 214Bi and 226Ra low level activity concentrations were investigated in ventilated laboratory. For this purpose, detectors which had 150% and 20% efficiencies were used. In spite of the air ventilation, radon background fluctuation effect on the low level activity measurement was observed. 20% efficiency detector measured activity discrepancies were 5.7% for 214Pb, 8.7% for 214Bi and 5.8% for 226Ra. 150% efficiency detector measured activity discrepancies were 13.8% for 214Pb, 11.5% for 214Bi and 12.7% for 226Ra.
Journal Section | Articles |
---|---|
Authors | |
Publication Date | May 12, 2017 |
Published in Issue | Year 2017 Volume: 21 Issue: 2 |
e-ISSN :1308-6529
Linking ISSN (ISSN-L): 1300-7688
All published articles in the journal can be accessed free of charge and are open access under the Creative Commons CC BY-NC (Attribution-NonCommercial) license. All authors and other journal users are deemed to have accepted this situation. Click here to access detailed information about the CC BY-NC license.